Optimizing Stem Length in Conversion Total Hip Arthroplasty: An Expanded Finite Element Analysis

被引:0
|
作者
Shimasaki, Koshiro [1 ]
Nishino, Tomofumi [1 ]
Yoshizawa, Tomohiro [1 ]
Watanabe, Ryunosuke [1 ]
Hirose, Fumi [1 ]
Yasunaga, Shota [1 ]
Mishima, Hajime [1 ]
机构
[1] Univ Tsukuba, Inst Med, Dept Orthopaed Surg, 1-1-1 Tennodai, Tsukuba 3058575, Japan
关键词
conversion total hip arthroplasty (cTHA); finite element analysis (FEA); stress concentration; periprosthetic fracture; stem length; FAILED INTERNAL-FIXATION; INTERTROCHANTERIC FRACTURE; PERIPROSTHETIC FRACTURES; CORTICAL DEFECTS; FEMORAL-NECK; SALVAGE; FEMUR; COMPLICATIONS; STRESS; BIOMECHANICS;
D O I
10.3390/jcm14041141
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background/Objectives: Stress concentration around distal screw-removal holes confers a major risk for periprosthetic fractures following conversion total hip arthroplasty (cTHA) for intertrochanteric femoral fractures. Optimal stem-selection criteria and guidelines for cTHA can improve clinical outcomes. We determined the influence of the cementless stem length on the stress distribution around distal screw-removal holes. Methods: For the finite element analysis, institutional data from preoperative CT scans of contralateral femurs of patients who underwent THA were used. To replicate the post-nail-removal state, we used 3D registration of standard triangulated language data of the intramedullary nail as an unused material to simulate distal screw-removal holes, located 135 mm from the proximal end of the intramedullary nail. Cementless stems of 130, 140, 150, and 160 mm were individually registered using STL data, and cTHA models were constructed accordingly. Using simulations under load conditions representing normal walking and stair climbing, the mean and maximum equivalent stress values around the distal screw-removal holes were calculated. For multiple comparisons, repeated-measures ANOVA with Bonferroni correction was employed. Results: Compared to the 130 mm stem, the 150 mm and 160 mm stems similarly reduced the maximum equivalent stress around the distal screw-removal holes, although the 140 mm stem showed no significant difference with other stems. Conclusions: A >= 150 mm stem length reliably mitigated stress concentration around distal screw-removal holes post-cTHA; it is the optimal choice for balancing effectiveness and risk of complications and may contribute to improved long-term clinical outcomes. This study provides practical evidence for stem selection in cTHA and offers valuable insights for future treatment guidelines.
引用
收藏
页数:18
相关论文
共 50 条
  • [21] Conversion Total Hip Arthroplasty in the Era of Bundled Care Payments Impacts on Costs of Care
    Hunter, Michael D.
    Desmarais, Jason D.
    Quilligan, Edward J.
    Scudday, Travis S.
    Patel, Jay J.
    Barnett, Steven L.
    Gorab, Robert S.
    Nassif, Nader A.
    JOURNAL OF ARTHROPLASTY, 2023, 38 (06) : 998 - 1003
  • [22] Finite element analysis of the mechanical behavior of the different angle hip femoral stem
    Guvercin, Yilmaz
    Yaylaci, Murat
    Olmez, Hasan
    Yaylaci, Ecren Uzun
    Ozdemir, Mehmet Emin
    Dizdar, Ayberk
    BIOMATERIALS AND BIOMECHANICS IN BIOENGINEERING, 2022, 6 (01): : 29 - 46
  • [23] Biomechanical consideration of total hip arthroplasty following failed fixation of femoral intertrochanteric fractures - A finite element analysis
    Chen, Dave W.
    Lin, Chun-Li
    Hu, Chih-Chien
    Tsai, Ming-Feng
    Lee, Mel S.
    MEDICAL ENGINEERING & PHYSICS, 2013, 35 (05) : 569 - 575
  • [24] Conversion total hip arthroplasty after previous intertrochanteric hip fractures: a review of the epidemiology, technical issues, and outcomes
    Rinehart, Dustin B.
    Fene, Evan S.
    Huo, Michael H.
    CURRENT ORTHOPAEDIC PRACTICE, 2022, 33 (05): : 409 - 413
  • [25] Stress shielding effect after total hip arthroplasty varies between combinations of stem design and stiffness-a comparing biomechanical finite element analysis
    Burchard, Rene
    Graw, Jan A. A.
    Soost, Christian
    Schmitt, Jan
    INTERNATIONAL ORTHOPAEDICS, 2023, 47 (08) : 1981 - 1987
  • [26] The biomechanical effect of the collar of a femoral stem on total hip arthroplasty
    Jeon, Insu
    Bae, Ji-Yong
    Park, Jin-Hong
    Yoon, Taek-Rim
    Todo, Mitsugu
    Mawatari, Masaaki
    Hotokebuchi, Takao
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2011, 14 (01) : 103 - 112
  • [27] Quantification of soft tissue balance in total knee arthroplasty using finite element analysis
    Oh, Kwang-Jun
    Park, Won Man
    Kim, Kyungsoo
    Kim, Yoon Hyuk
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2014, 17 (14) : 1630 - 1634
  • [28] Conversion of a Failed Hip Hemiarthroplasty to Total Hip Arthroplasty: A Systematic Review and Meta-Analysis
    Poursalehian, Mohammad
    Hassanzadeh, Ali
    Lot, Mohadeseh
    Mortazavi, Seyed Mohammad Javad
    ARTHROPLASTY TODAY, 2024, 28
  • [29] Primary hip arthroscopy and conversion to total hip arthroplasty: trends and survival analysis in the Medicare population
    Malik, Azeem T.
    Jain, Nikhil
    Scharschmidt, Thomas J.
    Glassman, Andrew H.
    Khan, Safdar N.
    HIP INTERNATIONAL, 2022, 32 (02) : 239 - 245
  • [30] Use of Hip Arthroscopy and Risk of Conversion to Total Hip Arthroplasty: A Population-Based Analysis
    Schairer, William W.
    Nwachukwu, Benedict U.
    McCormick, Frank
    Lyman, Stephen
    Mayman, David
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2016, 32 (04) : 587 - 593