Comparison of continuous non-invasive blood pressure measurement using Vitalstream™ to invasive Intraarterial pressure in pediatric surgery

被引:0
作者
Boretsky, Karen R. [1 ]
Nasr, Viviane G. [1 ]
Atkinson, Douglas [1 ]
Baruch, Martin [2 ]
机构
[1] Boston Childrens Hosp, Dept Anesthesiol Crit Care & Pain Med, 300 Longwood Ave, Boston, MA 02115 USA
[2] Caretaker Med, 941 Glenwood Stn Ln 301, Charlottesville, VA 22901 USA
关键词
Pediatric; Blood pressure; Noninvasive continuous blood pressure; monitoring; ACCURACY;
D O I
10.1016/j.jclinane.2025.111763
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Accurate blood pressure monitoring is essential in many clinical scenarios for adults and children and, when continuous measurement is critical, necessitates the insertion of an arterial line. A novel continuous non-invasive arterial pressure monitoring device using a pulse contour algorithm (Pulse Decomposition Analysis), VitalstreamTM, is approved by the United States Food and Drug Administration for use in adults. In this study the performance and accuracy of the device compared to intraarterial blood pressure monitoring were assessed in children ages 2-17 undergoing major surgeries. We report the results using comparison to aspects of the recently published ISO 81060-3:2022 standard for continuous automated blood pressure measurement. Methods: 31 children ages 2-17 years scheduled for major surgery requiring invasive arterial blood pressure monitoring were consented to participate. Systolic, diastolic, and mean arterial blood pressure readings were obtained from both systems during at least thirty minutes of simultaneous monitoring during hemodynamically stable periods of the surgical procedure and statistically compared. Results: The correlations of systolic and, diastolic, and mean arterial pressures were, respectively, 0.77, 0.68 and 0.7. The Bland-Altman comparisons yielded bias of -3.79 (9.74) mmHg, 1.72 (8.45) mmHg and 2.41 (8.75) mmHg respectively, for systolic, diastolic, and mean arterial pressures, (p < 0.001 for all comparisons). Concordances for systole, diastole and MAP were, respectively, 0.82, 0.85 and 0.83. Conclusions: Most values fell within +/-20mmhg of the corresponding arterial line values. While this meets the basic requirement for such devices published by professional societies, clinicians will need to be aware of the potential variances and make clinical decisions accordingly. The VitalstreamTM may offer low risk, accurate continuous pressure monitoring in children ages 2-17. Clinical trials.gov: NCT04817137 The trial was registered at clinicaltrials.gov (NCT048I7173).
引用
收藏
页数:8
相关论文
共 17 条
[1]  
[Anonymous], 2022, ISO 81060-3
[2]  
[Anonymous], [73] CFR - Code of Federal Regulations Title 21. Sec. 182.60 Synthetic flavoring substances and adjuvants., U.S. Food Drug Administration, 2020. [Online]. Available: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/cfrsearch.cfm?fr182.60. [Accessed: 04-Jun-2021].
[3]  
Baruch Martin C, 2011, Nonlinear Biomed Phys, V5, P1, DOI 10.1186/1753-4631-5-1
[4]   An Evaluation of the NICCI Monitor in Providing Continuous, Noninvasive Blood Pressure Readings in Children During Intraoperative Anesthetic Care [J].
Bekiroglu, Ismail ;
Owusu-Bediako, Kwaku ;
Rice-Weimer, Julie ;
Tobias, Joseph D. .
JOURNAL OF CLINICAL MEDICINE RESEARCH-CANADA, 2022, 14 (04) :158-164
[5]   A prospective comparison of invasive and non-invasive blood pressure in children undergoing cardiac catheterization [J].
Beringer, Richard ;
Keith, Alastair ;
Jones, Elin ;
Murphy, Tim ;
White, Paul .
PEDIATRIC ANESTHESIA, 2023, 33 (10) :816-822
[6]   THE RELATIONSHIP BETWEEN ARTERIAL PULSE-WAVE VELOCITY AND PULSE FREQUENCY AT DIFFERENT PRESSURES [J].
CALLAGHAN, FJ ;
BABBS, CF ;
BOURLAND, JD ;
GEDDES, LA .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1984, 8 (01) :15-18
[7]  
Committee on Standards and Practice Parameters (CSPP), 2020, Standards for Basic Anesthetic Monitoring
[8]   Accuracy of the CNAP monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure readings in the prone position [J].
Dewhirst, Elisabeth ;
Corridore, Marco ;
Klamar, Jan ;
Beebe, Allan ;
Rice, Julie ;
Barry, N'Diris ;
Tobias, Joseph D. .
JOURNAL OF CLINICAL ANESTHESIA, 2013, 25 (04) :309-313
[9]  
Flynn JT, 2017, PEDIATRICS, V140, DOI [10.1542/peds.2017-1904, 10.1542/peds.2017-3035]
[10]   Continuous Non-invasive finger cuff CareTaker® comparable to invasive intraarterial pressure in patients undergoing major intra-abdominal surgery [J].
Gratz, Irwin ;
Deal, Edward ;
Spitz, Francis ;
Baruch, Martin ;
Allen, I. Elaine ;
Seaman, Julia E. ;
Pukenas, Erin ;
Jean, Smith .
BMC ANESTHESIOLOGY, 2017, 17