Magnetic Resonance Imaging Monitoring of Thermal Lesions Produced by Focused Ultrasound

被引:0
|
作者
Antoniou, Anastasia [1 ]
Evripidou, Nikolas [1 ]
Nikolaou, Anastasia [1 ]
Georgiou, Andreas [1 ]
Giannakou, Marinos [1 ]
Chrysanthou, Antreas [2 ]
Georgiou, Leonidas [2 ]
Ioannides, Cleanthis [2 ]
Damianou, Christakis [1 ]
机构
[1] Cyprus Univ Technol, Dept Elect Engn Comp Engn & Informat, 30 Archbishop Kyprianou St, CY-3036 Limassol, Cyprus
[2] German Oncol Ctr, Dept Intervent Radiol, Limassol, Cyprus
关键词
Contrast; lesion; monitoring; magnetic resonance imaging; porcine; ultrasound; NONENHANCED MRI; IN-VIVO; ABLATION; CONTRAST; SURGERY;
D O I
10.4103/jmu.jmu_112_23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background:The main goal of the study was to find the magnetic resonance imaging (MRI) parameters that optimize contrast between tissue and thermal lesions produced by focused ultrasound (FUS) using T1-weighted (T1-W) and T2-weighted (T2-W) fast spin echo (FSE) sequences.Methods:FUS sonications were performed in ex vivo porcine tissue using a single-element FUS transducer of 2.6 MHz in 1.5 and 3 T MRI scanners. The difference in relaxation times as well as the impact of critical MRI parameters on the resultant contrast-to-noise ratio (CNR) between coagulated and normal tissues were assessed. Discrete and overlapping lesions were inflicted in tissue with simultaneous acquisition of T2-W FSE images.Results:FUS lesions are characterized by lower relaxation times than intact porcine tissue. CNR values above 80 were sufficient for proper lesion visualization. For T1-W imaging, repetition time values close to 1500 ms were considered optimum for obtaining sufficiently high CNR at the minimum time cost. Echo time values close to 50 ms offered the maximum lesion contrast in T2-W FSE imaging. Monitoring of acute FUS lesions during grid sonications was performed successfully. Lesions appeared as hypointense spots with excellent contrast from surrounding tissue.Conclusion:MRI monitoring of signal intensity changes during FUS sonication in grid patterns using optimized sequence parameters can provide useful information about lesion progression and the success of ablation. This preliminary study demonstrated the feasibility of the proposed monitoring method in ex vivo porcine tissue and should be supported by in vivo studies to assess its clinical potential.
引用
收藏
页码:297 / 308
页数:12
相关论文
共 50 条
  • [21] Combined passive acoustic mapping and magnetic resonance thermometry for monitoring phase-shift nanoemulsion enhanced focused ultrasound therapy
    Crake, Calum
    Meral, F. Can
    Burgess, Mark T.
    Papademetriou, Iason T.
    McDannold, Nathan J.
    Porter, Tyrone M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (15) : 6144 - 6163
  • [22] Intracranial Applications of Magnetic Resonance-guided Focused Ultrasound
    Lipsman, Nir
    Mainprize, Todd G.
    Schwartz, Michael L.
    Hynynen, Kullervo
    Lozano, Andres M.
    NEUROTHERAPEUTICS, 2014, 11 (03) : 593 - 605
  • [23] Magnetic resonance imaging-guided focused ultrasound surgery for symptomatic uterine fibroids: estimation of treatment efficacy using thermal dose calculations
    Yoon, Sang-Wook
    Cha, Sun Hee
    Ji, Young Geon
    Kim, Hyun Cheol
    Lee, Mee Hwa
    Cho, Jin Ho
    EUROPEAN JOURNAL OF OBSTETRICS & GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 2013, 169 (02) : 304 - 308
  • [24] Magnetic resonance-guided focused ultrasound for uterine fibroids
    Magalha, Pedro Felipe
    Messina, Marcos de Lorenzo
    Soares Junior, Jose Maria
    Baracat, Edmund Chada
    CLINICS, 2025, 80
  • [25] Identification of occult breast lesions detected by magnetic resonance imaging with targeted ultrasound: A prospective study
    Aracava, Mrcia M.
    Chojniak, Rubens
    Souza, Juliana A.
    Bitencourt, Almir G. V.
    Marques, Elvira F.
    EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (03) : 516 - 519
  • [26] Adaptive model-predictive controller for magnetic resonance guided focused ultrasound therapy
    de Bever, Joshua
    Todd, Nick
    Payne, Allison
    Christensen, Douglas A.
    Roemer, Robert B.
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2014, 30 (07) : 456 - 470
  • [27] Responsive Nanoparticles to Enable a Focused Ultrasound-Stimulated Magnetic Resonance Imaging Spotlight
    Deng, Tian
    Zhang, Le
    Li, Xinzhou
    Zink, Jeffrey, I
    Wu, Holden H.
    ACS NANO, 2021, 15 (09) : 14618 - 14630
  • [28] Bilateral Magnetic Resonance Imaging-Guided Focused Ultrasound Thalamotomy for Essential Tremor
    Fukutome, Kenji
    Hirabayashi, Hidehiro
    Osakada, Yosuke
    Kuga, Yoshihiro
    Ohnishi, Hideyuki
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2022, 100 (01) : 44 - 52
  • [29] Decorrelated compounding of synthetic aperture ultrasound imaging to detect low contrast thermal lesions induced by focused ultrasound
    Nguyen, Michael
    Zhao, Na
    Xu, Yuan
    Tavakkoli, Jahangir
    ULTRASONICS, 2023, 134
  • [30] Intra-articular benign bone lesions treated with Magnetic Resonance-guided Focused Ultrasound (MRgFUS): imaging follow-up and clinical results
    Arrigoni, Francesco
    Barile, Antonio
    Zugaro, Luigi
    Splendiani, Alessandra
    Di Cesare, Ernesto
    Caranci, Ferdinando
    Ierardi, Anna Maria
    Floridi, Chiara
    Angileri, Alessio Salvatore
    Reginelli, Alfonso
    Brunese, Luca
    Masciocchi, Carlo
    MEDICAL ONCOLOGY, 2017, 34 (04)