Motion artifact variability in biomagnetic wearable devices

被引:0
|
作者
Arekhloo, Negin Ghahremani [1 ,2 ]
Wang, Huxi [1 ,2 ]
Parvizi, Hossein [2 ]
Tanwear, Asfand [1 ]
Zuo, Siming [1 ]
McKinlay, Michael [2 ]
Nunez, Carlos Garcia [2 ]
Nazarpour, Kianoush [1 ,3 ]
Heidari, Hadi [1 ,2 ]
机构
[1] Neuranics Ltd, Glasgow, Scotland
[2] Univ Glasgow, James Watt Sch Engn, Microelect Lab, Glasgow City, Scotland
[3] Univ Edinburgh, Sch Informat, Edinburgh, Scotland
来源
FRONTIERS IN MEDICAL TECHNOLOGY | 2024年 / 6卷
基金
英国工程与自然科学研究理事会;
关键词
biomagnetic measurements; motion artifacts; wearable sensors; gradient background field; homogeneous background field; SIGNAL; ECG;
D O I
10.3389/fmedt.2024.1457535
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Motion artifacts can be a significant noise source in biomagnetic measurements when magnetic sensors are not separated from the signal source. In ambient environments, motion artifacts can be up to ten times stronger than the desired signals, varying with environmental conditions. This study evaluates the variability of these artifacts and the effectiveness of a gradiometer in reducing them in such settings. To achieve these objectives, we first measured the single channel output in varying magnetic field conditions to observe the effect of homogeneous and gradient background fields. Our analysis revealed that the variability in motion artifact within an ambient environment is primarily influenced by the gradient magnetic field rather than the homogeneous one. Subsequently, we configured a gradiometer in parallel and vertical alignment with the direction of vibration (X-axis). Our findings indicated that in a gradient background magnetic field ranging from 1 nT/mm to 10 nT/mm, the single-channel sensor output exhibited a change of 164.97 pT per mm unit increase, while the gradiometer output showed a change of only 0.75 pT/mm within the same range. Upon repositioning the gradiometer vertically (Y direction), perpendicular to the direction of vibration, the single-channel output slope increased to 196.85 pT, whereas the gradiometer output only increased by 1.06 pT/mm for the same range. Our findings highlight the influence of ambient environments on motion artifacts and demonstrate the potential of gradiometers to mitigate these effects. In the future, we plan to record biomagnetic signals both inside and outside the shielded room to compare the efficacy of different gradiometer designs under varying environmental conditions.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Robust wireless capacitive ECG system with adaptive signal quality and motion artifact reduction
    Castro, I. D.
    Morariu, R.
    Torfs, T.
    Van Hoof, C.
    Puers, R.
    2016 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2016, : 585 - 590
  • [42] Use of power-line interference for adaptive motion artifact removal in biopotential measurements
    Xu, Lin
    Rooijakkers, Michael J.
    Rabotti, Chiara
    Peuscher, Jan
    Mischi, Massimo
    PHYSIOLOGICAL MEASUREMENT, 2016, 37 (01) : 25 - 40
  • [43] GUI Creation for Removal of Motion Artifact in PPG Signals
    Yadhuraj, S. R.
    Sudarshan, B. G.
    Kumar, Prasanna S. C.
    2016 3RD INTERNATIONAL CONFERENCE ON ADVANCED COMPUTING AND COMMUNICATION SYSTEMS (ICACCS), 2016,
  • [44] Evaluation of motion artifact metrics for coronary CT angiography
    Ma, Hongfeng
    Gros, Eric
    Szabo, Aniko
    Baginski, Scott G.
    Laste, Zachary R.
    Kulkarni, Naveen M.
    Okerlund, Darin
    Schmidt, Taly G.
    MEDICAL PHYSICS, 2018, 45 (02) : 687 - 702
  • [45] Wavelet Based Motion Artifact Removal for ECG Signals
    Hashim, Fakroul Ridzuan
    Soraghan, John
    Petropoulakis, Lykourgos
    Safie, Sairul Izwan
    2012 IEEE EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES), 2012,
  • [46] Effect of pressure and padding on motion artifact of textile electrodes
    Comert, Alper
    Honkala, Markku
    Hyttinen, Jari
    BIOMEDICAL ENGINEERING ONLINE, 2013, 12
  • [47] Effect of pressure and padding on motion artifact of textile electrodes
    Alper Cömert
    Markku Honkala
    Jari Hyttinen
    BioMedical Engineering OnLine, 12
  • [48] Effect of Pressure on Skin-Electrode Impedance in Wearable Biomedical Measurement Devices
    Taji, Bahareh
    Chan, Adrian D. C.
    Shirmohammadi, Shervin
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2018, 67 (08) : 1900 - 1912
  • [49] Soldier Safety and Performance through Wearable Devices
    Murdock, Richard C.
    Hagen, Joshua A.
    MICRO- AND NANOTECHNOLOGY SENSORS, SYSTEMS, AND APPLICATIONS X, 2018, 10639
  • [50] Wearable Devices for Cardiac Rhythm Diagnosis and Management
    Ip, James E.
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2019, 321 (04): : 337 - 338