Simulated radiation levels and patterns of MRI without a Faraday shielded room

被引:0
作者
Kazemivalipour, Ehsan [1 ,2 ]
Guerin, Bastien [1 ,2 ]
Wald, Lawrence L. [1 ,2 ,3 ]
机构
[1] Massachusetts Gen Hosp, AA Martinos Ctr Biomed Imaging, Dept Radiol, Charlestown, MA USA
[2] Harvard Med Sch, Boston, MA USA
[3] Harvard MIT Div Hlth Sci Technol, Cambridge, MA USA
基金
美国国家卫生研究院;
关键词
birdcage coil; magnetic resonance imaging; radiation pattern; safety; shielded room; ABSORPTION RATE; EXPLICIT CONTROL; PULSE DESIGN; GLOBAL SAR; LOCAL SAR; POWER; INTERFERENCE; PET;
D O I
10.1002/mrm.30499
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeWe characterize electromagnetic (EM) radiation patterns and levels in conventional MRI systems as a function of field strength and load symmetry, providing a framework for mitigation strategies allowing operation without a shielded room.MethodsWe simulated the far-field radiation pattern and fields at a 10 m radius (|E|10m and |B|10m) for a solenoidal superconducting MRI with a body birdcage coil operated between 0.25T and 6.5T. Five load configurations probed the impact of load-symmetry, ranging from a sphere to a body load (least-symmetric). We also assessed simple layered EM absorbers at the bore-ends.ResultsAll configurations exceeded regulatory limits for realistic transmit levels. At 1.5T, a 300 Vrms RF-pulse is 2700-fold the |E|10m limit. Field strength and load symmetry strongly modulate radiation patterns and levels. The radiated power increased by more than four orders of magnitude from 0.25T to 6.5T. Spherical load radiation transitioned from a peak gain at the bore-ends (0.25-0.5T) to a donut-shaped pattern, suggesting current loops around the bore (1 T-1.5T), back to bore-axis-directed gain, suggesting propagating waves along the bore (2T-6.5T). Transition patterns were seen between these regimes; uniform radiation at 0.75T and a combined donut/bore-directed pattern at 1.75T. Load asymmetry increased both strength and pattern asymmetry, with the body load having the highest and least symmetric radiation with the legs facilitating wave propagation at high-fields. A simple optimized layered absorber at scanner's service-end reduced 3T peak radiation by 11 dB.ConclusionRadiation from unshielded scanners far exceeds regulatory limits, particularly at high-field. Mitigation strategies must address load-symmetry, field strength, and wave effects.
引用
收藏
页码:835 / 851
页数:17
相关论文
共 59 条
[1]   Low-field MRI: Clinical promise and challenges [J].
Arnold, Thomas Campbell ;
Freeman, Colbey W. ;
Litt, Brian ;
Stein, Joel M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2023, 57 (01) :25-44
[2]  
Balanis C. A., 2016, Antenna theory: analysis and design
[3]  
Biber S, 2025, MAGN RESON MED, V93, P1842, DOI 10.1002/mrm.30374
[4]   ESTIMATING RADIOFREQUENCY POWER DEPOSITION IN BODY NMR IMAGING [J].
BOTTOMLEY, PA ;
REDINGTON, RW ;
EDELSTEIN, WA ;
SCHENCK, JF .
MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (04) :336-349
[5]   Travelling-wave nuclear magnetic resonance [J].
Brunner, David O. ;
De Zanche, Nicola ;
Froehlich, Juerg ;
Paska, Jan ;
Pruessmann, Klaas P. .
NATURE, 2009, 457 (7232) :994-U2
[6]   The effects of temperature and frequency on the dielectric properties, electromagnetic interference shielding and microwave-absorption of short carbon fiber/silica composites [J].
Cao, Mao-Sheng ;
Song, Wei-Li ;
Hou, Zhi-Ling ;
Wen, Bo ;
Yuan, Jie .
CARBON, 2010, 48 (03) :788-796
[7]   Computation design and performance prediction towards a multi-layer microwave absorber [J].
Cao, MS ;
Zhu, J ;
Yuan, J ;
Zhang, TF ;
Peng, ZH ;
Gao, ZJ ;
Xiao, G ;
Qin, SM .
MATERIALS & DESIGN, 2002, 23 (06) :557-564
[8]  
CISPR 11, 2016, Industrial, scientific and medical equipment, Radiofrequency disturbance characteristics, Limits and methods of measurement
[9]   Signal-to-noise ratio and absorbed power as functions of main magnetic field strength, and definition of "90°" RF pulse for the head in the birdcage coil [J].
Collins, CM ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (04) :684-691
[10]   A portable scanner for magnetic resonance imaging of the brain [J].
Cooley, Clarissa Z. ;
McDaniel, Patrick C. ;
Stockmann, Jason P. ;
Srinivas, Sai Abitha ;
Cauley, Stephen F. ;
Sliwiak, Monika ;
Sappo, Charlotte R. ;
Vaughn, Christopher F. ;
Guerin, Bastien ;
Rosen, Matthew S. ;
Lev, Michael H. ;
Wald, Lawrence L. .
NATURE BIOMEDICAL ENGINEERING, 2021, 5 (03) :229-239