Reciprocating Torsional Fatigue and Mechanical Tests of Thermal-Treated Nickel Titanium Instruments

被引:2
|
作者
Vieira, Victor Talarico Leal [1 ,2 ,3 ]
Jaime, Alejandro [4 ]
Puente, Carlos Garcia [4 ]
Soimu, Giuliana [1 ]
Joa, Emmanuel
Silva, Nogueira Leal [1 ,3 ]
Elias, Carlos Nelson [2 ]
de Deus, Gustavo [3 ]
机构
[1] Grande Rio Univ UNIGRANRIO, Sch Dent, Rio De Janeiro, Brazil
[2] Mil Inst Engn, Mat Sci Dept, Rio De Janeiro, RJ, Brazil
[3] Fluminense Fed Univ, Dept Endodont, Niteroi, Brazil
[4] Maimonides Univ UMAI, Dept Endodont, Buenos Aires, Argentina
关键词
Mechanical tests; NiTi instruments; reciprocating torsional test; thermal-treated NiTi; CYCLIC FATIGUE; ENDODONTIC PREPARATION; MOVEMENT KINEMATICS; ROTARY INSTRUMENTS; RESISTANCE; FILES;
D O I
10.1016/j.joen.2024.12.013
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: This study aimed to evaluate the mechanical properties of Reciproc, Reciproc Blue, One RECI, and R-Motion, specifically focusing on their flexibility and buckling resistance and to assess their torsional fatigue behavior under 2 different reciprocation angles (150 degrees/30 degrees or 70 degrees/30 degrees). Methods: A total of 160 instruments (40 per brand) were tested. Flexion and buckling tests were conducted using a Universal testing machine (DL 200 MF). For the flexion test, instruments were positioned at a 45 degrees angle and flexed toward the ground plane. In the buckling test, the load was applied along the instrument's axis. A 20 N load cell and a test speed of 15 mm/s were used for both tests. Torsion test was performed using a torsiometer. The test speed was 100 rpm and the instruments were driven by an endodontic motor in 150 degrees/30 degrees and 70 degrees/30 degrees reciprocating angles. One-way analysis of variance post hoc Tukey tests and Student's t test were used for statistical comparisons (a = 5%). Results: The instruments exhibited significant differences in flexibility (P < .05), with One RECI demonstrating the highest flexibility and Reciproc the lowest. In terms of buckling resistance, Reciproc showed the highest values, while One RECI had the lowest (P < .05). Reciproc and Reciproc Blue displayed greater torque to fracture compared to One RECI and R-Motion (P < .05). Reducing the reciprocation angle did not affect the torque to fracture (P > .05). However, the reduced angle of 70 degrees/30 degrees significantly increased the time to fracture for all instruments when compared to angle of 150 degrees/30 degrees (P < .05). Conclusions: The tested instruments demonstrated notable differences in their mechanical performance, highlighting variability in flexibility, buckling resistance, and torsional behavior. Reducing the reciprocation angle prolonged the time to fracture without affecting maximum torque, underscoring the influence of reciprocation kinematics on instrument performance. (J Endod 2025;51:359- 366.)
引用
收藏
页码:359 / 366
页数:8
相关论文
共 50 条
  • [1] Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments
    Murilo Priori Alcalde
    Marco Antonio Hungaro Duarte
    Clovis Monteiro Bramante
    Bruno Carvalho de Vasconselos
    Mario Tanomaru-Filho
    Juliane Maria Guerreiro-Tanomaru
    Jader Camilo Pinto
    Marcus Vinicius Reis Só
    Rodrigo Ricci Vivan
    Clinical Oral Investigations, 2018, 22 : 1865 - 1871
  • [2] Evaluation of cyclic and torsional fatigue resistance of several heat-treated reciprocating nickel-titanium instruments
    Weissheimer, Theodoro
    Duarte, Marco Antonio Hungaro
    So, Marcus Vinicius Reis
    da Rosa, Ricardo Abreu
    Espinosa, Michel Klymus
    Vivan, Rodrigo Ricci
    Alcalde, Murilo Priori
    AUSTRALIAN ENDODONTIC JOURNAL, 2023, 49 (03) : 524 - 529
  • [3] Correction to: Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments
    Murilo Priori Alcalde
    Marco Antonio Hungaro Duarte
    Clovis Monteiro Bramante
    Bruno Carvalho de Vasconselos
    Mario Tanomaru-Filho
    Juliane Maria Guerreiro-Tanomaru
    Jader Camilo Pinto
    Marcus Vinicius Reis Só
    Rodrigo Ricci Vivan
    Clinical Oral Investigations, 2018, 22 : 1879 - 1879
  • [4] Cyclic Fatigue and Torsional Resistance of Four Martensite-Based Nickel Titanium Reciprocating Instruments
    Silva, Emmanuel J. N. L.
    Lima, Carolina O.
    Vieira, Victor T. L.
    Antunes, Henrique S.
    Moreira, Edson J. L.
    Versiani, Marco A.
    EUROPEAN ENDODONTIC JOURNAL, 2020, 5 (03): : 231 - 235
  • [5] Torsional Fatigue Resistance of Blue-treated Reciprocating Instruments
    Nogueira Leal Silva, Emmanuel Joao
    Hecksher, Fernanda
    Antunes, Henrique dos Santos
    De-Deus, Gustavo
    Elias, Carlos Nelson
    Leal Vieira, Victor Talarico
    JOURNAL OF ENDODONTICS, 2018, 44 (06) : 1038 - 1041
  • [6] Torsional, Static, and Dynamic Cyclic Fatigue Resistance of Reciprocating and Continuous Rotating Nickel-Titanium Instruments
    Pedulla, Eugenio
    Kharouf, Naji
    Caruso, Simone
    La Rosa, Giusy Rita Maria
    Jmal, Hamdi
    Haikel, Youssef
    Mancino, Davide
    JOURNAL OF ENDODONTICS, 2022, 48 (11) : 1421 - 1427
  • [7] Dynamic Cyclic Fatigue Resistance of Heat-treated Nickel Titanium Instruments in Reciprocating Motion
    Sobral, Thais Kauana Magalhaes
    Piasecki, Lucila
    Tomazinho, Flavia Sens Fagundes
    Kirchhoff, Alison Luis
    Gabardo, Marilisa Carneiro Leao
    Mattos, Natanael Henrique Ribeiro
    Filho, Flares Baratto
    Fariniuk, Luiz Fernando
    EUROPEAN ENDODONTIC JOURNAL, 2023, 8 (03): : 201 - 206
  • [8] Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments (vol 22, pg 1879, 2018)
    Alcalde, Murilo Priori
    Hungaro Duarte, Marco Antonio
    Bramante, Clovis Monteiro
    de Vasconselos, Bruno Carvalho
    Tanomaru-Filho, Mario
    Guerreiro-Tanomaru, Juliane Maria
    Pinto, Jader Camilo
    Reis So, Marcus Vinicius
    Vivan, Rodrigo Ricci
    CLINICAL ORAL INVESTIGATIONS, 2018, 22 (04) : 1879 - 1879
  • [9] Cyclic Fatigue Resistance of Rotary and Reciprocating Nickel-Titanium Instruments Subjected to Static and Dynamic Tests
    Thu, Myint
    Ebihara, Arata
    Maki, Keiichiro
    Miki, Nishijo
    Okiji, Takashi
    JOURNAL OF ENDODONTICS, 2020, 46 (11) : 1752 - 1757
  • [10] A comparison of cyclic fatigue resistance in reciprocating and rotary nickel-titanium instruments
    Wan, Jeffrey
    Rasimick, Brian J.
    Musikant, Barry L.
    Deutsch, Allan S.
    AUSTRALIAN ENDODONTIC JOURNAL, 2011, 37 (03) : 122 - 127