摘要
Introduction In atrial fibrillation (AF) ablation, fluoroscopy has been a standard tool for catheter guidance. However, the combination of electroanatomic mapping systems (EAMs) and intracardiac echocardiography (ICE) now allows for minimal or zero-fluoroscopy procedures. Concurrently, high-power short-duration (HPSD) ablation has emerged as a promising technique, offering enhanced resistive heating while reducing conductive heating. This approach potentially improves both safety and efficacy. Despite these advancements, there is a lack of comprehensive clinical data on the safety and effectiveness of HPSD ablation when used in conjunction with ICE-guided zero-fluoroscopy procedures.Objective To compare two different ablation strategies-high-power short-duration (HPSD) and low-power long-duration (LPLD)-both utilizing intracardiac echocardiography (ICE)-guided zero-fluoroscopy in the context of atrial fibrillation (AF) ablation.Methods This retrospective study included 173 consecutive patients with AF who underwent ICE-guided zero-fluoroscopy ablation. Patients were divided into two groups: HPSD and LPLD. All procedures were conducted using an EAM system with ICE guidance. Both groups underwent routine pulmonary vein isolation (PVI), with additional linear ablations performed for persistent AF when necessary. We compared treatment outcomes and the incidence of complications between the two groups.Results All procedures were successfully completed under ICE-guided zero-fluoroscopy, establishing a feasible and reliable workflow. The procedure and ablation times were significantly shorter in the HPSD group compared to the LPLD group. At one-year follow-up, sinus rhythm was maintained in 77 patients in the HPSD group and 74 patients in the LPLD group, with no significant difference between the two group. Postoperative complications occurred in 5 patients in the HPSD group and 3 patients in the LPLD group. Importantly, there were no major adverse cardiac and cerebrovascular events (MACCE) in either group.Conclusion A zero-fluoroscopy workflow utilizing an EAM system combined with ICE appears to be both feasible and safe for ablation in AF patients. In patients undergoing ICE-guided zero-fluoroscopy ablation, the HPSD strategy is comparable to LPLD ablation in effectiveness while offering the benefit of shorter procedure and ablation times.
引用
收藏
页数:10
相关论文
共 50 条
  • [11] High-power, short-duration ablation for atrial fibrillation: Pros and cons
    Knotts, Robert J.
    Barbhaiya, Chirag R.
    PROGRESS IN CARDIOVASCULAR DISEASES, 2021, 66 : 86 - 91
  • [12] Efficacy and safety of high-power short duration atrial fibrillation ablation in elderly patients
    Mueller, Julian
    Nentwich, Karin
    Berkovitz, Artur
    Ene, Elena
    Sonne, Kai
    Chakarov, Ivaylo
    Barth, Sebastian
    Waechter, Christian
    Luesebrink, Ulrich
    Behnes, Michael
    Akin, Ibrahim
    Deneke, Thomas
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (07) : 1425 - 1434
  • [13] High-power short-duration versus low-power long-duration ablation for pulmonary vein isolation: A substudy of the AWARE randomized controlled trial
    Joza, Jacqueline
    Nair, Girish M.
    Birnie, David H.
    Nery, Pablo B.
    Redpath, Calum J.
    Sarrazin, Jean-Francois
    Champagne, Jean
    Bernick, Jordan
    Wells, George A.
    Essebag, Vidal
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2024, 35 (01) : 136 - 145
  • [14] High-power short-duration ablation of atrial fibrillation: A contemporary review
    Naniwadekar, Aditi
    Dukkipati, Srinivas R.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (03): : 528 - 540
  • [15] Efficacy of high-power and short-duration ablation in patients with atrial fibrillation: a prospective randomized controlled trial
    Shin, Dong Geum
    Ahn, Jinhee
    Han, Sang-Jin
    Lim, Hong Euy
    EUROPACE, 2020, 22 (10): : 1495 - 1501
  • [16] High-power, short-duration ablation during Box isolation for atrial fibrillation
    Kumagai, Koichiro
    Toyama, Hideko
    JOURNAL OF ARRHYTHMIA, 2020, 36 (05) : 899 - 904
  • [17] Efficacy and Safety of High-power Short-duration Radio Frequency Ablation in the Treatment of Atrial Fibrillation: A Meta-analysis of Prospective Study
    Ma, Guohua
    Zhang, Weijiang
    HEART SURGERY FORUM, 2024, 27 (02) : E188 - E198
  • [18] Short and Long-Term Outcomes of Lesion Index-Guided High-Power Short-Duration Approach for Atrial Fibrillation Ablation
    Di Cori, Andrea
    Parollo, Matteo
    Gentile, Francesco
    Pistelli, Lorenzo
    Vitale, Carlo
    Della Volpe, Salvatore
    Santoro, Mario Giannotti
    Mazzocchetti, Lorenzo
    De Lucia, Raffaele
    Canu, Antonio
    Barletta, Valentina
    Grifoni, Gino
    Segreti, Luca
    Bongiorni, Maria Grazia
    Zucchelli, Giulio
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (15)
  • [19] Meta-analysis of high-power short-duration versus cryoballoon ablation for atrial fibrillation
    Lin, Limin
    Huang, Ying
    Huang, Qunying
    Yu, Fuling
    Mao, Yinjun
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2024, 47 (08): : 1013 - 1024
  • [20] HPSD ablation for AF high-power short-duration RF ablation for atrial fibrillation: A review
    Winkle, Roger A.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (10) : 2813 - 2823