Evaluation of a prototype array for daily quality assurance in spot scanning proton therapy

被引:0
作者
Flatten, Veronika [1 ,2 ]
Devendranath, Henry-Aravinth [3 ,4 ]
Kroh, Janik [2 ,5 ]
Witt, Matthias [2 ,6 ]
Baumann, Kilian-Simon [2 ,6 ,7 ]
Gall, Kenneth [1 ]
Simon, Bill [1 ]
Wulff, Joerg [3 ]
Schoenfeld, Andreas A. [1 ]
机构
[1] Mirion Med Co, SunNucl, Melbourne, FL 32940 USA
[2] Marburg Ion Beam Therapy Ctr MIT, Marburg, Germany
[3] Westdeutschen Protonentherapiezentrum Essen WPE, Essen, Germany
[4] Heinrich Heine Univ Dusseldorf, Dusseldorf, Germany
[5] Strahlentherapie MVZ Gesundheit Nordhessen, Kassel, Germany
[6] Univ Appl Sci, Inst Med Phys & Radiat Protect, Giessen, Germany
[7] Philipps Univ Marburg, Dept Radiotherapy & Radiooncol, Marburg, Germany
关键词
daily QA; PBS QA; proton therapy; DAILY QA; COMPACT;
D O I
10.1002/acm2.14454
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundQuality assurance (QA) on a daily basis is an essential task in radiotherapy. In pencil beam scanning proton therapy (PBS), there is a lack of available practical QA devices for routine daily QA in comparison to conventional radiotherapy.PurposeThe aim was to characterize and evaluate a prototype for the task of daily QA routine for PBS with parameters recommended by the AAPM TG 224, that is, the dose output constancy, the spot position and the distal range verification. Furthermore, a time efficient calibration method for fast and reliable daily QA routine was established for the prototype.MethodsFirst, a calibration routine was designed and evaluated, which characterizes the array response and allows for a conversion of the measured signal to clinically needed QA parameters. Finally, a time and resource efficient daily QA routine was developed and tested.ResultsThe prototype array can distinguish spot position deviations with sub-millimeter accuracy, as well as changes in the spot size in terms of FWHM with a 2%$\%$ sensitivity. The range and thus the energy can be evaluated at different depths also with sub-millimeter precision. After some training, the setup of the prototype device took roughly two minutes and the total beamtime was about one minute on cyclotron site and five minutes for synchrotrons.ConclusionsA prototype for daily QA in spot scanning proton therapy was evaluated, which features a fast and easy setup and allows for measuring relevant beam parameters, typically within less than a minute of beam time. All QA parameters as recommended by the AAPM TG 224 report can be analyzed with sufficient accuracy.
引用
收藏
页数:11
相关论文
共 16 条
[1]   A comprehensive and efficient daily quality assurance for PBS proton therapy [J].
Actis, O. ;
Meer, D. ;
Koenig, S. ;
Weber, D. C. ;
Mayor, A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (05) :1661-1675
[2]   AAPM task group 224: Comprehensive proton therapy machine quality assurance [J].
Arjomandy, Bijan ;
Taylor, Paige ;
Ainsley, Christopher ;
Safai, Sairos ;
Sahoo, Narayan ;
Pankuch, Mark ;
Farr, Jonathan B. ;
Park, Sung Yong ;
Klein, Eric ;
Flanz, Jacob ;
Yorke, Ellen D. ;
Followill, David ;
Kase, Yuki .
MEDICAL PHYSICS, 2019, 46 (08) :E678-E705
[3]   A fast and reliable method for daily quality assurance in spot scanning proton therapy with a compact and inexpensive phantom [J].
Bizzocchi, Nicola ;
Fracchiolla, Francesco ;
Schwarz, Marco ;
Algranati, Carlo .
MEDICAL DOSIMETRY, 2017, 42 (03) :238-246
[4]   An analytical approximation of the Bragg curve for therapeutic proton beams [J].
Bortfeld, T .
MEDICAL PHYSICS, 1997, 24 (12) :2024-2033
[5]   A simple and fast daily quality assurance solution for modulated scanning proton and carbon ion beams [J].
Chen, Jing-Ni ;
Moyers, Michael F. ;
Lin, Lien-Chun .
MEDICAL PHYSICS, 2021, 48 (10) :5684-5693
[6]   Quality assurance of scanned proton beams at different gantry angles using an ionization chamber array in a rotational phantom [J].
Decabooter, Esther ;
Roijen, Erik ;
Martens, Jonathan ;
Unipan, Mirko ;
Bosmans, Geert ;
Vilches-Freixas, Gloria .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 104 :67-74
[7]  
Ding X., 2021, Therap Radiol Oncol, V5, P22, DOI [10.21037/tro-21-11, DOI 10.21037/TRO-21-11]
[8]   A novel daily QA system for proton therapy [J].
Ding, Xiaoning ;
Zheng, Yuanshui ;
Zeidan, Omar ;
Mascia, Anthony ;
Hsi, Wen ;
Kang, Yixiu ;
Ramirez, Eric ;
Schreuder, Niek ;
Harris, Ben .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :115-126
[9]   MULTIPLE SCATTERING WITH ENERGY LOSS [J].
EYGES, L .
PHYSICAL REVIEW, 1948, 74 (10) :1534-1535
[10]   Daily QA in proton therapy using a single commercially available detector [J].
Lambert, Jamil ;
Baeumer, Christian ;
Koska, Benjamin ;
Ding, Xiaoning .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (06) :217-228