The MyFlex-ζ Foot: A Variable Stiffness ESR Ankle-Foot Prosthesis

被引:0
|
作者
Tabucol, Johnnidel [1 ]
Kooiman, Vera G. M. [2 ,3 ]
Leopaldi, Marco [1 ]
Leijendekkers, Ruud [4 ,5 ]
Selleri, Giacomo [1 ]
Mellini, Marcello [1 ]
Verdonschot, Nico [6 ,7 ]
Oddsson, Magnus [8 ]
Carloni, Raffaella [9 ]
Zucchelli, Andrea [1 ]
Brugo, Tommaso M. [1 ]
机构
[1] Univ Bologna, Dept Ind Engn, I-40131 Bologna, Italy
[2] Radboud Univ Nijmegen, Med Ctr, Orthoped Res Lab, Dept Rehabil, NL-6500 HB Nijmegen, Netherlands
[3] Radboud Univ Nijmegen, Med Ctr, Donders Inst Brain Cognit & Behav, NL-6500 HB Nijmegen, Netherlands
[4] Radboud Univ Nijmegen, Donders Inst Brain Cognit & Behav, Radboud Inst Hlth Sci, Orthoped Res Lab,Dept Rehabil,Med Ctr, NL-6500 HB Nijmegen, Netherlands
[5] Radboud Univ Nijmegen, Med Ctr, IQ Healthcare, NL-6500 HB Nijmegen, Netherlands
[6] Radboud Univ Nijmegen, Med Ctr, Orthoped Res Lab, NL-6500 HB Nijmegen, Netherlands
[7] Univ Twente, Fac Engn & Technol, Dept Biomech Engn, NL-7500 AE Enschede, Netherlands
[8] Ossur, IS-110 Reykjavik, Iceland
[9] Univ Groningen, Fac Sci & Engn, Dept Artificial Intelligence, Bernoulli Inst Math Comp Sci & Artificial Intelli, Groningen, Netherlands
关键词
Foot; Prosthetics; Legged locomotion; Ankle; Particle measurements; Force; Atmospheric measurements; Costs; Blades; Biomechanics; Amputation; biomechanical tests; prosthesis design; prosthetic feet; variable stiffness; TRANSTIBIAL AMPUTEES; POWERED KNEE; WALKING; AMPUTATION; DESIGN; ACTUATOR; FEET;
D O I
10.1109/TNSRE.2025.3534096
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Most commercially available foot prostheses are passive ESR feet, which store and release energy to reduce metabolic costs and improve comfort but cannot adjust to varying walking conditions. In contrast, bionic feet adapt to different tasks but are hindered by high weight, power consumption, and cost. This paper presents MyFlex- zeta , an ESR foot with a variable stiffness system, as a compromise between these two categories. MyFlex- zeta adjusts stiffness by varying the sagittal-plane distance between two key points, altering force interactions within the prosthesis and affecting overall stiffness. Clinical tests with three transfemoral amputees evaluated stiffness variation across two sessions: the first subjective, where participants assessed stiffness settings during different tasks, and the second biomechanical, measuring performance parameters. Two participants selected different stiffness settings for various tasks, while the third, with limited perception of stiffness changes, showed less distinction in outcomes. Greater sagittal-plane rotation and higher energy absorption were observed in most tasks with more compliant settings, although one participant's results were limited due to selecting close stiffness settings. Overall, these findings suggest MyFlex- zeta offers adaptability and performance improvements over traditional ESR feet. With further actuation and control system development, MyFlex- zeta could mark significant progress in prosthesis technology.
引用
收藏
页码:653 / 663
页数:11
相关论文
共 50 条
  • [11] Design of a powered ankle-foot prosthesis with an adjustable stiffness toe joint
    She, Haotian
    Zhu, Jinying
    Tian, Ye
    Wang, Yanchao
    Huang, Qiang
    ADVANCED ROBOTICS, 2020, 34 (10) : 689 - 697
  • [12] Design and characterization of a variable-stiffness ankle-foot orthosis
    Rong, Haochen
    Ramezani, Sepehr
    Ambro, Alex
    Lei, Chi Hou
    Choi, Hwan
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2024, 48 (06) : 720 - 726
  • [13] Implementation and Testing of Ankle-Foot Prosthesis With a New Compensated Controller
    Gao, Fei
    Liu, Yannan
    Liao, Wei-Hsin
    IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2019, 24 (04) : 1775 - 1784
  • [14] A Universal Ankle-Foot Prosthesis Emulator for Human Locomotion Experiments
    Caputo, Joshua M.
    Collins, Steven H.
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (03):
  • [15] Biomechanical response to ankle-foot orthosis stiffness during running
    Esposito, Elizabeth Russell
    Choi, Harmony S.
    Owens, Johnny G.
    Blanck, Ryan V.
    Wilken, Jason M.
    CLINICAL BIOMECHANICS, 2015, 30 (10) : 1125 - 1132
  • [16] A Robotic Ankle-Foot Prosthesis With Active Alignment
    LaPre, Andrew Kennedy
    Umberger, Brian R.
    Sup, Frank C.
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2016, 10 (02):
  • [17] Passive Ankle-Foot Prosthesis Prototype with Extended Push-Off
    Brackx, Branko
    Van Damme, Michael
    Matthys, Arnout
    Vanderborght, Bram
    Lefeber, Dirk
    INTERNATIONAL JOURNAL OF ADVANCED ROBOTIC SYSTEMS, 2013, 10
  • [18] STIFFNESS ANALYSIS OF A SIMPLIFIED ANKLE-FOOT ORTHOSIS
    Swierski, Ethan
    Burke, Molly
    Arenas, Maria
    Bernat, Jessica
    Manzer, James
    Piovesan, Davide
    PROCEEDINGS OF THE ASME 2020 INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, IMECE2020, VOL 5, 2020,
  • [19] Control of a Powered Ankle-Foot Prosthesis Based on a Neuromuscular Model
    Eilenberg, Michael F.
    Geyer, Hartmut
    Herr, Hugh
    IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2010, 18 (02) : 164 - 173
  • [20] A Study of Human Walking Biomechanics for Ankle-Foot Prosthesis Design
    Fandakli, Selin Aydin
    Okumus, Halil Ibrahim
    Ozturk, Mehmet
    2018 6TH INTERNATIONAL CONFERENCE ON CONTROL ENGINEERING & INFORMATION TECHNOLOGY (CEIT), 2018,