Accuracy of new intraocular lens power calculation formula for short and long eyes using segmental refractive indices

被引:5
|
作者
Kato, Yukihito [1 ]
Ayaki, Masahiko [2 ]
Tamaoki, Akeno [3 ]
Tanaka, Yoshiki [1 ]
Ichikawa, Kei [1 ]
Ichikawa, Kazuo [1 ]
机构
[1] Chukyo Eye Clin, 12-22 Sanbonmatsu Cho,Atsuta Ku, Nagoya, Aichi 4560032, Japan
[2] Otake Eye Clin, 521-8 Shimotsuruma, Yamato, Kanagawa 2420001, Japan
[3] Chukyo Hosp, Div Ophthalmol, Japan Community Hlth Care Org, Nagoya, Japan
来源
关键词
PARTIAL-COHERENCE INTERFEROMETRY; GRADIENT-INDEX; BIOMETRY; TOMOGRAPHY; ERROR;
D O I
10.1097/j.jcrs.0000000000001466
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the accuracy of a new intraocular lens power calculation formula using segmental refractive index-based axial length (AL). Setting: Chukyo Eye Clinic, Nagoya, Japan. Design: Retrospective observational study. Methods: This study included patients undergoing preoperative examination for cataract surgery with the new Barrett True AL (BTAL) and Emmetropia Verifying Optical (EVO) formulas using segmental refractive index, and conventional Barrett Universal II (BU II) formula using equivalent refractive index. The predicted refractive error of each formula was compared with the postoperative subjective spherical equivalent. Results: The mean prediction error (MPE) in the short AL group (<= 22 mm; 44 eyes) was 0.32 +/- 0.40 diopter (D) for BU II, 0.22 +/- 0.37 D for BTAL, and 0.10 +/- 0.37 D for EVO (P < .0001). MPE in the long AL group (>= 26 mm; 92 eyes) was 0.01 +/- 0.32 D for BU II, 0.04 +/- 0.32 D for BTAL, and 0.09 +/- 0.32 D for EVO (P < .0001). In patients with an AL >= 28 mm, BU II showed a myopic trend in 57.1% of cases, while BTAL and EVO showed a hyperopic trend in 71.4%. The MPE for patients with an AL >= 28 mm was -0.16 +/- 0.34 D for BU II, 0.18 +/- 0.33 D for BTAL, and 0.16 +/- 0.32 D for EVO (P < .0001). Conclusions: The new EVO and BTAL formulas showed higher accuracy than BU II in short eyes, whereas there was no difference in long eyes.
引用
收藏
页码:810 / 815
页数:6
相关论文
共 50 条
  • [1] ACCURACY OF THE NEWER GENERATION INTRAOCULAR-LENS POWER CALCULATION FORMULAS IN LONG AND SHORT EYES
    OLSEN, T
    THIM, K
    CORYDON, L
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1991, 17 (02): : 187 - 193
  • [2] Intraocular lens power calculation in short eyes
    Gavin, E. A.
    Hammond, C. J.
    EYE, 2008, 22 (07) : 935 - 938
  • [3] Intraocular lens power calculation in short eyes
    E A Gavin
    C J Hammond
    Eye, 2008, 22 : 935 - 938
  • [4] Accuracy of intraocular lens power calculation in eyes with keratoconus
    Kojima, Takashi
    Horai, Rie
    Isogai, Naoki
    Nakamura, Tomoaki
    Tamaoki, Akeno
    Ichikawa, Kazuo
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (09)
  • [5] A new formula for intraocular lens power calculation after refractive corneal surgery
    Camellin, M
    Calossi, A
    JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (02) : 187 - 199
  • [6] Accuracy of 14 intraocular lens power calculation formulas in extremely long eyes
    Li, Xinxin
    Song, Chunyuan
    Wang, Yong
    Wang, Jing
    Tang, Qiongyan
    Wu, Zheming
    Zhou, Yanwen
    Sun, Juan
    Jia, Yanhong
    Lin, Zhenlin
    Li, Shaowei
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2024, 262 (11) : 3619 - 3628
  • [7] Comparison of the accuracy of intraocular lens power calculation formulas for eyes after corneal refractive surgery
    Wang, Xiao-Zhen
    Cui, Rui
    Song, Xu-Dong
    Yun, Bo
    Qian, Jin
    Ding, Ning
    ANNALS OF TRANSLATIONAL MEDICINE, 2020, 8 (14)
  • [8] Meta-analysis of accuracy of intraocular lens power calculation formulas in short eyes
    Wang, Qiwei
    Jiang, Wu
    Lin, Tiao
    Wu, Xiaohang
    Lin, Haotian
    Chen, Weirong
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2018, 46 (04): : 356 - 363
  • [9] Network Meta-analysis of Intraocular Lens Power Calculation Formula Accuracy in 1016 Eyes With Long Axial Length
    Ma, Yiyuan
    Xiong, Ruilin
    Liu, Zhenzhen
    Young, Charlotte aimee
    Wu, Yue
    Zheng, Danying
    Zhang, Xinyu
    Jin, Guangming
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2024, 257 : 178 - 186
  • [10] Intraocular lens power calculation errors in long eyes
    Olsen, Thomas
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (04): : 733 - 734