Influence of Implant ScanBody Material and Intraoral Scanners on the Accuracy of Complete-Arch Digital Implant Impressions

被引:4
作者
Azevedo, Luis [1 ,2 ,3 ]
Marques, Tiago [2 ,4 ]
Karasan, Duygu [3 ]
Fehmer, Vincent [3 ]
Sailer, Irena [3 ]
Correia, Andre [2 ,4 ]
Polo, Miguel Gomez [1 ]
机构
[1] Univ Complutense Madrid, Dept Conservat Dent & Orofacial Prostheses, Madrid, Spain
[2] Univ Catolica Portuguesa, Ctr Interdisciplinary Res Hlth, Viseu, Portugal
[3] Univ Geneva, Univ Clin Dent Med, Div Fixed Prosthodont & Biomat, Geneva, Switzerland
[4] Univ Catol Portuguesa, Dept Dent Med, Viseu, Portugal
关键词
PRECISION; TRUENESS;
D O I
10.11607/ijp.8565
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To evaluate the accuracy of complete-arch digital implant impressions using different intraoral scan body (ISB) materials and intraoral scanners (IOSs). Materials and Methods: The mandibular dental cast of an edentulous patient with six tissue-level dental implants was used as the master cast. Two types of ISBs-polyether ether ketone (PEEK) and plasma-coated medical titanium-were used with five IOSs: TRIOS 4 (T4), Virtuo Vivo (VV), Medit i700 (Mi700), iTero 5D (i5D), and Primescan (PS). To assess accuracy, digital impressions (n = 10) with each IOS and ISB were compared to two reference models obtained by digitizing the master cast with each ISB type using a desktop scanner (IScan4D LS3i) and importing the scan data into metrology software (Geomagic Control X). Root mean square (RMS) error was employed to evaluate overall deviation values (trueness), while precision was determined using the SD of RMS values. Statistical significance was set at P < .05. Kruskal-Wallis test was used, followed by the pairwise comparison method with Bonferroni correction (alpha = .05). Results:An interaction between ISB material and IOS was found (P = .001). Plasma-coated medical titanium ISBs demonstrated significantly higher trueness and precision compared to PEEK ISBs with T4 (P = .001), Mi700 (P = .001; P = .004), and i5D (P = .001). Conversely, VV exhibited higher trueness and precision values with PEEK ISBs (P = .005; P = .003). PS provided the highest trueness and precision regardless of the ISB material (P = .912). T4 showed the lowest accuracy for PEEK ISBs, and VV showed the lowest accuracy for plasma-coated medical titanium ISBs.Conclusions: Except for PS, all IOSs showed significant differences between ISB materials. PS demonstrated the highest accuracy with both ISB materials, whereas T4 had the lowest accuracy for PEEK ISBs, and VV showed the lowest accuracy for plasma-coated medical titanium ISBs
引用
收藏
页码:575 / 582a
页数:9
相关论文
共 39 条
[1]   Trueness and marginal fit of implant-supported complete-arch fixed prosthesis frameworks made of high-performance polymers and titanium: An explorative in-vitro study [J].
Abou-Ayash, Samir ;
Schimmel, Martin ;
Ozcan, Mutlu ;
Ozcelik, Burak ;
Bragger, Urs ;
Yilmaz, Burak .
JOURNAL OF DENTISTRY, 2021, 113
[2]   In vitro fit of CAD-CAM complete arch screw-retained titanium and zirconia implant prostheses fabricated on 4 implants [J].
AL-Meraikhi, Hadi ;
Yilmaz, Burak ;
McGlumphy, Edwin ;
Brantley, William ;
Johnston, William M. .
JOURNAL OF PROSTHETIC DENTISTRY, 2018, 119 (03) :409-416
[3]   Digital vs. conventional full-arch implant impressions: a comparative study [J].
Amin, Sarah ;
Weber, Hans Peter ;
Finkelman, Matthew ;
El Rafie, Khaled ;
Kudara, Yukio ;
Papaspyridakos, Panos .
CLINICAL ORAL IMPLANTS RESEARCH, 2017, 28 (11) :1360-1367
[4]   Effect of scanning speed, scanning pattern, and tip size on the accuracy of intraoral digital scans [J].
An, Hongseok ;
Langas, Eleni E. ;
Gill, Aria S. .
JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (06) :1160-1167
[5]  
[Anonymous], 1994, Accuracy (trueness and precision) of measurement methods and results - Part 1: General principies and defmitions
[6]   Influence of implant scanbody material, position and operator on the accuracy of digital impression for complete-arch: A randomized in vitro trial [J].
Arcuri, Lorenzo ;
Pozzi, Alessandro ;
Lio, Fabrizio ;
Rompen, Eric ;
Zechner, Werner ;
Nardi, Alessandra .
JOURNAL OF PROSTHODONTIC RESEARCH, 2020, 64 (02) :128-136
[7]   Comparison of measured deviations in digital implant scans depending on software and operator [J].
Cakmak, Gulce ;
Marques, Vinicius Rizzo ;
Donmez, Mustafa Borga ;
Lu, Wei-En ;
Abou-Ayash, Samir ;
Yilmaz, Burak .
JOURNAL OF DENTISTRY, 2022, 122
[8]   Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A Comparative In Vitro Study [J].
Di Fiore, Adolfo ;
Graiff, Lorenzo ;
Savio, Gianpaolo ;
Granata, Stefano ;
Basilicata, Michele ;
Bollero, Patrizio ;
Meneghello, Roberto .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2022, 19 (08)
[9]   Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study [J].
Diker, Emre ;
Terzioglu, Hakan ;
Gouveia, Diogo N. M. ;
Donmez, Mustafa Borga ;
Seidt, Jeremy ;
Yilmaz, Burak .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2023, 25 (02) :419-425
[10]   Comparison of the Accuracy of Optical Impression Systems in Three Different Clinical Situations [J].
Doukantzi, Maria ;
Mojon, Philippe ;
Todorovic, Ana ;
Chebib, Najla ;
Pjetursson, Bjarni Elvar ;
Maniewicz, Sabrina ;
Sailer, Irena .
INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2021, 34 (04) :511-517