Motion artifact variability in biomagnetic wearable devices

被引:0
|
作者
Arekhloo, Negin Ghahremani [1 ,2 ]
Wang, Huxi [1 ,2 ]
Parvizi, Hossein [2 ]
Tanwear, Asfand [1 ]
Zuo, Siming [1 ]
McKinlay, Michael [2 ]
Nunez, Carlos Garcia [2 ]
Nazarpour, Kianoush [1 ,3 ]
Heidari, Hadi [1 ,2 ]
机构
[1] Neuranics Ltd, Glasgow, Scotland
[2] Univ Glasgow, James Watt Sch Engn, Microelect Lab, Glasgow City, Scotland
[3] Univ Edinburgh, Sch Informat, Edinburgh, Scotland
来源
FRONTIERS IN MEDICAL TECHNOLOGY | 2024年 / 6卷
基金
英国工程与自然科学研究理事会;
关键词
biomagnetic measurements; motion artifacts; wearable sensors; gradient background field; homogeneous background field; SIGNAL; ECG;
D O I
10.3389/fmedt.2024.1457535
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Motion artifacts can be a significant noise source in biomagnetic measurements when magnetic sensors are not separated from the signal source. In ambient environments, motion artifacts can be up to ten times stronger than the desired signals, varying with environmental conditions. This study evaluates the variability of these artifacts and the effectiveness of a gradiometer in reducing them in such settings. To achieve these objectives, we first measured the single channel output in varying magnetic field conditions to observe the effect of homogeneous and gradient background fields. Our analysis revealed that the variability in motion artifact within an ambient environment is primarily influenced by the gradient magnetic field rather than the homogeneous one. Subsequently, we configured a gradiometer in parallel and vertical alignment with the direction of vibration (X-axis). Our findings indicated that in a gradient background magnetic field ranging from 1 nT/mm to 10 nT/mm, the single-channel sensor output exhibited a change of 164.97 pT per mm unit increase, while the gradiometer output showed a change of only 0.75 pT/mm within the same range. Upon repositioning the gradiometer vertically (Y direction), perpendicular to the direction of vibration, the single-channel output slope increased to 196.85 pT, whereas the gradiometer output only increased by 1.06 pT/mm for the same range. Our findings highlight the influence of ambient environments on motion artifacts and demonstrate the potential of gradiometers to mitigate these effects. In the future, we plan to record biomagnetic signals both inside and outside the shielded room to compare the efficacy of different gradiometer designs under varying environmental conditions.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Evaluation of Data Processing and Artifact Removal Approaches Used for Physiological Signals Captured Using Wearable Sensing Devices during Construction Tasks
    Anwer, Shahnawaz
    Li, Heng
    Antwi-Afari, Maxwell Fordjour
    Mirza, Aquil Maud
    Rahman, Mohammed Abdul
    Mehmood, Imran
    Guo, Runhao
    Wong, Arnold Yu Lok
    JOURNAL OF CONSTRUCTION ENGINEERING AND MANAGEMENT, 2024, 150 (01)
  • [32] Electrocardiogram Removal to Establish a Motion Artifact Database
    Kulpa, Jonathan
    Chan, Adrian D. C.
    2023 IEEE CANADIAN CONFERENCE ON ELECTRICAL AND COMPUTER ENGINEERING, CCECE, 2023,
  • [33] Electrocardiographic motion artifact versus electrode impedance
    Wiese, SR
    Anheier, P
    Connemara, RD
    Mollner, AT
    Neils, TE
    Kahn, JA
    Webster, JG
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (01) : 136 - 139
  • [34] Optimization of view ordering for motion artifact suppression
    Nguyen, TD
    Ding, GL
    Watts, R
    Wang, Y
    MAGNETIC RESONANCE IMAGING, 2001, 19 (07) : 951 - 957
  • [35] ADAPTIVE MOTION ARTIFACT FILTER FOR PHOTOPLETHYSMOGRAM EXTRACTION
    Abdallah, O.
    Tarazona, A. Piera
    Roca, T. Martinez
    Alam, K. Abo
    Bolz, A.
    BIODEVICES 2010: PROCEEDINGS OF THE THIRD INTERNATIONAL CONFERENCE ON BIOMEDICAL ELECTRONICS AND DEVICES, 2010, : 187 - 190
  • [36] Motion artifact contaminated multichannel EEG dataset
    Ahmed, Sheikh Farhana Binte
    Amin, Md. Ruhul
    Islam, Md. Kafiul
    DATA IN BRIEF, 2024, 57
  • [37] Portable microwave radiometer for wearable devices
    Vesnin, Sergey G.
    Sedankin, Mikhail K.
    Ovchinnikov, Lev M.
    Gudkov, Alexander G.
    Leushin, Vitaly Yu
    Sidorov, Igor A.
    Goryanin, Igor I.
    NEXO REVISTA CIENTIFICA, 2021, 34 (04): : 1431 - 1447
  • [38] Portable microwave radiometer for wearable devices
    Vesnin, Sergey G.
    Sedankin, Mikhail K.
    Ovchinnikov, Lev M.
    Gudkov, Alexander G.
    Leushin, Vitaly Yu
    Sidorov, Igor A.
    Goryanin, Igor I.
    SENSORS AND ACTUATORS A-PHYSICAL, 2021, 318
  • [39] Properties of different types of dry electrodes for wearable smart monitoring devices
    Popovic-Maneski, Lana
    Ivanovic, Marija D.
    Atanasoski, Vladimir
    Miletic, Marjan
    Zdolsek, Sanja
    Bojovic, Bosko
    Hadzievski, Ljupco
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2020, 65 (04): : 405 - 415
  • [40] Wearable Sensor Networks for Motion Capture
    Arsenault, Dennis
    Whitehead, Anthony
    PROCEEDINGS OF THE 2015 7TH INTERNATIONAL CONFERENCE ON INTELLIGENT TECHNOLOGIES FOR INTERACTIVE ENTERTAINMENT, 2015, : 158 - 167