Comparisons of Fifth-, Sixth-, and Seventh-Generation Continuous Glucose Monitoring Systems

被引:14
|
作者
Welsh, John B. [1 ]
Psavko, Simon [1 ]
Zhang, Xiaohe [1 ]
Gao, Peggy [1 ]
Balo, Andrew K. [1 ]
机构
[1] Dexcom Inc, 6340 Sequence Dr, San Diego, CA 92121 USA
来源
JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY | 2024年 / 18卷 / 01期
关键词
accuracy; continuous glucose monitoring; Dexcom; G7; human factors; DAILY INSULIN INJECTIONS; REAL-TIME; GLYCEMIC CONTROL; YOUNG-ADULTS; HYPOGLYCEMIA; PERFORMANCE; MULTICENTER; ADOLESCENTS;
D O I
10.1177/19322968221099879
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Between-system differences for continuous glucose monitoring (CGM) devices have important clinical consequences.Purpose: Here we review attributes of Dexcom's fifth-, sixth-, and seventh-generation (G5, G6, and G7) CGM systems.Methods: Accuracy metrics were derived from preapproval trials of the three systems and compared after propensity score adjustments were used to balance baseline demographic characteristics. Metrics included mean absolute relative differences (MARD) between CGM and YSI values and the proportion of CGM values within 20% or 20 mg/dL of the YSI values ("%20/20"). Ease-of-use was evaluated by formal task analysis.Conclusions: Adjusted MARD and %20/20 agreement rates were 9.0%/93.1% (abdomen-placed G5), 9.9%/92.3% (abdomen-placed G6), 9.1%/93.2% (abdomen-placed G7), and 8.2%/95.3% (arm-placed G7). Task analysis favored G7 over earlier systems. Favorable clinical outcomes such as hemoglobin A1c reduction and hypoglycemia avoidance seen with G5 and G6 are anticipated with G7 use.
引用
收藏
页码:143 / 147
页数:5
相关论文
共 50 条
  • [41] Time lag characterization of two continuous glucose monitoring systems
    Garg, Satish K.
    Voelmle, Mary
    Gottlieb, Peter A.
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2010, 87 (03) : 348 - 353
  • [42] Time lag differences of two continuous glucose monitoring systems
    Garg, S. K.
    Voelmle, M.
    Gottlieb, P. A.
    DIABETOLOGIA, 2009, 52 : S364 - S364
  • [43] Feasibility of a Glucose Manipulation Procedure for the Standardized Performance Evaluation of Continuous Glucose Monitoring Systems
    Link, Manuela
    Eichenlaub, Manuel
    Waldenmaier, Delia
    Wehrstedt, Stephanie
    Pleus, Stefan
    Jendrike, Nina
    Oter, Sukru
    Haug, Cornelia
    Hossmann, Stefanie
    Rothenbuhler, Martina
    Brandt, Derek
    Freckmann, Guido
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2025,
  • [44] Prediction Quality of Glucose Trend Indicators in Two Continuous Tissue Glucose Monitoring Systems
    Freckmann, Guido
    Link, Manuela
    Westhoff, Antje
    Kamecke, Ulrike
    Pleus, Stefan
    Haug, Cornelia
    DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 (08) : 550 - 556
  • [45] Accuracy of the Third Generation of a 14-Day Continuous Glucose Monitoring System
    Alva, Shridhara
    Brazg, Ronald
    Castorino, Kristin
    Kipnes, Mark
    Liljenquist, David R.
    Liu, Hanqing
    DIABETES THERAPY, 2023, 14 (04) : 767 - 776
  • [46] Accuracy of the Third Generation of a 14-Day Continuous Glucose Monitoring System
    Shridhara Alva
    Ronald Brazg
    Kristin Castorino
    Mark Kipnes
    David R. Liljenquist
    Hanqing Liu
    Diabetes Therapy, 2023, 14 : 767 - 776
  • [47] Comparing Continuous Glucose Monitoring Systems in General Population: Case Report
    Gosak, Lucija
    Svensek, Adrijana
    Stiglic, Gregor
    PROCEEDINGS OF THE 2024 9TH INTERNATIONAL CONFERENCE ON INTELLIGENT INFORMATION TECHNOLOGY, ICIIT 2024, 2024, : 467 - 473
  • [48] Allergic contact dermatitis to continuous glucose monitoring systems and insulin patchpumps
    Kamann, S.
    Oppel, E.
    ALLERGOLOGIE, 2019, 42 (09) : 397 - 403
  • [49] Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems
    Garg, Satish K.
    Smith, James
    Beatson, Christie
    Lopez-Baca, Benita
    Voelmle, Mary
    Gottlieb, Peter A.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 (02) : 65 - 72
  • [50] Contact allergy to continuous glucose monitoring systems and insulin patch pumps
    Klein, Anna
    Kamann, Stefanie
    Hoegg, Christof
    Oppel, Eva Maria
    HAUTARZT, 2020, 71 (SUPPL 1): : 41 - 43