Performance Evaluation of New Large Area 3D CdZnTe Drift Strip Detectors

被引:1
作者
Hauptmann, Benjamin N. [1 ]
Owe, Selina R. H. [1 ]
Kuvvetli, Irfan [1 ]
Tcherniak, Denis [1 ]
Budtz-Jorgensen, Carl [1 ]
Bairstow, Ian [2 ]
Harris, Brian [3 ]
Cherlin, Alexander [2 ]
Cantwell, Ben [2 ]
机构
[1] Tech Univ Denmark, DTU Space, Elektrovej 328, Lyngby, Denmark
[2] Kromek Uk, Netpk,Thomas Wright Way, Sedgefield TS21 3FD, England
[3] Kromek US, 143 Zehner Sch Rd, Zelienople, PA 16063 USA
来源
SPACE TELESCOPES AND INSTRUMENTATION 2024: ULTRAVIOLET TO GAMMA RAY, PT 1 | 2024年 / 13093卷
关键词
CZT drift strip detector; Digitized pulse shape analysis; 3D CZT detectors; Gamma detectors; X-ray detectors and telescopes; Space instrumentation; Medical imaging; Molecular Breast Imaging; MOBILITY;
D O I
10.1117/12.3015044
中图分类号
P1 [天文学];
学科分类号
0704 ;
摘要
What do medical imaging and astronomy have in common? Whether it is detecting tumors or observing colliding neutron stars, both fields can benefit from the same sensor technology. When observing radiation in the MeV-band (0.1-100 MeV) from celestial sources, one faces challenges like low flux, limited interaction probability, three energy loss processes, and a high background radiation rate. Present MeV-band observatories suffer from poor sensitivity, and new state-of the-art detector technology will be a key contributor to improve sensitivity of future observatories. The detector group at DTU Space has developed a 3D CdZnTe drift strip detector technology to meet the demands of future high-energy detectors. This advancement also shows potential application for emerging Low Dose Molecular Breast Imaging (LD-MBI) systems for breast cancer diagnosis. In this study, we present the results of the novel large area 3D CZT drift strip detectors (4x4x0.5 cm(3)) developed in a collaboration between DTU Space and Kromek. We find the spatial resolution of these modules to be < 0.6mm in the x- and z-axis, and we measure the spectral resolutions to be < 7% FWHM at 122 keV, < 1.4% FWHM at 661.6 keV, and 1% FWHM above 1460 keV.
引用
收藏
页数:19
相关论文
共 13 条
[1]  
Cherlin A., 2021, 2021 IEEE NUCL SCI S, P1
[2]  
Erlandsson K., 2021, 16 INT M FULL 3D IM
[3]  
Hauptmann B. N., 2024, Characterization and performance evaluation of the 3d mbi detector modules
[4]   Direct measurement of product of the electron mobility and mean free drift time of CdZnTe semiconductors using position sensitive single polarity charge sensing detectors [J].
He, Z ;
Knoll, GF ;
Wehe, DK .
JOURNAL OF APPLIED PHYSICS, 1998, 84 (10) :5566-5569
[5]  
Hutton B. F., 2021, IEEE NSS MIC
[6]  
Owe S., 2023, Three-Dimensional Mapping of Carrier Lifetime and Mo- bility, P87
[7]  
Owe S., 2023, Development of 3D Imaging Detectors for High Energy Astronomy Instrumentation
[8]  
Owe S. H., 2018, Characterization of new generation 3d czt high resolution xand gamma-ray detector for high energy space mission concept
[9]   Carrier Lifetime and Mobility Characterization Using the DTU 3-D CZT Drift Strip Detector [J].
Owe, S. Howalt ;
Kuvvetli, I ;
Budtz-Jorgensen, C. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2021, 68 (09) :2440-2446
[10]   Evaluation of a Compton camera concept using the 3D CdZnTe drift strip detectors [J].
Owe, S. Howalt ;
Kuvvetli, I. ;
Budtz-Jorgensen, C. ;
Zoglauer, A. .
JOURNAL OF INSTRUMENTATION, 2019, 14