Transcatheter aortic valve replacement in a bicuspid aortic valve with membranous interventricular septum aneurysm communicating with aortic root: a case report

被引:0
|
作者
Xi, Ziwei [1 ]
Yao, Jing [1 ]
Song, Guangyuan [1 ]
机构
[1] Capital Med Univ, Beijing Anzhen Hosp, Intervent Ctr Valvular Heart Dis, 2 Anzhen Rd, Beijing 100029, Peoples R China
关键词
Transcatheter aortic valve replacement; Membranous interventricular septum aneurysm; Case report;
D O I
10.1093/ehjcr/ytae523
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Membranous interventricular septum aneurysm (MISA) is a rare abnormality occurring in 0.3% of patients with congenital heart disease, which thereby increases anatomical complexity.Case summary Transcatheter aortic valve replacement (TAVR) procedure was planned for a 71-year-old female patient from East Asia with a type 1 bicuspid aortic valve diagnosed with severe aortic stenosis by transthoracic echocardiography (TTE). Pre-procedural multidetector computed tomography (MDCT) clearly revealed an extremely horizontal aorta and a MISA originating from the sub-annulus with the upper edge extending 7.2 mm above the annulus. A probable communicating flow between the left ventricle and the aorta was confirmed by reviewing the TTE images. Moreover, there was a calcified raphe between the left- and right-coronary cusps. A downsized balloon-expandable valve (a 23 mm Sapien 3 valve with an additional 2 mL dilation) was therefore chosen and deployed with a 100/0 aortic/ventricular ratio position. The TTE post-implantation indicated a trace perivalvular leakage. The cardiac MDCT performed post-procedure, at the 6-month, and 12-month follow-ups demonstrated complete sealing and significant healing of the aneurysm.Discussion Transcatheter aortic valve replacement utilizing a balloon-expandable valve was successfully performed for a case with membranous interventricular septum aneurysm extending above the annulus. Comprehensive imaging analysis before the procedure is crucial for TAVR with challenging anatomical conditions.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Transcatheter aortic valve replacement in membranous interventricular septum aneurysm with left ventricular outflow tract extension
    Banga, Sandeep
    Barzallo, Marco A.
    Nighswonger, Casey L.
    Mungee, Sudhir
    WORLD JOURNAL OF CARDIOLOGY, 2018, 10 (01): : 1 - 5
  • [2] Successful Transcatheter Aortic Valve Replacement for Bicuspid Aortic Valve
    Maeda, Koichi
    Kuratani, Toru
    Torikai, Kei
    Shimamura, Kazuo
    Ueno, Takayoshi
    Toda, Koichi
    Sawa, Yoshiki
    ANNALS OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 20 : 705 - 708
  • [3] Transcatheter Aortic Valve Replacement in Bicuspid Aortic Valve Disease
    Mylotte, Darren
    Lefevre, Thierry
    Sondergaard, Lars
    Watanabe, Yusuke
    Modine, Thomas
    Dvir, Danny
    Bosmans, Johan
    Tchetche, Didier
    Kornowski, Ran
    Sinning, Jan-Malte
    Theriault-Lauzier, Pascal
    O'Sullivan, Crochan J.
    Barbanti, Marco
    Debry, Nicolas
    Buithieu, Jean
    Codner, Pablo
    Dorfmeister, Magdalena
    Martucci, Giuseppe
    Nickenig, Georg
    Wenaweser, Peter
    Tamburino, Corrado
    Grube, Eberhard
    Webb, John G.
    Windecker, Stephan
    Lange, Ruediger
    Piazza, Nicolo
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 64 (22) : 2330 - 2339
  • [4] Transcatheter aortic valve replacement in bicuspid aortic valve disease
    Kochman, Janusz
    Rymuza, Bartosz
    Huczek, Zenon
    CURRENT OPINION IN CARDIOLOGY, 2015, 30 (06) : 594 - 602
  • [5] Transcatheter Aortic Valve Replacement in Bicuspid Aortic Valve Stenosis
    Vincent, Flavien
    Ternacle, Julien
    Denimal, Tom
    Shen, Mylene
    Redfors, Bjorn
    Delhaye, Cedric
    Simonato, Matheus
    Debry, Nicolas
    Verdier, Basile
    Shahim, Bahira
    Pamart, Thibault
    Spillemaeker, Hugues
    Schurtz, Guillaume
    Pontana, Francois
    Thourani, Vinod H.
    Pibarot, Philippe
    Van Belle, Eric
    CIRCULATION, 2021, 143 (10) : 1043 - 1061
  • [6] Transcatheter Aortic Valve Replacement in a Bicuspid Aortic Valve With Type A Aortic Dissection
    Putthapiban, Prapaipan
    Alani, Anas
    Jutzy, Kenneth
    Abudayyeh, Islam
    JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (18) : E243 - E245
  • [7] Research Progress of Transcatheter Aortic Valve Replacement in Aortic Valve Stenosis due to Bicuspid Aortic Valve
    Chen, Liyan
    Jiang, Li
    Zou, Mei
    Chen, Xuanlan
    Wu, Zhiyong
    HEART SURGERY FORUM, 2024, 27 (04) : E431 - E438
  • [8] Transcatheter aortic valve replacement for aortic regurgitation following valve sparing root replacement: a case series
    Sharma, Harish
    Mechery, Anthony
    Lawton, Ewa
    Nadir, M. Adnan
    Doshi, Sagar N.
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2025, 9 (01)
  • [9] Recent Advances in Aortic Valve Disease: Highlights From a Bicuspid Aortic Valve to Transcatheter Aortic Valve Replacement
    Augoustides, John G. T.
    Wolfe, Yanika
    Walsh, Elizabeth K.
    Szeto, Wilson Y.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2009, 23 (04) : 569 - 576
  • [10] Late Contained Aortic Root Rupture After Transcatheter Aortic Valve Replacement for Bicuspid Aortic Stenosis
    Ochiai, Tomoki
    Yoon, Sung-Han
    Flint, Nir
    Sharma, Rahul
    Patel, Chinar
    Patel, Vivek
    Chakravarty, Tarun
    Nakamura, Mamoo
    Cheng, Wen
    Makkar, Raj
    JACC-CARDIOVASCULAR INTERVENTIONS, 2019, 12 (14) : E121 - E122