Image quality of virtual monochromatic and material density iodine images for evaluation of head and neck neoplasms using deep learning-based CT image reconstruction - A retrospective observational study

被引:0
作者
Buerckenmeyer, Florian [1 ]
Graeger, Stephanie [1 ]
Mlynska, Lucja [1 ]
Guettler, Felix [1 ]
Ingwersen, Maja [1 ]
Teichgraeber, Ulf [1 ]
Kraemer, Martin [1 ]
机构
[1] Friedrich Schiller Univ Jena, Jena Univ Hosp, Dept Diagnost & Intervent Radiol, Am Klinikum 1, D-07747 Jena, Germany
关键词
Computed tomography; Computer-assisted image processing; Deep-learning image reconstruction; Head and neck neoplasms;
D O I
10.1016/j.ejrad.2024.111806
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the quality of deep learning image reconstructed (DLIR) virtual monochromatic images (VMI) and material density (MD) iodine images from dual-energy computed tomography (DECT) for the evaluation of head and neck neoplasms with CT scans from a conventional single-energy protocol. Method: A total of 294 head and neck CT scans (98 VMIs operated at 60 keV, 102 MD iodine images, and 94 images from a 120 kVp single-energy CT (SECT) protocol) were retrospectively evaluated. VMIs and MD iodine images were generated using the Gemstone Spectral Imaging (GSI) mode using DLIR and metal artifact reduction (MAR) algorithms. SECT images were generated using adaptive statistical iterative reconstruction (ASIR-V). Images were scored by two independent readers on a 6-point Likert-type scale for overall image quality, vessel contrast, soft tissue contrast, noise texture, noise intensity, artifact reduction, and sharpness. Results: Subjective overall image quality was rated as superior or excellent in 98 % of DLIR-based MD iodine images and VMIs, but only in 55 % of ASIR-V-based SECT images. For each individual quality criterion, image quality of VMIs and MD iodine images was rated as better than that of SECT images (p < 0.001 in each case). Noise texture and intensity were rated better in MD iodine images than in VMIs. Conclusion: DECT using both DLIR and MAR for the generation of VMIs and MD iodine images resulted in higher subjective quality of oncologic head and neck images than ASIR-V-based SECT. Noise reduction and noise texture were best achieved with DLIR-based MD iodine images.
引用
收藏
页数:5
相关论文
共 18 条
[1]   Dual-energy computed tomography for improved visualization of internal jugular chain neck lymph node metastasis and nodal necrosis in head and neck squamous cell carcinoma [J].
Baba, Akira ;
Kurokawa, Ryo ;
Kurokawa, Mariko ;
Rivera-de Choudens, Roberto ;
Srinivasan, Ashok .
JAPANESE JOURNAL OF RADIOLOGY, 2023, 41 (12) :1351-1358
[2]   Iterative Metal Artifact Reduction in Head and Neck CT Facilitates Tumor Visualization of Oral and Oropharyngeal Cancer Obscured by Artifacts From Dental Hardware [J].
Bayerl, Nadine ;
May, Matthias Stefan ;
Wuest, Wolfgang ;
Roth, Jan-Peter ;
Kramer, Manuel ;
Hofmann, Christian ;
Schmidt, Bernhard ;
Uder, Michael ;
Ellmann, Stephan .
ACADEMIC RADIOLOGY, 2023, 30 (12) :2962-2972
[3]   Image quality comparison of single-energy and dual-energy computed tomography for head and neck patients: a prospective randomized study [J].
Bedernik, Andreas ;
Wuest, Wolfgang ;
May, Matthias Stefan ;
Heiss, Rafael ;
Uder, Michael ;
Wiesmueller, Marco .
EUROPEAN RADIOLOGY, 2022, 32 (11) :7700-7709
[4]   Pros and Cons of Dual-Energy CT Systems: "One Does Not Fit All" [J].
Borges, Ana P. P. ;
Antunes, Celia ;
Curvo-Semedo, Luis .
TOMOGRAPHY, 2023, 9 (01) :195-216
[5]   Low-dose CT urography using deep learning image reconstruction: a prospective study for comparison with conventional CT urography [J].
Cheng, Yannan ;
Han, Yangyang ;
Li, Jianying ;
Fan, Ganglian ;
Cao, Le ;
Li, Junjun ;
Jia, Xiaoqian ;
Yang, Jian ;
Guo, Jianxin .
BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1120)
[6]   Advanced dual-energy CT applications for the evaluation of the soft tissues of the neck [J].
Forghani, R. ;
Mukherji, S. K. .
CLINICAL RADIOLOGY, 2018, 73 (01) :70-80
[7]   Spectral CT imaging: Technical principles of dual-energy CT and multi-energy photon-counting CT [J].
Greffier, Joel ;
Villani, Nicolas ;
Defez, Didier ;
Dabli, Djamel ;
Si-Mohamed, Salim .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2023, 104 (04) :167-177
[8]   Ionizing radiation sensitivity of the ocular lens and its dose rate dependence [J].
Hamada, Nobuyuki .
INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2017, 93 (10) :1024-1034
[9]   Deep Learning Image Reconstruction for CT: Technical Principles and Clinical Prospects [J].
Koetzier, Lennart R. ;
Mastrodicasa, Domenico ;
Szczykutowicz, Timothy P. ;
van der Werf, Niels R. ;
Wang, Adam S. ;
Sandfort, Veit ;
van der Molen, Aart J. ;
Fleischmann, Dominik ;
Willemink, Martin J. .
RADIOLOGY, 2023, 306 (03)
[10]   Head and neck single- and dual-energy CT: differences in radiation dose and image quality of 2nd and 3rd generation dual-source CT [J].
Lenga, Lukas ;
Lange, Marvin ;
Martin, Simon S. ;
Albrecht, Moritz H. ;
Booz, Christian ;
Yel, Ibrahim ;
Arendt, Christophe T. ;
Vogl, Thomas J. ;
Leithner, Doris .
BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1122)