Accuracy and consistency of effective atomic number over object size using deep silicon photon-counting detector CT

被引:0
作者
Shapiro, Teva N. [1 ]
Salyapongse, Aria M. [1 ]
Lubner, Meghan G. [1 ]
Toia, Giuseppe, V [1 ]
Yin, Zhye [4 ]
Slavic, Scott [4 ]
Szczykutowicz, Timothy P. [1 ,2 ,3 ]
机构
[1] Univ Wisconsin Madison, Dept Radiol, Madison, WI USA
[2] Univ Wisconsin Madison, Dept Med Phys, Madison, WI USA
[3] Univ Wisconsin Madison, Dept Biomed Engn, Madison, WI USA
[4] GE Healthcare, Niskayuna, WI USA
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2025年 / 131卷
关键词
CT; Photon-counting; Effective atomic number; COMPUTED-TOMOGRAPHY; ENERGY; SIMULATION; DENSITY; PHANTOM;
D O I
10.1016/j.ejmp.2025.104945
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Photon-counting detector (PCD) CT is the newest generation of CT detector technology. It is critical to characterize its performance in measuring important biomarkers used in quantitative CT including effective atomic number (Zeff). More accurate Zeff measurements could be beneficial in tissue classification and proton therapy tasks. Methods: A phantom of varying water-equivalent diameter (WED) containing clinically relevant inserts was scanned using a prototype deep silicon PCD CT and a dual-energy (DE) energy integrating detector (EID) CT. Zeff maps were generated. Measured Zeff values were compared across WEDs and to theoretical values. Results: The measured Zeff of the polystyrene, solid water, iodine, and bone (50% CaCO3) inserts differed from the theoretical value by a maximum of-14.0%, 4.6%, 8.4% and 13.0% respectively on EID vs 4.5%, 5.2%, 2.3% and 7.2% on PCD. The maximum variation in Zeff over the WED range on EID was 0.71, 0.27, 0.66, and 1.22 vs 0.47, 0.64, 0.1 and 0.22 on PCD for polystyrene, solid water, iodine, and bone (50% CaCO3) respectively. Conclusion: This is the first study to evaluate Zeff measurements made using a prototype whole body PCD CT system. We found that PCD CT outperformed the EID CT in terms of Zeff accuracy and consistency over the WED range on most of the insert materials. Similarly, PCD CT outperformed most previous study's findings using EID CT. The high consistency and accuracy of measured Zeff using deep silicon PCD CT could make quantitative CT increasingly possible over a large range of patient sizes.
引用
收藏
页数:7
相关论文
共 39 条
  • [1] Mayneord W.V., The significance of the Röntgen, Acta — Unio Internationalis Contra Cancrum, 2, pp. 271-282, (1937)
  • [2] Spiers F.W., Effective atomic number and energy absorption in tissues, Br J Radiol, 19, 218, pp. 52-63, (1946)
  • [3] Vrbaski S., Pena L.M.A., Brombal L., Donato S., Taibi A., Contillo A., Et al.
  • [4] Mileto A., Allen B.C., Pietryga J.A., Farjat A.E., Zarzour J.G., Bellini D., Et al., Characterization of incidental renal mass with dual-energy CT: diagnostic accuracy of effective atomic number maps for discriminating nonenhancing cysts from enhancing masses, AJR Am J Roentgenol, 209, 4, pp. W221-W230, (2017)
  • [5] Nakajima S., Ito H., Mitsuhashi T., Kubo Y., Matsui K., Tanaka I., Et al., Clinical application of effective atomic number for classifying non-calcified coronary plaques by dual-energy computed tomography, Atherosclerosis, 261, pp. 160-162, (2017)
  • [6] Wohlfahrt P., Richter C., Status and innovations in pre-treatment CT imaging for proton therapy, Br J Radiol, 93, 1107, (2020)
  • [7] Peters N., Wohlfahrt P., Christian C., Mohler C., Menkel S., Tschiche M., Et al., Reduction of clinical safety margins in proton therapy enabled by the clinical implementation of dual-energy CT for direct stopping-power prediction, Radiother Oncol, 166, pp. 71-78, (2022)
  • [8] Hua C., Shapira N., Merchant T.E., Klahr P., Yagil Y., Accuracy of electron density, effective atomic number, and iodine concentration determination with a dual-layer dual-energy computed tomography system, Med Phys, 45, 6, pp. 2486-2497, (2018)
  • [9] Heismann B.J., Leppert J., Stierstorfer K., Density and atomic number measurements with spectral x-ray attenuation method, J Appl Phys, 1, 94, pp. 2073-2079, (2003)
  • [10] Szczykutowicz T.P., Qi Z., Chen G.H., A simple image based method for obtaining electron density and atomic number in dual energy CT, Proc of SPIE, 16, (2011)