Paul versus Baerveldt 350 glaucoma drainage implants: One-year comparative analysis

被引:0
|
作者
Oliver-Gutierrez, David [1 ,2 ]
Segura-Duch, Gloria [1 ,3 ]
avila-Marron, Elena [1 ]
Arciniegas-Perasso, Carlos A. [1 ]
Duch-Tuesta, Susana [1 ]
机构
[1] VERTE ICO Ophthalmol, Via Augusta 61, Barcelona 08006, Spain
[2] Univ Hosp Valle Hebron, Paseo Valle Hebron 119-129, Barcelona 08035, Spain
[3] Univ Autonoma Barcelona, Inst Univ Barraquer, Ctr Oftalmol Barraquer, C-Muntaner 314, Barcelona 08021, Spain
关键词
Baerveldt glaucoma implant; glaucoma drainage implants; Paul glaucoma implant; refractory glaucoma; tube surgery; VALVE IMPLANTATION; MITOMYCIN-C; SURGERY; OUTCOMES; TUBE;
D O I
10.4103/IJO.IJO_2595_24
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:This study aimed to evaluate and compare the effectiveness and safety of Paul glaucoma implant (PGI) and Baerveldt 350 glaucoma implant (BGI) over a 1-year follow-up period. Methods:This retrospective study was conducted in a private clinic. It compared 27 patients consecutively treated with the PGI to a historical cohort of 29 eyes that received the BGI prior to the introduction of the PGI between 2018 and 2023. Intervention: A new drainage device, the Paul implant, was placed in the anterior or posterior chamber of eyes with uncontrolled intraocular pressure (IOP). Main Outcome Measures:The main outcome measures were success rate, final mean IOP, IOP reduction percentage, medication reduction, and complications. Statistical analyses, including adjustments for confounders, were used to compare the performance of PGI and BGI over at least a 12-month period. Results:Significant reductions in IOP were observed in both groups (P < 0.001). At 1 year, no significant differences were found in mean IOP (BGI: 12.0 mmHg, SD 2.9; PGI: 11.2 mmHg, SD 6.0) or medication usage (P > 0.05). The failure rates were 7% for BGI and 18% for PGI, with complete success rates of 56% for BGI and 32% for PGI. Hypertensive phases occurred in 32% of the BGI cases and in 19% of the PGI cases. No significant differences in the complication rates or postoperative visual acuity were observed between the groups. Conclusion:PGI and BGI exhibited comparable efficacy and safety profiles after 1 year.
引用
收藏
页码:S317 / S323
页数:7
相关论文
共 50 条
  • [31] Ahmed Versus Baerveldt Glaucoma Drainage Implantation in Patients With Markedly Elevated Intraocular Pressure (≥30mm Hg)
    Resende, Arthur F.
    Moster, Marlene R.
    Patel, Neal S.
    Lee, Daniel
    Dhami, Hermandeep
    Pro, Michael J.
    Waisbourd, Michael
    JOURNAL OF GLAUCOMA, 2016, 25 (09) : 738 - 743
  • [32] One-year result of XEN45 implant for glaucoma: efficacy, safety, and postoperative management
    Tan, S. Z.
    Walkden, A.
    Au, L.
    EYE, 2018, 32 (02) : 324 - 332
  • [33] RETRACTED: Modified Trabeculectomy versus Glaucoma Drainage Implant Surgery: A Retrospective Comparative Study for Refractory Glaucoma Treatment (Retracted Article)
    He, Yuan
    He, Beilei
    Ji, Zhi
    Zhang, Ruixue
    Quan, Zhuoya
    Xie, Guijun
    Pu, Xiaoli
    OXIDATIVE MEDICINE AND CELLULAR LONGEVITY, 2022, 2022
  • [34] Short Dental Implants (≤8.5 mm) versus Standard Dental Implants (≥10 mm): A One-Year Post-Loading Prospective Observational Study
    Pardo-Zamora, Guillermo
    Ortiz-Ruiz, Antonio Jose
    Camacho-Alonso, Fabio
    Martinez-Marco, Jose Francisco
    Molina-Gonzalez, Juan Manuel
    Pique-Clusella, Nuria
    Vicente-Hernandez, Ascension
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2021, 18 (11)
  • [35] One-year outcomes of stand-alone ab externo SIBS microshunt implantation in refractory glaucoma
    Durr, Georges M.
    Schlenker, Matthew B.
    Samet, Saba
    Ahmed, Iqbal Ike K.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2022, 106 (01) : 71 - 79
  • [36] One-Year Outcomes of Micropulse Cyclophototherapy for Primary Open-angle Glaucoma
    Tong, Weihan
    Shen, Tina Yu Ting
    Wong, Hung Chew
    Aquino, Maria Cecilia
    Chew, Paul
    Lim, Dawn
    JOURNAL OF GLAUCOMA, 2021, 30 (10) : 911 - 920
  • [37] One-Year Outcomes of Standalone XEN Gel Stent Versus Nonpenetrating Deep Sclerectomy
    Touboul, Arnaud
    Fels, Audrey
    Morin, Adele
    Bensmail, Djawed
    Lachkar, Yves
    JOURNAL OF GLAUCOMA, 2022, 31 (12) : 955 - 965
  • [38] Efficacy of bleb-independent penetrating canaloplasty in primary angle-closure glaucoma: one-year results
    Zhang, Shaodan
    Hu, Cheng
    Cheng, Huanhuan
    Gu, Juan
    Samuel, Kwizera
    Lin, Haishuang
    Deng, Yuxuan
    Xie, Yanqian
    Hu, Jingjing
    Le, Rongrong
    Xu, Shuxia
    Tham, Clement C.
    Liang, Yuanbo
    ACTA OPHTHALMOLOGICA, 2022, 100 (01) : E213 - E220
  • [39] Comparative effectiveness of hysterectomy versus myomectomy on one-year health-related quality of life in women with uterine fibroids
    Wallace, Kedra
    Zhang, Shuaiqi
    Thomas, Laine
    Stewart, Elizabeth A.
    Nicholson, Wanda Kay
    Wegienka, Ganesa R.
    Wise, Lauren A.
    Laughlin-Tommaso, Shannon K.
    Diamond, Michael P.
    Marsh, Erica E.
    Jacoby, Vanessa L.
    Anchan, Raymond M.
    Venable, Sateria
    Larry, G. Maxwell
    Lytle, Barbara
    Wang, Tracy
    Myers, Evan R.
    FERTILITY AND STERILITY, 2020, 113 (03) : 618 - 626
  • [40] One-year versus five-year hospital readmission after ischemic stroke and TIA
    Bjerkreim, Anna Therese
    Naess, Halvor
    Khanevski, Andrej Netland
    Thomassen, Lars
    Waje-Andreassen, Ulrike
    Logallo, Nicola
    BMC NEUROLOGY, 2019, 19 (1)