Predicting the biomechanical behavior of lumbar intervertebral Discs: A comparative finite element analysis of a novel artificial disc design

被引:0
|
作者
Khanna, Ashutosh [1 ]
Jain, Pushpdant [1 ]
Paul, C. P. [2 ]
机构
[1] VIT Bhopal Univ, Sch Mech Engn, Indore Bhopal Highway, Sehore 466114, Madhya Pradesh, India
[2] Raja Ramanna Ctr Adv Technol, Laser Technol Div, Laser Addit Mfg Lab, Indore 452013, Madhya Pradesh, India
关键词
Artificial Intervertebral Disc; Finite Element Analysis; Computer-Aided Design; Biomechanics; Lumbar Spine; Implants; Stress Analysis; SPINE; FIXATION; TITANIUM; SEGMENTS; MODELS; STRAIN; METAL; ROD;
D O I
10.1016/j.jocn.2024.110960
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Osseointegration along with better mimicry of natural bone behaviour addresses the long-term performance of artificial intervertebral disc prosthesis. Here the effect of a novel artificial intervertebral disc geometry on stress, deformation and strain on lumbar segments to restore movement of the spine was investigated. The process involved, using CT image data, and solid modelling, simulation-driven design and finite element (FE) analysis, hexahedral mesh sensitivity analysis, implant placements. The range of motion (ROM) was calculated using an ANSYS deformation probe. The intact lumbar spine model established was compared with two implants, replacement at segment L4-L5 level, and biomechanical results were compared using axial loads of 500 N, 800 N, 1000 N and 10Nm moment. The two lumbosacral FE models, a novel implant Titanium Conix (TIC) and another FDA approved SB ChariteTM (SBC) implant were considered. Novel TIC implant geometry exhibited comparable ROM values in four physiological motions, which were comparable to as required for restoring natural motion. The result shows that the proposed TIC observed the deformation during flexion, extension, bending and twist as 3.43 mm, 3.19 mm, 3.33 mm and 3.48 mm respectively. Similarly strain of 0.01 during flexion, 0.02 during extension, 0.01 during bending and 0.02 during twist. The implants designed in this study demonstrate the suitability of titanium alloy in endplates and annulus. The FE models in the study with their biomechanical parameters can be considered before clinical implementation of any implants, pre-surgery evaluations, implant placement simulations, postsurgical response, follow-up revisions, implant customization and manufacturing.
引用
收藏
页数:13
相关论文
共 50 条
  • [31] Finite Element Study of a Lumbar Intervertebral Disc Nucleus Replacement Device
    Coogan, Jessica S.
    Francis, W. Loren
    Eliason, Travis D.
    Bredbenner, Todd L.
    Stemper, Brian D.
    Yoganandan, Narayan
    Pintar, Frank A.
    Nicolella, Daniel P.
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2016, 4
  • [32] Comparison of novel stabilisation device with various stabilisation approaches: A finite element based biomechanical analysis
    Jain, Pushpdant
    Khan, Mohammed Rajik
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2022, 45 (05) : 514 - 522
  • [33] Improvement in the Process of Designing a New Artificial Human Intervertebral Lumbar Disc Combining Soft Computing Techniques and the Finite Element Method
    Lostado Lorza, Ruben
    Somovilla Gomez, Fatima
    Fernandez Martinez, Roberto
    Garcia, Ruben Escribano
    Corral Bobadilla, Marina
    INTERNATIONAL JOINT CONFERENCE SOCO'16- CISIS'16-ICEUTE'16, 2017, 527 : 223 - 232
  • [34] BIOMECHANICAL PERFORMANCE OF A MODIFIED DESIGN OF DYNAMIC CERVICAL IMPLANT COMPARED TO CONVENTIONAL BALL AND SOCKET DESIGN OF AN ARTIFICIAL INTERVERTEBRAL DISC IMPLANT: A FINITE ELEMENT STUDY
    Bhattacharya, Shambo
    Roy, Sandipan
    Rana, Masud
    Banerjee, Sreerup
    Karmakar, Santanu Kumar
    Biswas, Jayanta Kumar
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2019, 19 (04)
  • [35] Effect of single and multilevel artificial inter-vertebral disc replacement in lumbar spine: A finite element study
    Biswas, Jayanta Kumar
    Rana, Masud
    Malas, Anindya
    Roy, Sandipan
    Chatterjee, Subhomoy
    Choudhury, Sandeep
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2022, 45 (02) : 193 - 199
  • [36] Geometric determinants of the mechanical behavior of image-based finite element models of the intervertebral disc
    Fleps, Ingmar
    Newman, Harrah R.
    Elliott, Dawn M.
    Morgan, Elise F.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2024, 42 (06) : 1343 - 1355
  • [37] Effect of asymptomatic intervertebral flexion patterns on lumbar disc pressure: A finite element analysis study
    Nematimoez, Mehdi
    Haddas, Ram
    Breen, Alexander
    INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2024, 40 (11)
  • [38] Biomechanical Analysis of Different Lumbar Interspinous Process Devices: A Finite Element Study
    Shen, Hangkai
    Fogel, Guy R.
    Zhu, Jia
    Liao, Zhenhua
    Liu, Weiqiang
    WORLD NEUROSURGERY, 2019, 127 : E1112 - E1119
  • [39] Biomechanical Analysis of the External Fixation in a Lumbar Fracture Model: A Finite Element Study
    Liao, Yijia
    Yan, Yiguo
    Kang, Yu
    Wang, Wenjun
    Song, Xizheng
    Peng, Wen
    Fu, Haibin
    Chen, Huarong
    Wang, Cheng
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2022, 42 (04) : 469 - 478
  • [40] Application of Finite Element Analysis for Investigation of Intervertebral Disc Degeneration: from Laboratory to Clinic
    Bin-wu Hu
    Xiao Lv
    Song-feng Chen
    Zeng-wu Shao
    Current Medical Science, 2019, 39 : 7 - 15