Effect of different airway devices on ventilation during cardiopulmonary resuscitation

被引:0
作者
Segond, N. [1 ,2 ,3 ]
Fischer, M. [1 ,2 ]
Fontecave-Jallon, J. [3 ]
Podsiadlo, P. [4 ]
Lurie, K. [5 ,6 ]
Bellier, A. [7 ]
Debaty, G. [1 ,2 ,3 ]
机构
[1] Univ Hosp Grenoble Alpes, Emergency Dept, Grenoble, France
[2] Univ Hosp Grenoble Alpes, Mobile Intens Care Unit, Grenoble, France
[3] Univ Grenoble Alpes, CNRS, UMR 5525, VetAgro Sup,Grenoble INP,TIMC, F-38000 Grenoble, France
[4] Jan Kochanowski Univ, Dept Emergency Med, Kielce, Poland
[5] Hennepin Healthcare, Minneapolis, MN USA
[6] Univ Minnesota, Minneapolis, MN USA
[7] Univ Grenoble Alpes, Dept Anat LADAF, Grenoble, France
关键词
Cardiopulmonary resuscitation; Heart arrest; Mechanical ventilation; Respiratory physiology; Respiratory mechanics; HOSPITAL CARDIAC-ARREST; CEREBRAL PERFUSION-PRESSURE; BAG-MASK VENTILATION; ENDOTRACHEAL INTUBATION; MECHANICAL VENTILATION; LARYNGEAL TUBE; HEAD; GUIDELINES; COUNCIL; STRATEGY;
D O I
10.1016/j.resuscitation.2025.110584
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Purpose: This study compared face mask, supraglottic airway device (SGA), and endotracheal tube (ETT) ventilation with mechanical ventilation (MV) during cardiopulmonary resuscitation (CPR) in the flat position and with head and thorax elevation (HTE). Methods: Using thawed, fresh-frozen human cadavers this randomized cross-over study compared face mask, SGA, and ETT ventilation using an automated ventilator in the flat and HTE positions. Tidal volume (TV) was set to 8 mL/kg ideal predicted body weight, and expiratory TV (VTe) (mL/ kg) was the primary endpoint. Secondary endpoints included inspiratory TV (VTi), maximal inspiratory airway pressure (Pmax), and leakage between inspiratory and expiratory tidal volumes (VTi-VTe). Results: Data from 8 cadavers and 2302 ventilation cycles were analyzed. In the flat position, VTe was 7.66 +/- 3.75 with ETT, 5.01 +/- 3.14 with SGA, and 5.63 +/- 2.83 with face mask, respectively. A mixed linear model showed the airway device significantly impacted VTe, VTi, Pmax, and VTi-VTe (p < 0.001 for each). Compared with intubation, both face mask and SGA ventilation resulted in lower VTe, lower Pmax, and higher VTi-VTe (p < 0.001 for each). No significant differences were observed between face mask and SGA ventilation. There were higher VTe and lower VTiVTe values (p < 0.001 for each) with HTE versus the flat position (p < 0.001). Conclusions: In human cadavers undergoing CPR, mechanical ventilation through a face mask or SGA versus an ETT was associated with lower VTe, lower Pmax, and higher leakage values in human cadavers during CPR. Head and thorax elevation reduce face mask and SGA airway leakage during CPR and increase VTe.
引用
收藏
页数:8
相关论文
共 36 条
  • [1] Aufderheide Tom P, 2004, Crit Care Med, V32, pS345, DOI 10.1097/01.CCM.0000134335.46859.09
  • [2] Effect of a Strategy of a Supraglottic Airway Device vs Tracheal Intubation During Out-of-Hospital Cardiac Arrest on Functional Outcome The AIRWAYS-2 Randomized Clinical Trial
    Benger, Jonathan R.
    Kirby, Kim
    Black, Sarah
    Brett, Stephen J.
    Clout, Madeleine
    Lazaroo, Michelle J.
    Nolan, Jerry P.
    Reeves, Barnaby C.
    Robinson, Maria
    Scott, Lauren J.
    Smartt, Helena
    South, Adrian
    Stokes, Elizabeth A.
    Taylor, Jodi
    Thomas, Matthew
    Voss, Sarah
    Wordsworth, Sarah
    Rogers, Chris A.
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2018, 320 (08): : 779 - 791
  • [3] The impact of introducing real time feedback on ventilation rate and tidal volume by ambulance clinicians in the North East in cardiac arrest simulations
    Charlton, Karl
    McClelland, Graham
    Millican, Karen
    Haworth, Daniel
    Aitken-Fell, Paul
    Norton, Michael
    [J]. RESUSCITATION PLUS, 2021, 6
  • [4] EFFECT OF BODY POSITION ON PULMONARY-FUNCTION
    DEAN, E
    [J]. PHYSICAL THERAPY, 1985, 65 (05): : 613 - 618
  • [5] The impact of real-time feedback on ventilation quality during out-of-hospital cardiac arrest: A before-and-after study
    Drennan, Ian R.
    Lee, Meji
    Heroux, Jean-Philippe
    Lee, Andrew
    Riches, John
    Peppler, Jonathan
    Poitras, Annabel
    Cheskes, Sheldon
    [J]. RESUSCITATION, 2024, 204
  • [6] Duhem H, 2019, Cadaver models for cardiac arrest: A systematic review and perspectives, P9
  • [7] Effect of automated head-thorax elevation during chest compressions on lung ventilation: a model study
    Duhem H.
    Terzi N.
    Segond N.
    Bellier A.
    Sanchez C.
    Louis B.
    Debaty G.
    Guérin C.
    [J]. Scientific Reports, 13 (1)
  • [8] Improving post-cardiac arrest cerebral perfusion pressure by elevating the head and thorax
    Duhem, Helene
    Moore, Johanna C.
    Rojas-Salvador, Carolina
    Salverda, Bayert
    Lick, Michael
    Pepe, Paul
    Labarere, Jose
    Debaty, Guillaume
    Lurie, Keith G.
    [J]. RESUSCITATION, 2021, 159 : 45 - 53
  • [9] Manual ventilation quality is improved with a real-time visual feedback system during simulated resuscitation
    Gould, Jeffrey R.
    Campana, Lisa
    Rabickow, Danielle
    Raymond, Richard
    Partridge, Robert
    [J]. INTERNATIONAL JOURNAL OF EMERGENCY MEDICINE, 2020, 13 (01)
  • [10] European Resuscitation Council Guidelines 2021 Epidemiology of cardiac arrest in Europe
    Graesner, Jan-Thorsten
    Herlitz, Johan
    Tjelmeland, Ingvild B. M.
    Wnent, Jan
    Masterson, Siobhan
    Lilja, Gisela
    Bein, Berthold
    Boettiger, Bernd W.
    Rosell-Ortiz, Fernando
    Nolan, Jerry P.
    Bossaert, Leo
    Perkins, Gavin D.
    [J]. RESUSCITATION, 2021, 161 : 61 - 79