Optimizing Stem Length in Conversion Total Hip Arthroplasty: An Expanded Finite Element Analysis

被引:0
作者
Shimasaki, Koshiro [1 ]
Nishino, Tomofumi [1 ]
Yoshizawa, Tomohiro [1 ]
Watanabe, Ryunosuke [1 ]
Hirose, Fumi [1 ]
Yasunaga, Shota [1 ]
Mishima, Hajime [1 ]
机构
[1] Univ Tsukuba, Inst Med, Dept Orthopaed Surg, 1-1-1 Tennodai, Tsukuba 3058575, Japan
关键词
conversion total hip arthroplasty (cTHA); finite element analysis (FEA); stress concentration; periprosthetic fracture; stem length; FAILED INTERNAL-FIXATION; INTERTROCHANTERIC FRACTURE; PERIPROSTHETIC FRACTURES; CORTICAL DEFECTS; FEMORAL-NECK; SALVAGE; FEMUR; COMPLICATIONS; STRESS; BIOMECHANICS;
D O I
10.3390/jcm14041141
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background/Objectives: Stress concentration around distal screw-removal holes confers a major risk for periprosthetic fractures following conversion total hip arthroplasty (cTHA) for intertrochanteric femoral fractures. Optimal stem-selection criteria and guidelines for cTHA can improve clinical outcomes. We determined the influence of the cementless stem length on the stress distribution around distal screw-removal holes. Methods: For the finite element analysis, institutional data from preoperative CT scans of contralateral femurs of patients who underwent THA were used. To replicate the post-nail-removal state, we used 3D registration of standard triangulated language data of the intramedullary nail as an unused material to simulate distal screw-removal holes, located 135 mm from the proximal end of the intramedullary nail. Cementless stems of 130, 140, 150, and 160 mm were individually registered using STL data, and cTHA models were constructed accordingly. Using simulations under load conditions representing normal walking and stair climbing, the mean and maximum equivalent stress values around the distal screw-removal holes were calculated. For multiple comparisons, repeated-measures ANOVA with Bonferroni correction was employed. Results: Compared to the 130 mm stem, the 150 mm and 160 mm stems similarly reduced the maximum equivalent stress around the distal screw-removal holes, although the 140 mm stem showed no significant difference with other stems. Conclusions: A >= 150 mm stem length reliably mitigated stress concentration around distal screw-removal holes post-cTHA; it is the optimal choice for balancing effectiveness and risk of complications and may contribute to improved long-term clinical outcomes. This study provides practical evidence for stem selection in cTHA and offers valuable insights for future treatment guidelines.
引用
收藏
页数:18
相关论文
共 50 条
  • [41] The outcomes of conversion of hemiarthroplasty to total hip arthroplasty, a systematic review and meta-analysis
    Schmitz, Peter P.
    van Susante, Job L. C.
    Sierevelt, Inger N.
    Somford, Matthijs P.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2024, 144 (07) : 2993 - 3001
  • [42] Perioperative Predictors of Length of Stay After Total Hip Arthroplasty
    Petis, Stephen M.
    Howard, James L.
    Lanting, Brent A.
    Somerville, Lyndsay E.
    Vasarhelyi, Edward M.
    JOURNAL OF ARTHROPLASTY, 2016, 31 (07) : 1427 - 1430
  • [43] Impact of the Femoral Head Position on Moment Arms in Total Hip Arthroplasty: A Parametric Finite Element Study
    Ruediger, Hannes A.
    Parvex, Valerie
    Terrier, Alexandre
    JOURNAL OF ARTHROPLASTY, 2016, 31 (03) : 715 - 720
  • [44] Finite element model of a novel short stemmed total hip arthroplasty implant developed from cross sectional CT scans
    Lerch, Matthias
    Weigel, Nelly
    Windhagen, Henning
    Ettinger, Max
    Thorey, Fritz
    Kurtz, Agnes
    Stukenborg-Colsman, Christina
    Bouguecha, Anas
    TECHNOLOGY AND HEALTH CARE, 2013, 21 (05) : 493 - 500
  • [45] Outcomes and survivorship after conversion of failed hip hemiarthroplasty to total hip arthroplasty
    Sarpong, Nana O.
    Grosso, Matthew J.
    Lakra, Akshay
    Herndon, Carl L.
    Jennings, Emma
    Shah, Roshan P.
    Geller, Jeffrey A.
    Cooper, H. John
    HIP INTERNATIONAL, 2021, 31 (03) : 388 - 392
  • [46] Hemiarthroplasty Conversion: A Comparison to Primary and Revision Total Hip Arthroplasty
    Sarpong, Nana O.
    Grosso, Matthew J.
    Lakra, Akshay
    Held, Michael B.
    Herndon, Carl L.
    Cooper, H. John
    JOURNAL OF ARTHROPLASTY, 2019, 34 (06) : 1168 - 1173
  • [47] Are leg length and hip offset comparable after hip resurfacing and cementless total hip arthroplasty?
    Patel, S.
    Thakrar, R. R.
    Bhamra, J.
    Hossain, F.
    Tengrootenhuysen, M.
    Haddad, F. S.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2011, 93 (06) : 465 - 469
  • [48] Finite element analysis of a novel design approach to resisting total hip dislocation
    Scifert, CF
    Brown, TD
    Lipman, JD
    CLINICAL BIOMECHANICS, 1999, 14 (10) : 697 - 703
  • [49] Conversion to total hip arthroplasty in posttraumatic arthritis: short-term clinical outcomes
    Taheriazam, Afshin
    Saeidinia, Amin
    ORTHOPEDIC RESEARCH AND REVIEWS, 2019, 11 : 41 - 46
  • [50] The effect of the hip joint endoprosthesis length after a total hip arthroplasty: A biomechanical study
    Smyris, Andreas-Foivos
    Potsika, Vassiliki T.
    Farmakis, Ioannis-Ilias K.
    Tachos, Nikolaos
    Fotiadis, Dimitrios, I
    Xenakis, Theodoros A.
    Pakos, Emilios E.
    JOURNAL OF ORTHOPAEDICS TRAUMA AND REHABILITATION, 2019, 26 (01) : 61 - 66