A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy

被引:1
作者
Alzahrani, Meshal [1 ,2 ,3 ]
Broadbent, David A. [2 ,3 ]
Teh, Irvin [3 ]
Al-Qaisieh, Bashar [2 ]
Johnstone, Emily [4 ]
Speight, Richard [2 ]
机构
[1] King Abdulaziz Univ, Fac Appl Med Sci, Dept Radiol Sci, Jeddah, Saudi Arabia
[2] Leeds Teaching Hosp NHS Trust, Dept Med Phys & Engn, Leeds, England
[3] Univ Leeds, Leeds Inst Cardiovasc & Metab Med, Leeds, England
[4] Christie NHS Fdn Trust, Christie Med Phys & Engn, Manchester, England
关键词
Magnetic resonance imaging; Radiotherapy; MRI in RT; Quality assurance; ACR phantom; Multimodal anthropomorphic phantom; IMPLEMENTATION; SIMULATION;
D O I
10.1016/j.phro.2025.100707
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: The use of magnetic resonance imaging (MRI) for radiotherapy (RT) simulation has grown, prompting quality assurance (QA) guidelines by the Institute of Physics and Engineering in Medicine (IPEM) and the American Association of Physicists in Medicine (AAPM). This study compares a novel multimodality anthropomorphic phantom to an American College of Radiology (ACR) phantom for a subset of these MRI-specific QA tests in RT. Materials and methods: A novel 3D-printed multimodality head-and-neck anthropomorphic phantom was compared to an ACR large MRI phantom. IPEM and AAPM-recommended QA tests were conducted, including informatics/connectivity/data transfer, MRI-CT registration, end-to-end QA, and signal-to-noise ratio (SNR)/ percentage integral uniformity (PIU) assessments using RT accessories. Results: Both phantoms were suitable for informatics/connectivity/data transfer. In MRI-CT registration, no errors were found; the ACR phantom offered more quantitative landmarks, while the anthropomorphic phantom provided limited structures. Both phantoms achieved target registration errors (TREs) below 0.97 mm and dice similarity coefficient (DSC) values above 0.9, meeting guidelines. For end-to-end QA, the anthropomorphic phantom facilitated dose measurements of 1.994 Gy versus a calculated 2.01 Gy (-0.8 %). SNR and PIU assessments showed higher values in radiology setups compared to RT setups for both phantoms. Conclusions: Multimodality anthropomorphic phantoms compatible with dosimetric equipment allow realistic end-to-end QA, unlike the ACR phantom. While the ACR phantom is suitable for informatics and MRI-CT registration, anthropomorphic phantoms better represent clinical scenarios. For comprehensive QA, both ACR and anthropomorphic phantoms are required. Additionally, large field-of-view (FOV) phantoms are crucial for evaluating large FOV MRI distortions.
引用
收藏
页数:7
相关论文
共 26 条
[1]  
American College of Radiology ACR, 2015, Am Coll Radiol
[2]  
[Anonymous], 2013, Phantom patient for stereotactic end-to-end verification
[3]   An anthropomorphic maxillofacial phantom using 3-dimensional printing, polyurethane rubber and epoxy resin for dental imaging and dosimetry [J].
Badiuk, Sawyer Rhae ;
Sasaki, David K. ;
Rickey, Daniel W. .
DENTOMAXILLOFACIAL RADIOLOGY, 2022, 51 (01)
[4]   Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task [J].
Brock, Kristy K. ;
Mutic, Sasa ;
McNutt, Todd R. ;
Li, Hua ;
Kessler, Marc L. .
MEDICAL PHYSICS, 2017, 44 (07) :E43-E76
[5]  
CIRS, 2014, MRI Distortion Phantom for SRS
[6]   Development and evaluation of a novel MR-compatible pelvic end-to-end phantom [J].
Cunningham, Justine M. ;
Barberi, Enzo A. ;
Miller, John ;
Kim, Joshua P. ;
Glide-Hurst, Carri K. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01) :265-275
[7]   IPEM code of practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service [J].
Eaton, David J. ;
Bass, Graham ;
Booker, Paul ;
Byrne, John ;
Duane, Simon ;
Frame, John ;
Grattan, Mark ;
Thomas, Russell A. S. ;
Thorp, Natalie ;
Nisbet, Andrew .
PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (19)
[8]   Design, development, and implementation of the Radiological Physics Center's pelvis and thorax anthropomorphic quality assurance phantoms [J].
Followill, David S. ;
Evans, DeeAnn Radford ;
Cherry, Christopher ;
Molineu, Andrea ;
Fisher, Gary ;
Hanson, William F. ;
Lbbott, Geoffrey S. .
MEDICAL PHYSICS, 2007, 34 (06) :2070-2076
[9]   An anthropomorphic multimodality (CT/MRI) head phantom prototype for end-to-end tests in ion radiotherapy [J].
Gallas, Raya R. ;
Huenemohr, Nora ;
Runz, Armin ;
Niebuhr, Nina I. ;
Jaekel, Oliver ;
Greilich, Steffen .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2015, 25 (04) :391-399
[10]   Task group 284 report: magnetic resonance imaging simulation in radiotherapy: considerations for clinical implementation, optimization, and quality assurance [J].
Glide-Hurst, Carri K. ;
Paulson, Eric S. ;
McGee, Kiaran ;
Tyagi, Neelam ;
Hu, Yanle ;
Balter, James ;
Bayouth, John .
MEDICAL PHYSICS, 2021, 48 (07) :E636-E670