Epidural magnetic stimulation of the motor cortex using an implantable coil

被引:0
作者
Lee, Kyeong Jae [1 ]
Jang, Jae-Won [1 ]
Kim, June Sic [2 ]
Kim, Sohee [1 ]
机构
[1] Daegu Gyeongbuk Inst Sci & Technol DGIST, Dept Robot & Mechatron Engn, Daegu, South Korea
[2] Konkuk Univ, Clin Res Inst, Med Ctr, Seoul, South Korea
关键词
Magnetic stimulation; Epidural stimulation; Low-intensity magnetic stimulation; Neuromodulation; Motor evoked potential; Implantable coil; CORTICAL STIMULATION; WAVE-FORM; RODENT;
D O I
10.1016/j.brs.2024.10.001
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Magnetic stimulation, represented by transcranial magnetic stimulation (TMS), is used to treat neurological diseases. Various strategies have been explored to improve the spatial resolution of magnetic stimulation. While reducing the coil size is the most impactful approach for increasing the spatial resolution, it decreases the stimulation intensity and increases heat generation. Objective: We aim to demonstrate the feasibility of magnetic stimulation using an epidurally implanted millimeter-sized coil and that it does not damage the cortical tissue via heating even when a repetitive stimulation protocol is used. Methods: A coil with dimensions of 3.5 x 3.5 x 2.6 mm(3) was epidurally implanted on the left motor cortex of rat, corresponding to the right hindlimb. Before and after epidural magnetic stimulation using a quadripulse stimulation (QPS) protocol, changes in the amplitude of motor evoked potentials (MEPs) elicited by a TMS coil were compared. Results: The experimental group showed an average increase of 88 % in MEP amplitude in the right hindlimb after QPS, whereas the MEP amplitude in the left hindlimb increased by 18 % on average. The control group showed no significant change in MEP amplitude after QPS in either hindlimb. The temperature changes at the coil surface remained <2 C-degrees during repetitive stimulation, meeting the thermal safety limit for implantable medical devices. Conclusion: These results demonstrate the feasibility of epidural magnetic stimulation using an implantable coil to induce neuromodulation effects. This novel method is expected to be a promising alternative for focal magnetic stimulation with an improved spatial resolution and lowered stimulus current than previous magnetic stimulation methods.
引用
收藏
页码:1157 / 1166
页数:10
相关论文
共 62 条
[1]   Epidural cortical stimulation enhances motor function after sensorimotor cortical infarcts in rats [J].
Adkins, DeAnna L. ;
Campos, Peter ;
Quach, David ;
Borromeo, Mark ;
Schallert, Kellan ;
Jones, Theresa A. .
EXPERIMENTAL NEUROLOGY, 2006, 200 (02) :356-370
[2]  
[Anonymous], MagPro R20 User Guide
[3]  
BARKER AT, 1991, ELECTROEN CLIN NEURO, P227
[4]   Theta-Burst Transcranial Magnetic Stimulation Alters Cortical Inhibition [J].
Benali, Alia ;
Trippe, Joern ;
Weiler, Elke ;
Mix, Annika ;
Petrasch-Parwez, Elisabeth ;
Girzalsky, Wolfgang ;
Eysel, Ulf T. ;
Erdmann, Ralf ;
Funke, Klaus .
JOURNAL OF NEUROSCIENCE, 2011, 31 (04) :1193-1203
[5]   Repetitive Transcranial Magnetic Stimulation to the Unilateral Hemisphere of Rat Brain [J].
Beom, Jaewon ;
Lee, Jung Chan ;
Paeng, Jin Chul ;
Han, Tai Ryoon ;
Bang, Moon Suk ;
Oh, Byung-Mo .
JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2016, (116)
[6]   Novel magnetic stimulation methodology for low-current implantable medical devices [J].
Bernardo, Rodrigo ;
Rodrigues, Andre ;
Soares dos Santos, Marco P. ;
Carneiro, Pedro ;
Lopes, Antonio ;
Amaral, Joao Sequeira ;
Amaral, Vitor Sequeira ;
Morais, Raul .
MEDICAL ENGINEERING & PHYSICS, 2019, 73 :77-84
[7]   Microscopic magnetic stimulation of neural tissue [J].
Bonmassar, Giorgio ;
Lee, Seung Woo ;
Freeman, Daniel K. ;
Polasek, Miloslav ;
Fried, Shelley I. ;
Gale, John T. .
NATURE COMMUNICATIONS, 2012, 3
[8]   In vivo low-intensity magnetic pulses durably alter neocortical neuron excitability and spontaneous activity [J].
Boyer, Manon ;
Baudin, Paul ;
Stengel, Chloe ;
Valero-Cabre, Antoni ;
Lohof, Ann M. ;
Charpier, Stephane ;
Sherrard, Rachel M. ;
Mahon, Severine .
JOURNAL OF PHYSIOLOGY-LONDON, 2022, 600 (17) :4019-4037
[9]   Epidural cortical stimulation and aphasia therapy [J].
Cherney, Leora R. ;
Harvey, Richard L. ;
Babbitt, Edna M. ;
Hurwitz, Rosalind ;
Kaye, Rosalind C. ;
Lee, Jaime B. ;
Small, Steven L. .
APHASIOLOGY, 2012, 26 (09) :1192-1217
[10]   Transcranial magnetic stimulation - Which part of the current waveform causes the stimulation? [J].
Corthout, E ;
Barker, AT ;
Cowey, A .
EXPERIMENTAL BRAIN RESEARCH, 2001, 141 (01) :128-132