Additional Screw Added to the Femoral Neck System Could Enhance the Stability of Pauwel Type III Femoral Neck Fractures: a Finite Element Analysis

被引:0
作者
Cha, Yonghan [1 ]
Park, Sunghoon [2 ]
Jung, Chang-Ho [3 ]
Kim, Jin-Woo [4 ]
Yoo, Jun-Il [5 ]
Kim, Jung-Taek [6 ]
Jeon, Yongho
Han, Kyeong Jin [6 ]
机构
[1] Eulji Univ, Daejeon Eulji Med Ctr, Dept Orthoped Surg, Sch Med, Daejeon, South Korea
[2] Ajou Univ, Med Ctr, Sch Med, Dept Radiol, Suwon 16499, South Korea
[3] Ajou Univ, Dept Mech Engn, Suwon, South Korea
[4] Eulji Univ, Nowon Eulji Med Ctr, Dept Orthoped Surg, Sch Med, Seoul, South Korea
[5] Inha Univ, Inha Univ Hosp, Sch Med, Dept Orthoped Surg, Incheon, South Korea
[6] Ajou Univ, Med Ctr, Sch Med, Dept Orthoped Surg, 164 World Cup Ro, Suwon 16499, South Korea
基金
新加坡国家研究基金会;
关键词
Femoral neck fractures; Finite element analysis; Biomechanical phenomena; Orthopedic fixation devices; Fracture fixation; TROCHANTERIC FRACTURES; HIP-JOINT; POSITION; FAILURE; MODELS;
D O I
10.4055/cios24194
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: This study explores effective fixation methods for Pauwel type III femoral neck fractures by evaluating the biomechanical benefits of adding a screw to the Femoral Neck System (FNS). Methods: Computed tomography (CT) scans of an 82-year-old female patient with an intertrochanteric fracture were used to establish a finite element femur model with heterogeneous material properties. Finite element models of Pauwel type III fractures were created with and without an additional screw. The central and inferior trajectories of the FNS bolt were examined separately and combined with an additional screw for virtual fixation. Walking and stair-climbing loads were applied. Results: With the addition of a screw, both peak maximum and minimum principal strains consistently stayed comparable or decreased in models with both central and inferior bolt trajectories, while the volume of elements with principal strain exceeding 1% decreased by more than half. The peak von Mises stress observed in the implants ranged from 215.7 to 359.3 MPa, remaining below the titanium alloy's yield strength of 800 MPa. For normal walking, the addition of a screw to the central bolt trajectory model decreased the fracture gap by 50.6% and reduced sliding distance by 8.6%. For the inferior bolt trajectory, the gap was reduced by 57.9% and sliding distance by 25.0%. Under stair-climbing conditions, these improvements were also evident; the central trajectory model saw a halved fracture gap and a 7.9% decrease in sliding distance, while the inferior trajectory model experienced a 55.7% gap reduction and a 27.2% decrease in sliding distance. The additional screw increased the area ratio of the fracture site experiencing interfragmentary compression 34%-39%, while the additional screw alleviated peak interfragmentary compression by 12%-18% under both normal walking and stair-climbing conditions. Conclusions: The addition of a screw reduced the fracture gap, sliding distance, and peak interfragmentary compression, while increasing the area ratio of interfragmentary compression under both walking and stair-climbing loads, regardless of the FNS bolt trajectory, suggesting a better mechanical environment for fracture healing.
引用
收藏
页码:204 / 215
页数:12
相关论文
共 30 条
  • [1] ASTM International, 2016, Standard specification for wrought titanium-6aluminum-7niobium alloy for surgical implant applications (UNS R56700) Internet
  • [2] HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS
    BERGMANN, G
    GRAICHEN, F
    ROHLMANN, A
    [J]. JOURNAL OF BIOMECHANICS, 1993, 26 (08) : 969 - 990
  • [3] Bonnaire FA, 2002, INJURY, V33, pS24
  • [4] Biomechanical Evaluation of Internal Fixation of Pauwels Type III Femoral Neck Fractures: A Systematic Review of Various Fixation Methods
    Cha, Yong-Han
    Yoo, Jun-Il
    Hwang, Seok-Young
    Kim, Kap-Jung
    Kim, Ha-Yong
    Choy, Won-Sik
    Hwang, Sun-Chul
    [J]. CLINICS IN ORTHOPEDIC SURGERY, 2019, 11 (01) : 1 - 14
  • [5] Pre-sliding of femoral neck system improves fixation stability in pauwels type III femoral neck fracture: a finite element analysis
    Cha, Yonghan
    Chung, Jun Young
    Jung, Chang-Ho
    Kim, Jin-Woo
    Lee, Jeyoon
    Yoo, Jun-Il
    Kim, Jung-Taek
    Jeon, Yongho
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
  • [6] Improved control over implant anchorage under the use of the femoral neck system for fixation of femoral neck fractures: a technical note
    Cha, Yonghan
    Song, Ji-Ung
    Yoo, Jun-Il
    Park, Ki Hoon
    Kim, Jung-Taek
    Park, Chan Ho
    Choy, Won-Sik
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2021, 22 (01)
  • [7] The efficacy of dynamic compression locking system vs. dynamic hip screw in the treatment of femoral neck fractures: a comparative study
    Chang, Jian-Zhong
    Xiao, Ya-Ping
    Li, Ling
    Bei, Ming-Jian
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [8] Femoral head necrosis: A finite element analysis of common and novel surgical techniques
    Cilla, Myriam
    Checa, Sara
    Preininger, Bernd
    Winkler, Tobias
    Perka, Carsten
    Duda, Georg N.
    Pumberger, Matthias
    [J]. CLINICAL BIOMECHANICS, 2017, 48 : 49 - 56
  • [9] DePuy Synthes, 2024, Surgical technique of Femoral Neck System Internet
  • [10] Management of proximal femur fractures in the elderly: current concepts and treatment options
    Fischer, H.
    Maleitzke, T.
    Eder, C.
    Ahmad, S.
    Stoeckle, U.
    Braun, K. F.
    [J]. EUROPEAN JOURNAL OF MEDICAL RESEARCH, 2021, 26 (01)