Evaluating femoral head collapse risk post-fixation removal: a finite element analysis

被引:0
作者
Li, Xishan [1 ]
Zhou, Xiang [1 ,2 ]
Yang, Jie [1 ]
Boeker, Kai Oliver [1 ]
Schilling, Arndt F. [1 ]
Lehmann, Wolfgang [1 ]
机构
[1] Univ Med Ctr Gottingen, Dept Trauma Surg Orthoped & Plast Surg, Gottingen, Germany
[2] Fourth Peoples Hosp Guiyang, Dept Articular & Traumat Orthoped Surg, Guiyang, Guizhou, Peoples R China
来源
FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY | 2025年 / 13卷
关键词
finite element analysis; femoral head collapse; screw hole configuration; internal fixation removal; femoral neck fracture; NECK FRACTURE; INTERNAL-FIXATION; AVASCULAR NECROSIS; CORTICAL BONE; OSTEONECROSIS; COMPLICATIONS; MORTALITY; NONUNION; FEMUR;
D O I
10.3389/fbioe.2025.1441026
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Background Femoral neck fractures are prevalent in orthopedic injuries, often leading to complications such as nonunion and osteonecrosis of the femoral head (ONFH). Studies indicate that after healing and removal of internal fixation devices, some patients develop ONFH, while others experience osteosclerosis around the screw holes due to prolonged fixation, increasing ONFH risk. Despite such observations, biomechanical studies on this phenomenon are limited. This study assesses the risk of femoral head collapse post-internal fixation device removal and investigates the biomechanical effects of bone grafting at screw removal sites.Methods Using CT data, femoral anatomy was reconstructed. For control, the femoral head's collapse area was identified. Experimental models, divided into those with and without bone grafts in screw holes, incorporated three fixation techniques, namely, triple cannulated screws (3CS), dynamic hip screws with cannulated screws (DHS+CS), and the femoral neck system (FNS), further subclassified into normal and sclerotic screw-hole models. Stress distribution, stress values, stress index, and strain range were assessed.Results In both models, DHS+CS showed the highest stress in the overall model, while 3CS had the highest stress in the collapse area. The 3CS configuration also resulted in the largest strain range, which was observed in the central pillar of normal screw-hole models and the lateral pillar of sclerotic screw-hole models. The bone graft models exhibited lower peak, average stress, and strain values than the normal and sclerotic screw-hole models.Conclusion The FNS screw hole demonstrates a relatively lower mechanical risk of femoral head collapse. In contrast, sclerotic screw holes increase this risk, while bone grafting may improve the biomechanical behavior after fixation removal, potentially reducing the likelihood of femoral head collapse.
引用
收藏
页数:11
相关论文
共 57 条
  • [1] Excess mortality following hip fracture: a systematic epidemiological review
    Abrahamsen, B.
    van Staa, T.
    Ariely, R.
    Olson, M.
    Cooper, C.
    [J]. OSTEOPOROSIS INTERNATIONAL, 2009, 20 (10) : 1633 - 1650
  • [2] Logistic regression analysis of factors associated with avascular necrosis of the femoral head following femoral neck fractures in middle-aged and elderly patients
    Ai, Zi-Sheng
    Gao, You-Shui
    Sun, Yuan
    Liu, Yue
    Zhang, Chang-Qing
    Jiang, Cheng-Hua
    [J]. JOURNAL OF ORTHOPAEDIC SCIENCE, 2013, 18 (02) : 271 - 276
  • [3] Resorbable fillers reduce stress risers from empty screw holes
    Alford, J. Winslow
    Bradley, Michael P.
    Fadale, Paul D.
    Crisco, Joseph J.
    Moore, Douglas C.
    Ehrlich, Michael G.
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2007, 63 (03): : 647 - 654
  • [4] Comparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue
    Bayraktar, HH
    Morgan, EF
    Niebur, GL
    Morris, GE
    Wong, EK
    Keaveny, TM
    [J]. JOURNAL OF BIOMECHANICS, 2004, 37 (01) : 27 - 35
  • [5] Management of Acute Hip Fracture
    Bhandari, Mohit
    Swiontkowski, Marc
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2017, 377 (21) : 2053 - 2062
  • [6] BONE STRENGTH - EFFECT OF SCREW HOLES
    BURSTEIN, AH
    VESSELY, JC
    FRANKEL, VH
    LUNSETH, P
    CURREY, J
    HEIPLE, KG
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1972, 54 (06) : 1143 - +
  • [7] FATIGUE BEHAVIOR OF ADULT CORTICAL BONE - THE INFLUENCE OF MEAN STRAIN AND STRAIN RANGE
    CARTER, DR
    CALER, WE
    SPENGLER, DM
    FRANKEL, VH
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (05): : 481 - 490
  • [8] Finite element analysis of two cephalomedullary nails in treatment of elderly reverse obliquity intertrochanteric fractures: zimmer natural nail and proximal femoral nail antirotation-II
    Chen, Jian
    Ma, Jian-xiong
    Wang, Ying
    Bai, Hao-hao
    Sun, Lei
    Wang, Yan
    Lu, Bin
    Dong, Ben-chao
    Tian, Ai-xian
    Ma, Xin-long
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (01)
  • [9] Chu Kai, 2020, Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi, V34, P1091, DOI 10.7507/1002-1892.202001076
  • [10] Optimizing Limb Position for Measuring Knee Anatomical Axis Alignment from Standing Knee Radiographs
    Cooke, T. Derek V.
    Sled, Elizabeth A.
    [J]. JOURNAL OF RHEUMATOLOGY, 2009, 36 (03) : 472 - 477