Commissioning evaluation of a deviceless 4DCT scanner

被引:0
作者
Tillery, Hunter [1 ]
Windsor, Cheyann [1 ]
Aguilera, Christopher [2 ]
机构
[1] Northwest Med Phys Ctr, 21031 67th Ave West, Lynnwood, WA 98036 USA
[2] Oregon Hlth & Sci Univ, Dept Radiat Med, Portland, OR USA
关键词
CT Simulation; Deviceless; 4D; D4D; 4DCT;
D O I
10.1002/acm2.70048
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundThe utilization of four-dimensional computed tomography (4DCT) for radiation therapy has not seen major advances to the method of data binning since shortly after inception. Recently there is increased interest in the utilization of an alternative binning method rather than more established techniques. At this point routine quality assurance and commissioning of 4DCT have been well studied and established with traditional binning methods. Due to this new "deviceless" technique relying on algorithms instead of an external breathing signal, established dynamic phantoms and equipment typically used in the commissioning and quality assurance workflow have proven to no longer be compatible.MethodsA commercially available phantom was modified to include components that the deviceless 4D algorithm uses for binning. Typical 4DCT commissioning datasets were acquired and reconstructed using both deviceless and device-based binning techniques. Both regular and irregular breathing curves were evaluated for performance, similar to what would be seen with typical radiation therapy patients.ResultsDeviceless and device-based binning methods performed similarly and well for regular breathing curves. As datasets became more irregular, the deviceless algorithm was better able to reconstruct 4DCTs.ConclusionCommissioning datasets for both device-based and deviceless 4DCTs were evaluated to test if modifications to a commercially available phantom would allow for an accurate comparison between binning systems. It was shown that not only did these modifications work but also highlighted a difference in the way that these systems binned data, which could be applied to patients with breathing irregularities.
引用
收藏
页数:7
相关论文
共 10 条
[1]   Validation of four-dimensional computed tomography without external reference respiratory signals for radiation treatment planning of lung tumors [J].
Doi, Yasuhiro ;
Shimohigashi, Yoshinobu ;
Kai, Yudai ;
Maruyama, Masato ;
Toya, Ryo .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (05)
[2]   Comparison of two different respiratory monitoring systems with 4D-CT images for target volume definition in patients undergoing para-aortic nodal irradiation [J].
Ergen, Sefika Arzu ;
Karacam, Songul ;
Catal, Tuba Kurt ;
Dincbas, Fazilet Oner ;
Oksuz, Didem Colpan ;
Sahinler, Ismet .
NORTHERN CLINICS OF ISTANBUL, 2024, 11 (02) :120-126
[3]   Evaluation of target volume and dose accuracy in intrafractional cases of lung cancer based on 4D-CT and 4D-CBCT images using an in-house dynamic thorax phantom [J].
Haqqi, Firyal Dhiyaul ;
Sudarmaji, Arief ;
Wibowo, Wahyu Edy ;
Nasution, Nuruddin ;
Handoko, Handoko ;
Fauzia, Annisa Rahma ;
Qudsi, Jihadil ;
Pawiro, Supriyanto Ardjo .
REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (02) :360-370
[4]   4D CT sorting based on patient internal anatomy [J].
Li, Ruijiang ;
Lewis, John H. ;
Cervino, Laura I. ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (15) :4821-4833
[5]   Evaluation of a New GE Device-Less Cine 4D-CT [J].
Martin, R. ;
Chandler, A. ;
Doan, D. ;
Rowland, C. ;
Pan, T. .
MEDICAL PHYSICS, 2015, 42 (06) :3645-3645
[6]   A comprehensive quality assurance procedure for 4D CT commissioning and periodic QA [J].
Polizzi, Mitchell ;
Kim, Siyong ;
Rosu-Bubulac, Mihaela .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (11)
[7]  
Sprouts D, 2017, MED PHYS, V44, P3046
[8]   Quality assurance of a breathing controlled four-dimensional computed tomography algorithm [J].
Szkitsak, Juliane ;
Karius, Andre ;
Hofmann, Christian ;
Fietkau, Rainer ;
Bert, Christoph ;
Speer, Stefan .
PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 23 :85-91
[9]   4DCT is long overdue for improvement [J].
Tryggestad, Erik ;
Li, Heng ;
Rong, Yi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (04)
[10]   Intelligent 4D CT sequence scanning (i4DCT): First scanner prototype implementation and phantom measurements of automated breathing signal-guided 4D CT [J].
Werner, Rene ;
Sentker, Thilo ;
Madesta, Frederic ;
Schwarz, Annette ;
Vornehm, Marc ;
Gauer, Tobias ;
Hofmann, Christian .
MEDICAL PHYSICS, 2020, 47 (06) :2408-2412