Helmet continuous positive airway pressure for patients' transport using a single oxygen cylinder: A bench study

被引:1
|
作者
Capsoni, N. [1 ,2 ]
Zadek, F. [2 ]
Privitera, D. [1 ]
Parravicini, G. [1 ]
Zoccali, G. V. [3 ]
Galbiati, F. [1 ]
Bombelli, M. [2 ]
Fumagalli, R. [2 ,3 ]
Langer, T. [2 ,3 ]
机构
[1] ASST Grande Osped Metropolitano Niguarda, Dept Emergency Med, Piazza Osped Maggiore 3, I-20162 Milan, Italy
[2] Univ Milano Bicocca, Dept Med & Surg, Monza, Italy
[3] Niguarda Ca Granda, Dept Anesthesia & Intens Care Med, Milan, Italy
来源
PULMONOLOGY | 2025年 / 31卷 / 01期
关键词
Continuous positive airway pressure; Non-invasive ventilation; Hypoxemic respiratory failure; Emergency medical services; Transportation of patients; Ambulances; ACUTE RESPIRATORY-FAILURE; PREHOSPITAL USE; VENTILATION;
D O I
10.1016/j.pulmoe.2023.09.007
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Continuous positive airway pressure (CPAP) is frequently used to treat patients with acute respiratory failure in out-of-hospital settings. Compared to a facemask, the helmet has many advantages for the patient but requires a minimum gas flow of 60 L/min to avoid CO2 rebreathing. The aim of the present bench study was to evaluate the performance of four Venturi devices, connected to a single oxygen cylinder, in delivering helmet-CPAP with clinically relevant gas flow, fraction of inspired oxygen (FiO(2))(,) and positive end-expiratory pressure (PEEP) values. Methods: Three double-inlet Venturi systems (EasyVent, Ventuplus, Compact-HAR) were connected to full 5-L oxygen cylinders using a double flowmeter, and their oxygen requirements to reach different setups (flow 60-80 L/min; FiO(2) 0.4-0.5-0.6, PEEP 7.5-10-12.5 cmH(2)O) were tested. The fourth Venturi system (O2-MAX) was directly attached to the tank, and the flow and FiO(2) delivered at preset FiO(2) 0.3 and 0.6 were recorded. The runtime of the cylinder was assessed. Results: EasyVent, Ventuplus, and O2-MAX were able to deliver helmet-CPAP with clinically useful setups when connected to a single oxygen cylinder, while Compact-HAR did not. The runtime of the cylinders ranged between 28 and 60 minutes according to the preset flow and FiO(2). The delivered gas flow decreased slowly and linearly with the drop in cylinder pressure until its exhaustion. Conclusions: Helmet-CPAP might be provided using portable Venturi systems connected to an oxygen cylinder, but not all of them are able to deliver it. The use of a double flowmeter allows delivery of both high flow and high FiO(2) when double-inlet Venturi systems are used. Due to the flow drop observed during the cylinder consumption, a flow >60 L/min should be set when helmet-CPAP is started. Considering the flow drop phenomenon, the estimated duration of the tank runtime can be used with a margin of safety when planning patient transport.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Head helmet versus face mask for non-invasive continuous positive airway pressure: a physiological study
    Nicolò Patroniti
    Giuseppe Foti
    Annamaria Manfio
    Anna Coppo
    Giacomo Bellani
    Antonio Pesenti
    Intensive Care Medicine, 2003, 29 : 1680 - 1687
  • [42] Air-oxygen helmet delivered continuous positive airway pressure to manage respiratory failure due to bronchiolitis: authors' response
    Kapadia, J.
    Brazier, A.
    Stone, S.
    Farrer, K.
    ACTA PAEDIATRICA, 2012, 101 (07) : E266 - E267
  • [43] Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices
    Rigau, Jordi
    Montserrat, Josep M.
    Woehrle, Holger
    Plattner, Diana
    Schwaibold, Matthias
    Navajas, Daniel
    Farre, Ramon
    CHEST, 2006, 130 (02) : 350 - 361
  • [44] Randomized clinical trial comparing helmet continuous positive airway pressure (hCPAP) to facemask continuous positive airway pressure (fCPAP) for the treatment of acute respiratory failure in the emergency department
    Adi, Osman
    Via, Gabriele
    Salleh, Siti Hafsah
    Chuan, Tan Wan
    Ab Rahman, Jamalludin
    Muhammad, Nik Azlan Nik
    Atan, Rafidah
    Anaes, M.
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2021, 49 : 385 - 392
  • [45] Continuous positive airway pressure in patients with heart failure
    Tulaimat, A
    Mokhlesi, B
    Stevens, D
    NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (01): : 93 - 94
  • [46] Noninvasive Continuous Positive Airway Pressure in Acute Respiratory Failure: Helmet Versus Facial Mask
    Chidini, Giovanna
    Calderini, Edoardo
    Cesana, Bruno Mario
    Gandini, Cristiano
    Prandi, Edi
    Pelosi, Paolo
    PEDIATRICS, 2010, 126 (02) : E330 - E336
  • [47] Continuous positive airway pressure by "helmet" in acute pulmonary oedema. Report of 60 cases
    Tresoldi, S
    Maggiolini, S
    Bozzano, A
    Vitale, G
    Pesenti, A
    Fumagalli, R
    EUROPEAN HEART JOURNAL, 2002, 23 : 713 - 713
  • [48] Bench evaluation of flow limitation detection by automated continuous positive airway pressure devices
    Lofaso, Frederic
    Desmarais, Gilbert
    Leroui, Karl
    Zalc, Vincent
    Fodil, Redouane
    Isabey, Daniel
    Louis, Bruno
    CHEST, 2006, 130 (02) : 343 - 349
  • [49] Helmet-delivered continuous positive airway pressure with heliox in respiratory syncytial virus bronchiolitis
    Mayordomo-Colunga, J.
    Medina, A.
    Rey, C.
    Concha, A.
    Los Arcos, M.
    Menendez, S.
    ACTA PAEDIATRICA, 2010, 99 (02) : 308 - 311
  • [50] Ocular findings in sleep apnoea patients using continuous positive airway pressure
    Kadyan, A.
    Asghar, J.
    Dowson, L.
    Sandramouli, S.
    EYE, 2010, 24 (05) : 843 - 850