In-line spectroscopy for iodine quantification in dynamic contrast-enhanced dedicated breast CT

被引:1
作者
Goris, Liselot C. [1 ,2 ]
Mikerov, Mikhail [2 ]
Pautasso, Juan J. [2 ]
Sechopoulos, Ioannis [1 ,2 ,3 ]
机构
[1] Univ Twente, Tech Med Ctr, Enschede, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Med Imaging, Geert Grootepl 10,Post 766, NL-6525 GA Nijmegen, Netherlands
[3] Dutch Expert Ctr Screening LRCB, Nijmegen, Netherlands
基金
欧洲研究理事会;
关键词
4D DCE-BCT; breast cancer; breast phantom; iodine quantification; spectroscopy; SPECTROPHOTOMETRIC DETERMINATION; PHANTOM; MAMMOGRAPHY;
D O I
10.1002/mp.17497
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundIn 4D dynamic contrast-enhanced dedicated breast computed tomography (4D DCE-bCT), the functional properties of the breast will be characterized by monitoring the uptake and washout of iodine-based contrast agents over time. This information could be valuable in breast cancer treatment. However, prior to clinical implementation, it is crucial to validate the quantitative estimates of iodine concentrations at each time point during acquisition.PurposeTo develop an in-line spectroscopy system capable of measuring iodine concentrations in a dynamic x-ray breast phantom in real-time.MethodsPotassium iodide served as the contrast agent. The system was set-up at both the entrance and exit of the phantom. It comprises a fiber-coupled green LED and collimator, which together ensure that a parallel beam passes through the sample holder. Transmitted light is captured by a collimator on the opposite side and directed through a fiber optic cable to a photodetector for intensity measurement. The relationship between 13 iodine concentrations (0-6 mg I/mL) and light transmission was tested, and the system's repeatability and accuracy were determined.ResultsThe system exhibited a strong correlation between iodine concentration and transmission values, achieving a root-mean-square error of 0.007. The repeated measurements had relative standard deviations of 0.04% and 0.1% for repeated water measurements at the phantom's entrance and exit, respectively. Furthermore, the accuracy measurements gave a mean error of fitting of 0.008 (+/- 0.07) mg I/mL.ConclusionThe in-line spectroscopy system can effectively monitor iodine concentrations in a dynamic breast phantom, providing a reliable method for quantitative validation of 4D DCE-bCT.
引用
收藏
页码:1037 / 1044
页数:8
相关论文
共 31 条
[1]   Spectrophotometric determination of urinary iodine by the Sandell-Kolthoff reaction subsequent to dry alkaline ashing.: Results from the Czech Republic in the period 1994-2002 [J].
Bílek, R ;
Bednár, J ;
Zamrazil, V .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2005, 43 (06) :573-580
[2]  
Born M.A. X., 1980, Principles of Optics, VSixth, P109, DOI 10.1016/B978-0-08-026482-0.50010-4
[3]   Patient-based 4D digital breast phantom for perfusion contrast-enhanced breast CT imaging [J].
Caballo, Marco ;
Mann, Ritse ;
Sechopoulos, Ioannis .
MEDICAL PHYSICS, 2018, 45 (10) :4448-4460
[4]   Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly undersampled projection data sets [J].
Chen, Guang-Hong ;
Tang, Jie ;
Leng, Shuai .
MEDICAL PHYSICS, 2008, 35 (02) :660-663
[5]   The diagnostic performance of dual-layer spectral detector CT for distinguishing breast cancer biomarker expression and molecular subtypes [J].
Chen, Lanjing ;
Xiao, Zhengyuan ;
Fu, Jianmei ;
Huang, Jingrong ;
Lan, Yongshu .
SCIENTIFIC REPORTS, 2024, 14 (01)
[6]   SCATTERING OF LIGHT BY AN AIR BUBBLE IN WATER [J].
DAVIS, GE .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1955, 45 (07) :572-581
[7]   An iterative maximum-likelihood polychromatic algorithm for CT [J].
De Man, B ;
Nuyts, J ;
Dupont, P ;
Marchal, G ;
Suetens, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (10) :999-1008
[8]   2 SIMPLE METHODS FOR MEASURING IODINE IN URINE [J].
DUNN, JT ;
CRUTCHFIELD, HE ;
GUTEKUNST, R ;
DUNN, AD .
THYROID, 1993, 3 (02) :119-123
[9]   Catalytic spectrophotometric determination of iodide in pharmaceutical preparations and edible salt [J].
El-Ries, M. A. ;
Khaled, Elmorsy ;
Zidane, F. I. ;
Ibrahim, S. A. ;
Abd-Elmonem, M. S. .
DRUG TESTING AND ANALYSIS, 2012, 4 (02) :129-135
[10]  
George M., 2011, J ANAL CHEM+, V6, P129