Magnetic Resonance Imaging Monitoring of Thermal Lesions Produced by Focused Ultrasound

被引:0
|
作者
Antoniou, Anastasia [1 ]
Evripidou, Nikolas [1 ]
Nikolaou, Anastasia [1 ]
Georgiou, Andreas [1 ]
Giannakou, Marinos [1 ]
Chrysanthou, Antreas [2 ]
Georgiou, Leonidas [2 ]
Ioannides, Cleanthis [2 ]
Damianou, Christakis [1 ]
机构
[1] Cyprus Univ Technol, Dept Elect Engn Comp Engn & Informat, 30 Archbishop Kyprianou St, CY-3036 Limassol, Cyprus
[2] German Oncol Ctr, Dept Intervent Radiol, Limassol, Cyprus
关键词
Contrast; lesion; monitoring; magnetic resonance imaging; porcine; ultrasound; NONENHANCED MRI; IN-VIVO; ABLATION; CONTRAST; SURGERY;
D O I
10.4103/jmu.jmu_112_23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background:The main goal of the study was to find the magnetic resonance imaging (MRI) parameters that optimize contrast between tissue and thermal lesions produced by focused ultrasound (FUS) using T1-weighted (T1-W) and T2-weighted (T2-W) fast spin echo (FSE) sequences.Methods:FUS sonications were performed in ex vivo porcine tissue using a single-element FUS transducer of 2.6 MHz in 1.5 and 3 T MRI scanners. The difference in relaxation times as well as the impact of critical MRI parameters on the resultant contrast-to-noise ratio (CNR) between coagulated and normal tissues were assessed. Discrete and overlapping lesions were inflicted in tissue with simultaneous acquisition of T2-W FSE images.Results:FUS lesions are characterized by lower relaxation times than intact porcine tissue. CNR values above 80 were sufficient for proper lesion visualization. For T1-W imaging, repetition time values close to 1500 ms were considered optimum for obtaining sufficiently high CNR at the minimum time cost. Echo time values close to 50 ms offered the maximum lesion contrast in T2-W FSE imaging. Monitoring of acute FUS lesions during grid sonications was performed successfully. Lesions appeared as hypointense spots with excellent contrast from surrounding tissue.Conclusion:MRI monitoring of signal intensity changes during FUS sonication in grid patterns using optimized sequence parameters can provide useful information about lesion progression and the success of ablation. This preliminary study demonstrated the feasibility of the proposed monitoring method in ex vivo porcine tissue and should be supported by in vivo studies to assess its clinical potential.
引用
收藏
页码:297 / 308
页数:12
相关论文
共 50 条
  • [1] Magnetic resonance cavitation imaging for the monitoring of ultrasound therapies
    Ishak, Ounay
    Breton, Elodie
    Cabras, Paolo
    Dumont, Erik
    Mondou, Paul
    Novell, Anthony
    Larrat, Benoit
    Vappou, Jonathan
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (21)
  • [2] Thermal dosimetry of a focused ultrasound beam in vivo by magnetic resonance imaging
    Chung, AH
    Jolesz, FA
    Hynynen, K
    MEDICAL PHYSICS, 1999, 26 (09) : 2017 - 2026
  • [3] Ultrasound and Magnetic Resonance Imaging Correlation of Adnexal Lesions
    Neeharika, C.
    Ravindran, C.
    JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2021, 9 (05): : 282 - 289
  • [4] Monitoring imaging of lesions induced by high intensity focused ultrasound based on a matching pursuit method
    Song, Weidong
    Zhang, Siyuan
    Wan, Jinjin
    Wan, Mingxi
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [5] A long arm for ultrasound: A combined robotic focused ultrasound setup for magnetic resonance-guided focused ultrasound surgery
    Krafft, Axel J.
    Jenne, Juergen W.
    Maier, Florian
    Stafford, R. Jason
    Huber, Peter E.
    Semmler, Wolfhard
    Bock, Michael
    MEDICAL PHYSICS, 2010, 37 (05) : 2380 - 2393
  • [6] Magnetic Resonance Imaging-Guided Focused Ultrasound Positioning System for Preclinical Studies in Small Animals
    Drakos, Theoharis
    Giannakou, Marinos
    Menikou, Georgios
    Damianou, Christakis
    JOURNAL OF ULTRASOUND IN MEDICINE, 2021, 40 (07) : 1343 - 1352
  • [7] Magnetic Resonance Thermometry of Focused Ultrasound Using a Preclinical Focused Ultrasound Robotic System at 3T
    Filippou, Antria
    Evripidou, Nikolas
    Georgiou, Andreas
    Georgiou, Leonidas
    Chrysanthou, Antreas
    Ioannides, Cleanthis
    Damianou, Christakis
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (04) : 583 - 596
  • [8] Modeling for Quantitative Analysis of Nakagami Imaging in Accurate Detection and Monitoring of Therapeutic Lesions by High-Intensity Focused Ultrasound
    Han, Meng
    Song, Weidong
    Zhang, Fengshou
    Li, Zhenwei
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2023, 49 (07) : 1575 - 1585
  • [9] Imaging Findings in Papillary Breast Lesions: An Analysis of Ductal Findings on Magnetic Resonance Imaging and Ultrasound
    Sarica, Ozgur
    Dokdok, Murat
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (04) : 542 - 551
  • [10] Magnetic resonance imaging of boiling induced by high intensity focused ultrasound
    Khokhlova, Tatiana D.
    Canney, Michael S.
    Lee, Donghoon
    Marro, Kenneth I.
    Crum, Lawrence A.
    Khokhlova, Vera A.
    Bailey, Michael R.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2009, 125 (04) : 2420 - 2431