Hall probe calibration in high-precision magnetic field mapping system of superconducting cyclotron

被引:0
作者
Xu, Manman [1 ]
Cao, Chuqing [2 ]
Fu, Yili [3 ]
Wu, Yonghong [4 ]
Wang, Xiangdong [5 ]
Su, Yongsheng [1 ]
Xi, Lin [1 ]
Liu, Yongming [1 ]
机构
[1] Anhui Polytech Univ, Sch Mech Engn, Wuhu 241000, Peoples R China
[2] Wuhu HIT Robot Technol Res Inst Co LTD, Wuhu 241000, Peoples R China
[3] Harbin Inst Technol, Harbin 150001, Peoples R China
[4] Anhui Univ Sci & Technol, Anhui Key Lab Mine Intelligent Equipment & Technol, Huainan 232001, Peoples R China
[5] Wuhu Magnet Wheel Transmiss Technol Co LTD, Wuhu 241000, Peoples R China
基金
中国国家自然科学基金;
关键词
Magnetic field mapping; Measurement accuracy; Polynomial fitting; Probe calibration; Superconducting cyclotron;
D O I
10.2478/nuka-2024-0026
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
IntroductionSuperconducting cyclotron can generate high-energy proton beams and are mainly used for radiation therapy of tumors and cancers. In the superconducting cyclotron SC200, the maximum magnetic induction intensity can typically reach up to 4.6 T, and the magnetic field accuracy is 1e-4. Hall probes are commonly used tools for measuring high-intensity magnetic fields.ObjectiveThrough comprehensive consideration, this study selects the SENIS Low-Noise Teslameter 3MH5 and Hall probe C to measure the magnetic field. When the magnetic field exceeds the range of 2 T, the measurement accuracy of the Hall probe is less than 1e-4, and the Hall probe needs to be calibrated to improve its measurement accuracy.MethodsThe Hall probes are calibrated using Swiss METROLAB PT2025 nuclear magnetic resonance (NMR) Tesla instrument and 1062 probe. Based on the calibration principle, a calibration system platform was built, test data were collected, and calibration curves were obtained. At the same time, the calibration data were analyzed through cross-validation experiments using the cubic polynomial fitting method.ResultsThe results indicate that the test deviation range is from -0.1 g to 0.1 g, and the measurement accuracy can reach 1e-4.ConclusionIn summary, the Hall probe can accurately measure the magnetic field distribution of the superconducting cyclotron. It can provide accurate and important data for the calculation and analysis of particle beam dynamics.
引用
收藏
页码:185 / 193
页数:9
相关论文
共 17 条
[11]  
Parameter Generation and Control, 2021, Walk-in stability rooms
[12]  
Popovic R., 1991, HALL EFFECT DEVICES
[13]   Advances in Magnetic Field Sensors [J].
Ripka, Pavel ;
Janosek, Michal .
IEEE SENSORS JOURNAL, 2010, 10 (06) :1108-1116
[14]  
Shen J. X., 2014, Instrument Technique and Sensor, V3, P1
[15]   Magnetic field sensor using the fiber loop ring-down technique and an etched fiber coated with magnetic fluid [J].
Shen, Tao ;
Feng, Yue ;
Sun, Binchao ;
Wei, Xinlao .
APPLIED OPTICS, 2016, 55 (04) :673-678
[16]   Integrated Fluxgate Magnetometer for Use in Isolated Current Sensing [J].
Snoeij, Martijn F. ;
Schaffer, Viola ;
Udayashankar, Sudarshan ;
Ivanov, Mikhail V. .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2016, 51 (07) :1684-1694
[17]  
[杨云 Yang Yun], 2018, [宇航计测技术, Journal of Astronautic Metrology and Measurement], V38, P37