Commissioning of a motion management system for a 1.5T Elekta Unity MR-Linac: A single institution experience

被引:0
作者
Smith, Blake R. [1 ]
St-Aubin, Joel [1 ]
Hyer, Daniel E. [1 ]
机构
[1] Univ Iowa, Dept Radiat Oncol, Iowa City, IA 52242 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2025年
关键词
commissioning; gating; motion management; MR-Linac; unity; ADAPTIVE RADIOTHERAPY; RESPIRATORY MOTION; RADIATION-THERAPY; QUALITY-ASSURANCE; IMPLEMENTATION; TRACKING; LUNG;
D O I
10.1002/acm2.70005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThis work describes a single institution experience of commissioning a real-time target tracking and beam control system, known as comprehensive motion management, for a 1.5 T Elekta MR-Linac.MethodsAnatomical tracking and radiation beam control were tested using the MRI4D Quasar motion phantom. Multiple respiratory breathing traces were modeled across a range of realistic regular and irregular breathing patterns ranging between 10 and 18 breaths per minute. Each of the breathing traces was used to characterize the anatomical position monitoring (APM) accuracy, and beam latency, and to quantify the dosimetric impact of both parameters during a respiratory-gated delivery using EBT3 film dosimetry. Additional commissioning tasks were performed to verify the dosimetric constancy during beam gating and to expand our existing quality assurance program.ResultsIt was determined that APM correctly predicted the 3D position of a dynamically moving tracking target to within 1.5 mm for 95% of the imaging frames with no deviation exceeding 2 mm. Among the breathing traces investigated, the mean latency ranged between -21.7 and 7.9 ms with 95% of all observed latencies within 188.3 ms. No discernable differences were observed in the relative profiles or cumulative output for a gated beam relative to an ungated beam with minimal dosimetric impact observed due to system latency. Measured dose profiles for all gated scenarios retained a gamma pass rate of 97% or higher for a 3%/2 mm criteria relative to a theoretical gated dose profile without latency or tracking inaccuracies.ConclusionMRI-guided target tracking and automated beam delivery control were successfully commissioned for the Elekta Unity MR-Linac. These gating features were shown to be highly accurate with an effectively small beam latency for a range of regular and irregular respiratory breathing traces.
引用
收藏
页数:13
相关论文
共 37 条
  • [1] Rankine L., Wan H., Parikh P., Et al., Cone-beam computed tomography internal motion tracking should be used to validate 4-dimensional computed tomography for abdominal radiation therapy patients, Int J Radiat Oncol Biol Phys, 95, 2, pp. 818-826, (2016)
  • [2] Seppenwoolde Y., Shirato H., Kitamura K., Et al., Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy, Int J Radiat Oncol Biol Phys, 53, 4, pp. 822-834, (2002)
  • [3] Hamming V.C., Visser C., Batin E., Et al., Evaluation of a 3D surface imaging system for deep inspiration breath-hold patient positioning and intra-fraction monitoring, Radiat Oncol, 14, pp. 1-8, (2019)
  • [4] Schwaab J., Prall M., Sarti C., Et al., Ultrasound tracking for intra-fractional motion compensation in radiation therapy, Physica Med, 30, 5, pp. 578-582, (2014)
  • [5] Charters J.A., Abdulkadir Y., O'connell D., Yang Y., Lamb J.M., Dosimetric evaluation of respiratory gating on a 0.35-T magnetic resonance–guided radiotherapy linac, J Appl Clin Med Phys, 23, 9, (2022)
  • [6] Godley A., Zheng D., Rong Y., MR-linac is the best modality for lung SBRT, J Appl Clin Med Phys, 20, 6, pp. 7-11, (2019)
  • [7] Green O.L., Rankine L.J., Cai B., Et al., First clinical implementation of real-time, real anatomy tracking and radiation beam control, Med Phys, 45, 8, pp. 3728-3740, (2018)
  • [8] Lee S.L., Mahler P., Olson S., Et al., Reduction of cardiac dose using respiratory-gated MR-linac plans for gastro-esophageal junction cancer, Med Dosim, 46, 2, pp. 152-156, (2021)
  • [9] Redler G., Stevens T., Cammin J., Et al., Dosimetric feasibility of utilizing the viewray magnetic resonance guided linac system for image-guided spine stereotactic body radiation therapy, Cureus, 11, 12, (2019)
  • [10] Tanabe Y., Kiritani M., Deguchi T., Hira N., Tomimoto S., Patient-specific respiratory motion management using lung tumors vs fiducial markers for real-time tumor-tracking stereotactic body radiotherapy, Phys Imaging Radiat Oncol, 25, (2023)