Temperature dependence compensation of X-ray spectrometry with MiniPIX Timepix3 SiC detector

被引:0
|
作者
Kurucova, N. [1 ]
Granja, C. [2 ]
Hladik, D. [2 ]
Sagatova, A. [1 ]
Zatko, B. [3 ]
机构
[1] Slovak Univ Technol Bratislava, Inst Nucl & Phys Engn, Fac Elect Engn & Informat Technol, Ilkovicova 3, Bratislava 84104, Slovakia
[2] ADVACAM, Pergamenky 12, Prague 17000, Czech Republic
[3] Slovak Acad Sci, Inst Elect Engn, Dubravska Cesta 9, Bratislava 84104, Slovakia
来源
JOURNAL OF INSTRUMENTATION | 2025年 / 20卷 / 02期
关键词
Detector design and construction technologies and materials; Radiation-hard detectors; Radiation-hard electronics; X-ray detectors;
D O I
10.1088/1748-0221/20/02/C02033
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Timepix3 is a hybrid pixel radiation detector, from the CERN-designed Timepix family, which contains 65.536 pixels. The detectors based on Timepix3 read-out chips have started to be used in a wide range of applications, such as medicine, particle physics, neutron detection and space applications. For applications in harsh environments such as space, particle therapy and nuclear power engineering, silicon carbide (SiC) stands as suitable semiconductor sensor thanks to its chemical stability, wide temperature operation range and enhanced radiation hardness. Newly developed MiniPIX Timepix3 SiC cameras are undergoing physical and operating tests for various applications. Since their main applications are in harsh environments where they are exposed to a wide range of temperatures, it is important to examine and describe the behaviour of these detectors under the influence of different temperatures. Consequently, it is necessary to compensate these temperature effects. Thus, this study investigates the temperature dependence of the spectral (deposited energy) response of the Timepix3 detectors equipped with a 4H-SiC sensor over a temperature range of 10 degrees C to 60 degrees C, with a per-pixel energy calibration performed at 20 degrees C. The detector was exposed to fluorescence X-rays with energies between 8 keU and 58 keU and between 31 keU and 40 keU from the radioactive sources 133Ba and 152Eu, respectively. It was found that with increasing temperature there is a shift of the energy spectra to lower values, which increases with increasing incident energy. A compensation method is proposed to correct for the distortion in the measured energy range.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Timepix3 as X-ray detector for time resolved synchrotron experiments
    Yousef, Hazem
    Crevatin, Giulio
    Gimenez, Eva N.
    Horswell, Ian
    Omar, David
    Tartoni, Nicola
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 845 : 639 - 643
  • [2] Hard X-ray imaging and particle detection with TIMEPIX3
    Frojdh, Christer
    Krapohl, David
    Thungstrom, Goran
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XVIII, 2016, 9968
  • [3] Stray radiation produced in FLASH electron beams characterized by the MiniPIX Timepix3 Flex detector
    Oancea, C.
    Balan, C.
    Pivec, J.
    Granja, C.
    Jakubek, J.
    Chvatil, D.
    Olsansky, V
    Chis, V
    JOURNAL OF INSTRUMENTATION, 2022, 17 (01):
  • [4] Timepix3: Temperature influence on X-ray measurements in counting mode with Si sensor
    Urban, Martin
    Doubravova, Daniela
    RADIATION MEASUREMENTS, 2021, 141
  • [5] Timepix3: Temperature influence on X-ray measurements in counting mode with Si sensor
    Urban, Martin
    Doubravova, Daniela
    RADIATION MEASUREMENTS, 2021, 141
  • [6] Dose estimation in X-ray backscatter imaging with Timepix3 and TLD detectors
    Parra, J. S. Useche
    Schuetz, M. K.
    Roque, G.
    Fey, J.
    Avila, C.
    Fiederle, M.
    Procz, S.
    JOURNAL OF INSTRUMENTATION, 2023, 18 (05)
  • [7] Detection resolving power of SiC Timepix3 detector to electrons, neutrons, ions and protons
    Granja, Carlos
    Barber, Corinne
    Barna, Sandra
    Chancellor, Jeffery
    Chvatil, David
    Grevillot, Loic
    Inzalaco, David
    Jakubek, Jan
    Kohout, Zdenek
    Magrin, Giulio
    Marek, Lukas
    Mihai, Radu
    Oancea, Cristina
    Olsansky, Vaclav
    Olsen, Trevor
    Poklop, Dusan
    Pospisil, Stanislav
    Resch, Andreas
    Sagatova, Andrea
    Sivertz, Michael
    Stursa, Jan
    Sykora, Rudolf
    Wilhelm, Ivan
    Zache, Vaclav
    Zatko, Bohumir
    JOURNAL OF INSTRUMENTATION, 2024, 19 (11):
  • [8] Timepix: Influence of temperature and vacuum on equalisation of X-ray detector and its verification
    Urban, Martin
    Nentvich, Ondrej
    Doubravova, Daniela
    Petr, Ondrej
    Inneman, Adolf
    Hudec, Rene
    Sieger, Ladislav
    UV, X-RAY, AND GAMMA-RAY SPACE INSTRUMENTATION FOR ASTRONOMY XXI, 2019, 11118
  • [9] Timepix detector as a tool for X-ray and gamma dosimetry
    Rubovic, Peter
    Bergmann, Benedikt
    Ekendahl, Daniela
    Hulka, Jiri
    Judas, Libor
    Kohout, Zdenek
    Pospisil, Stanislav
    Stekl, Ivan
    RADIATION MEASUREMENTS, 2017, 107 : 39 - 42
  • [10] Timepix detector at the X-ray Correlation Spectroscopy instrument at LCLS
    van Bakel, N. A.
    Robert, A.
    Sikorski, M.
    Ford, C.
    van Beveren, V.
    van der Heijden, B.
    van Beuzekom, M.
    11TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI 2012), 2013, 425