A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy

被引:0
|
作者
Alzahrani, Meshal [1 ,2 ,3 ]
Broadbent, David A. [2 ,3 ]
Teh, Irvin [3 ]
Al-Qaisieh, Bashar [2 ]
Johnstone, Emily [4 ]
Speight, Richard [2 ]
机构
[1] King Abdulaziz Univ, Fac Appl Med Sci, Dept Radiol Sci, Jeddah, Saudi Arabia
[2] Leeds Teaching Hosp NHS Trust, Dept Med Phys & Engn, Leeds, England
[3] Univ Leeds, Leeds Inst Cardiovasc & Metab Med, Leeds, England
[4] Christie NHS Fdn Trust, Christie Med Phys & Engn, Manchester, England
来源
PHYSICS & IMAGING IN RADIATION ONCOLOGY | 2025年 / 33卷
关键词
Magnetic resonance imaging; Radiotherapy; MRI in RT; Quality assurance; ACR phantom; Multimodal anthropomorphic phantom; IMPLEMENTATION; SIMULATION;
D O I
10.1016/j.phro.2025.100707
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: The use of magnetic resonance imaging (MRI) for radiotherapy (RT) simulation has grown, prompting quality assurance (QA) guidelines by the Institute of Physics and Engineering in Medicine (IPEM) and the American Association of Physicists in Medicine (AAPM). This study compares a novel multimodality anthropomorphic phantom to an American College of Radiology (ACR) phantom for a subset of these MRI-specific QA tests in RT. Materials and methods: A novel 3D-printed multimodality head-and-neck anthropomorphic phantom was compared to an ACR large MRI phantom. IPEM and AAPM-recommended QA tests were conducted, including informatics/connectivity/data transfer, MRI-CT registration, end-to-end QA, and signal-to-noise ratio (SNR)/ percentage integral uniformity (PIU) assessments using RT accessories. Results: Both phantoms were suitable for informatics/connectivity/data transfer. In MRI-CT registration, no errors were found; the ACR phantom offered more quantitative landmarks, while the anthropomorphic phantom provided limited structures. Both phantoms achieved target registration errors (TREs) below 0.97 mm and dice similarity coefficient (DSC) values above 0.9, meeting guidelines. For end-to-end QA, the anthropomorphic phantom facilitated dose measurements of 1.994 Gy versus a calculated 2.01 Gy (-0.8 %). SNR and PIU assessments showed higher values in radiology setups compared to RT setups for both phantoms. Conclusions: Multimodality anthropomorphic phantoms compatible with dosimetric equipment allow realistic end-to-end QA, unlike the ACR phantom. While the ACR phantom is suitable for informatics and MRI-CT registration, anthropomorphic phantoms better represent clinical scenarios. For comprehensive QA, both ACR and anthropomorphic phantoms are required. Additionally, large field-of-view (FOV) phantoms are crucial for evaluating large FOV MRI distortions.
引用
收藏
页数:7
相关论文
共 41 条
  • [11] A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems
    Yu, Amy S.
    Fowler, Tyler L.
    Dubrowski, Piotr
    MEDICAL PHYSICS, 2018, 45 (07) : 2857 - 2863
  • [12] Cone beam computed tomography for dose calculation quality assurance for magnetic resonance-only radiotherapy
    Wyatt, Jonathan J.
    Pearson, Rachel A.
    Walker, Christopher P.
    Brooks, Rachel L.
    Pilling, Karen
    McCallum, Hazel M.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 : 71 - 76
  • [13] Task group 284 report: magnetic resonance imaging simulation in radiotherapy: considerations for clinical implementation, optimization, and quality assurance
    Glide-Hurst, Carri K.
    Paulson, Eric S.
    McGee, Kiaran
    Tyagi, Neelam
    Hu, Yanle
    Balter, James
    Bayouth, John
    MEDICAL PHYSICS, 2021, 48 (07) : E636 - E670
  • [14] A Novel Magnetic Resonance Quality Assurance Phantom (KMRP-4) Multi-Site Comparison With the : American College of Radiology Phantom
    Kumar, Suchit
    Kim, Jong-Min
    Cha, Jang Gyu
    Hwang, Ji Young
    Jung, Seung Eun
    Lee, Chulhyun
    Oh, Chang-Hyun
    INVESTIGATIVE MAGNETIC RESONANCE IMAGING, 2023, 27 (01) : 10 - 20
  • [15] Factors influencing daily quality assurance measurements of magnetic resonance imaging scanners
    Owusu, Nana
    Magnotta, Vincent A.
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2021, 14 (04) : 396 - 401
  • [16] Factors influencing daily quality assurance measurements of magnetic resonance imaging scanners
    Nana Owusu
    Vincent A. Magnotta
    Radiological Physics and Technology, 2021, 14 : 396 - 401
  • [17] Design of a fused phantom for quantitative evaluation of brain metabolites and enhanced quality assurance testing for magnetic resonance imaging and spectroscopy
    Song, Kyu-Ho
    Kim, Sang-Young
    Lee, Do-Wan
    Jung, Jin-Young
    Lee, Jung-Hoon
    Baek, Hyeon-Man
    Choe, Bo-Young
    JOURNAL OF NEUROSCIENCE METHODS, 2015, 255 : 75 - 84
  • [18] A Novel Polymer Gel-Loading Phantom for Magnetic Resonance Imaging
    Cheng, Kai-Yuan
    Chen, Sih-Ying
    Shieh, Jiunn-I
    Wu, Yi-Ying
    Hsieh, Ling-Ling
    ARABIAN JOURNAL FOR SCIENCE AND ENGINEERING, 2016, 41 (01) : 163 - 167
  • [19] A Novel Polymer Gel-Loading Phantom for Magnetic Resonance Imaging
    Kai-Yuan Cheng
    Sih-Ying Chen
    Jiunn-I Shieh
    Yi-Ying Wu
    Ling-Ling Hsieh
    Arabian Journal for Science and Engineering, 2016, 41 : 163 - 167
  • [20] An Automatic Image Processing Workflow for Daily Magnetic Resonance Imaging Quality Assurance
    Juha I. Peltonen
    Teemu Mäkelä
    Alexey Sofiev
    Eero Salli
    Journal of Digital Imaging, 2017, 30 : 163 - 171