Study of a gel dosimeter based on Ag nanoparticles for applications in radiation therapy with synchrotron X-rays at ultrahigh dose rate compared to 60Co γ-rays

被引:0
|
作者
Soliman, Yasser S. [1 ]
Capron, Marie [3 ]
Pontoni, Diego [3 ]
Krisch, Michael [2 ]
Pellicioli, Paolo [2 ,4 ]
机构
[1] Egyptian Atom Energy Author AEA, Natl Ctr Radiat Res & Technol NCRRT, POB 8029, Cairo 11787, Egypt
[2] ESRF European Synchrotron, ID17 Biomed Beamline,71 Ave Martyrs, F-38000 Grenoble, France
[3] ESRF European Synchrotron, Partnership Soft Condensed Matter PSCM, 71 Ave Martyrs, F-38000 Grenoble, France
[4] Univ Bern, Inst Anat, Baltzerstr 2, CH-3012 Bern, Switzerland
关键词
Gel dosimeter; Ag nanoparticles; Dosimetry; X-ray synchrotron radiation; Radiotherapy; SILVER NANOPARTICLES; QUALITY-ASSURANCE; GOLD NANOPARTICLES; OPTICAL-PROPERTIES; ALANINE DOSIMETER; MEDICAL BEAMLINE; FILM DOSIMETRY; MICROBEAM; RADIOTHERAPY; GELATIN;
D O I
10.1016/j.radphyschem.2024.112351
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
A silver nitrate (AgNO3) gel was developed and evaluated as a dosimeter for synchrotron X-ray-based studies in microbeam radiation therapy (MRT) and FLASH radiation therapy. The gel was irradiated at the European Synchrotron Radiation Facility (ID17 Biomedical beamline) with a continuous X-ray spectrum in the 50-600 key range and a dose rate of 11.6 kGy/s. The spectrophotometric response after irradiation at this beamline was compared with the response at a conventional 60Co gamma-ray source providing a dose rate of 0.27 Gy/s. Ag + ions in the gel dosimeter undergo reduction to Ago nanoparticles, as detected at 450 nm, which is a surface plasmon resonance band. The intensity of this band increased linearly with increasing absorbed dose up to 100 Gy, and improved with the addition of glycerol. The gel dosimeter exhibited the lowest detectable dose (LDD) of 75.9% lower and a dose deposition of 76% higher for ID17 beamline irradiation than for 60Co irradiation. The theoretical relative response HQ,Qo for the gel irradiated at the synchrotron was 1.70, which is in close agreement with the experimental relative response FQ,Qo = 1.76. These results confirm the dose enhancement in the gel irradiated with orthovoltage X-rays. The good properties of the silver nitrate gel, together with its increased sensitivity to orthovoltage X-ray irradiation and modest overall uncertainty of 5.8% (2 sigma), confirm that the gel is a valid dosimeter for measurements at ultrahigh dose rates for synchrotron radiotherapy. The dosimeter is also a promising candidate for dose enhancement factor measurements in nanoparticle-based radiotherapy techniques.
引用
收藏
页数:11
相关论文
共 8 条
  • [1] Comparison of the dosimetric response of two Sr salts irradiated with 60Co?-rays and synchrotron X-rays at ultra-high dose rate
    Soliman, Yasser S.
    Fahim, Ramy Amer
    Pellicioli, Paolo
    Krisch, Michael
    Beshir, W. B.
    Abdel-Fattah, A. A.
    El-Moghny, Muhammad G. Abd
    El-Deab, Mohamed S.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 208
  • [2] Characterization of ferrous-agarose-xylenol gel dosimeter at 60Co γ-rays beam therapy unit
    Moussous, Ouiza
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2021, 14 (01) : 105 - 112
  • [3] Characterization of ferrous-agarose-xylenol gel dosimeter at 60Co γ-rays beam therapy unit
    Ouiza Moussous
    Radiological Physics and Technology, 2021, 14 : 105 - 112
  • [4] Radiation dose rate affects the radiosensitization of MCF-7 and HeLa cell lines to X-rays induced by dextran-coated iron oxide nanoparticles
    Khoshgard, Karim
    Kiani, Parvaneh
    Haghparast, Abbas
    Hosseinzadeh, Leila
    Eivazi, Mohammad Taghi
    INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2017, 93 (08) : 757 - 763
  • [5] A study on radiation interactions, dose enhancement, and hydrolysis with metallic nanoparticles irradiated by 6 megavoltage X-rays: Geant4 Monte Carlo simulation
    Batooei, Sepehr
    Moslehi, Amir
    Islamian, Jalil Pirayesh
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2022, 526 : 19 - 28
  • [6] Photon activated therapy (PAT) using monochromatic Synchrotron x-rays and iron oxide nanoparticles in a mouse tumor model: feasibility study of PAT for the treatment of superficial malignancy
    Choi, Gi-Hwan
    Seo, Seung-Jun
    Kim, Ki-Hong
    Kim, Hong-Tae
    Park, Sung-Hwan
    Lim, Jae-Hong
    Kim, Jong-Ki
    RADIATION ONCOLOGY, 2012, 7
  • [7] Dose-average linear energy transfer of electrons released in liquid water and LiF:Mg,Ti by low-energy x-rays, 137Cs and 60Co gamma
    Massillon-JL, G.
    Cabrera-Santiago, A.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (03)
  • [8] Dosimetric comparison of 192Ir high-dose-rate brachytherapy vs. 50 kV x-rays as techniques for breast intraoperative radiation therapy: Conceptual development of image-guided intraoperative brachytherapy using a multilumen balloon applicator and in-room CT imaging
    Jones, Ryan
    Libby, Bruce
    Showalter, Shayna L.
    Brenin, David R.
    Wilson, David D.
    Schroen, Anneke
    Morris, Monica
    Reardon, Kelli A.
    Morrison, John
    Showalter, Timothy N.
    BRACHYTHERAPY, 2014, 13 (05) : 502 - 507