Comparison of radial immunodiffusion, turbidimetric immunoassay, and Brix refractometry for determining bovine colostrum quality

被引:1
|
作者
Westhoff, T. A. [1 ]
Behling-Kelly, E. L. [1 ]
Mann, S. [1 ]
机构
[1] Cornell Univ, Coll Vet Med, Dept Populat Med & Diagnost Sci, Ithaca, NY 14853 USA
来源
JDS COMMUNICATIONS | 2024年 / 5卷 / 06期
基金
美国食品与农业研究所;
关键词
IMMUNOGLOBULIN-G CONCENTRATION; IMMUNOTURBIDIMETRY; ACCURACY; IGG;
D O I
10.3168/jdsc.2024-0604
中图分类号
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号
0905 ;
摘要
Determining the concentration of IgG in colostrum is critical for assessment of colostrum quality. On-farm use of a Brix refractometer to estimate colostrum IgG concentration is widespread, whereas radial immunodiffusion (RID) is the laboratory reference method. Turbidimetric immunoassay (TIA) might offer an alternative method to quantify IgG in colostrum, but the agreement with RID, as well as critical thresholds to determine high-quality colostrum, remain uncertain. The objective of this study was to determine the level of agreement between RID, Brix %, and TIA for evaluation of colostrum quality. Composite colostrum samples (n = 58) from Holstein cows were evaluated using a digital Brix refractometer at the time of collection and stored at -20 degrees C until analysis. The concentration of IgG was determined using RID and TIA. Data were analyzed using Passing-Bablok regression and Bland-Altman plots. Critical thresholds for TIA and Brix measurements to identify colostrum with an IgG concentration >= 50 and >= 100 g/L based on the reference method were determined using logistic regression and receiver operating characteristic curves. Results revealed that both TIA (Kendall's'r = 0.91) and Brix % ('r = 0.78) had a strong correlation with RID. Passing-Bablok regression identified a systematic (6.91 [4.33 to 8.98] g/L) and proportional (0.69 [0.67 to 0.72] g/L) bias between RID and TIA. The optimum thresholds to identify samples with an IgG concentration >= 50 and >= 100 g/L were 40.6 g/L (area under the curve [AUC]: 1.0; sensitivity (Se): 100; specificity [Sp]: 100) and 85.8 g/L (AUC: 0.99; Se: 96.6; Sp: 96.6) for TIA and 18.4% (AUC: 1.0; Se: 100; Sp: 100) and 25.8% (AUC: 0.99; Se: 82.8; Sp: 93.1) for Brix %, respectively. Using the identified thresholds, our results show that both Brix and TIA were highly accurate for identifying high-quality colostrum, but because of a proportional bias, direct comparison of IgG concentration results obtained by RID and TIA are cautioned.
引用
收藏
页码:679 / 683
页数:6
相关论文
共 9 条
  • [1] Investigation of colostrum quality in beef cattle by radial immunodiffusion and brix refractometry
    Vandeputte, S.
    Detilleux, J.
    Rollin, F.
    VETERINARY RECORD, 2014, 175 (14) : 353
  • [2] Comparative estimation of colostrum quality by Brix refractometry in bovine, caprine, and ovine colostrum
    Kessler, E. C.
    Bruckmaier, R. M.
    Gross, J. J.
    JOURNAL OF DAIRY SCIENCE, 2021, 104 (02) : 2438 - 2444
  • [3] Evaluation of Brix Refractometry for the Estimation of Colostrum Quality in Jennies
    Turini, Luca
    Nocera, Irene
    Bonelli, Francesca
    Mele, Marcello
    Sgorbini, Micaela
    JOURNAL OF EQUINE VETERINARY SCIENCE, 2020, 92
  • [4] Associations between the metabolic status of the cow and colostrum quality as determined by Brix refractometry
    Immler, Manuela
    Failing, Klaus
    Gaertner, Tanja
    Wehrend, Axel
    Donat, Karsten
    JOURNAL OF DAIRY SCIENCE, 2021, 104 (09) : 10131 - 10142
  • [5] Technical note: Comparison of radial immunodiffusion and ELISA for quantification of bovine immunoglobulin G in colostrum and plasma
    Gelsinger, S. L.
    Smith, A. M.
    Jones, C. M.
    Heinrichs, A. J.
    JOURNAL OF DAIRY SCIENCE, 2015, 98 (06) : 4084 - 4089
  • [6] Comparison of single radial immunodiffusion and ELISA for the quantification of immunoglobulin G in bovine colostrum, milk and calf sera
    Dunn, Amanda
    Duffy, Catherine
    Gordon, Alan
    Morrison, Steven
    Arguello, Anastasio
    Welsh, Michael
    Earley, Bernadette
    JOURNAL OF APPLIED ANIMAL RESEARCH, 2018, 46 (01) : 758 - 765
  • [7] Evaluation of colostrum quality in the Czech Republic using radial immunodiffusion and different types of refractometers
    Pechova, Alena
    Slosarkova, Sona
    Stanek, Stanislav
    Nejedla, Eliska
    Fleischer, Petr
    VETERINARNI MEDICINA, 2019, 64 (02) : 51 - 59
  • [8] Diagnostic accuracy of refractometry for assessing bovine colostrum quality: A systematic review and meta-analysis
    Buczinski, S.
    Vandeweerd, J. M.
    JOURNAL OF DAIRY SCIENCE, 2016, 99 (09) : 7381 - 7394
  • [9] Validation of a Point-of-Care Quantitative Equine IgG Turbidimetric Immunoassay and Comparison of IgG Concentrations Measured with Radial Immunodiffusion and a Point-of-Care IgG ELISA
    Ujvari, S.
    Schwarzwald, C. C.
    Fouche, N.
    Howard, J.
    Schoster, A.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2017, 31 (04): : 1170 - 1177