The objective of this note was to systematically study the effects from nonpoint radiation sources on the nuclear medicine imaging room or uptake room shielding design. The dose from a nonpoint source to a location at a certain distance to the center of the source was calculated. A correction factor could be applied to the dose calculated using a point source model with the same distance to compensate the nonpoint source effects. The correction factor was a function only depending on the source relative dimension, the ratio of the source's geometric dimension to the distance. Theoretical and numerical calculation were performed to calculate the correction factor on a one-dimensional line source, two-dimensional circular disc source, three-dimensional sphere, and cylinder sources. Our study indicated that for large nuclear medicine imaging rooms, the point source model was a good approximation for shielding design. For a small uptake room, a nonpoint source correction may be considered to calculate the barrier thickness.
引用
收藏
页码:413 / 419
页数:7
相关论文
共 7 条
[1]   The Nuclear Medicine Patient as a Line Source: The Source Length Is Certainly Not the Patient Height, But It Is a Reasonable Approximation [J].
Broggio, David .
HEALTH PHYSICS, 2022, 123 (03) :208-217
[2]  
Cember H., 2009, New York: McGraw Hill, V2009, P516
[3]   Comparing Dose Rates Near a Radioactive Patient Evaluated Using Various Source Models: Point, Line, Cylinder, and Anthropomorphic Phantoms [J].
Liu, Yi-Ching ;
Lee, Kuo-Wei ;
Sheu, Rong-Jiun .
HEALTH PHYSICS, 2015, 109 (01) :69-77
[4]   AAPM task group 108: PET and PET/CT shielding requirements [J].
Madsen, MT ;
Anderson, JA ;
Halama, JR ;
Kleck, J ;
Simpkin, DJ ;
Votaw, JR ;
Wendt, RE ;
Williams, LE ;
Yester, MV .
MEDICAL PHYSICS, 2006, 33 (01) :4-15
[5]  
Mortality Tables, 2023, Body measurements online
[6]  
Siegel JA, 2002, J NUCL MED, V43, P1241
[7]   Correction factors for more accurate estimates of exposure rates near radioactive patients: Experimental, point, and line source models [J].
Willegaignon, J. ;
Guimaraes, Maria I. C. ;
Stabin, Michael G. ;
Sapienza, Marcelo T. ;
Malvestiti, Luiz F. ;
Marone, Marilia M. S. ;
Sordi, Gian-Maria A. A. .
HEALTH PHYSICS, 2007, 93 (06) :678-688