Are previously validated blood pressure self-measurement devices accepted under the Universal Standard? A systematic review

被引:0
作者
de Oliveira, Estefanie Siqueira Vigato [1 ]
de Albuquerque, Nila Larisse Silva [1 ]
Dordetto, Priscila Rangel [1 ]
Lamas, Jose Luiz Tatagiba [1 ]
机构
[1] Univ Campinas UNICAMP, Sch Nursing, Campinas, SP, Brazil
关键词
arterial pressure; auscultation; blood pressure; blood pressure determination; blood pressure monitors; hypertension; oscillometry; protocols; sphygmomanometers; validation study; MEDICAL INSTRUMENTATION/EUROPEAN SOCIETY; HYPERTENSION/INTERNATIONAL ORGANIZATION; ADVANCEMENT; ASSOCIATION; GUIDANCE;
D O I
10.1097/HJH.0000000000003859
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
This study aimed to analyze whether oscillometric blood pressure devices validated for the general population may be considered approved under Universal Standard criteria. A systematic review was conducted, with searches in nine databases, up to September 2023, including 32 validation studies of noninvasive arm cuff devices for self-measurement. The British Hypertension Society protocol was most common (68%), followed by the Association for the Advancement of Medical Instrumentation (40%). Most devices met Universal Standard criterion 1, but only 17 (53%) met criterion 2. Few studies contained details about the choice of cuffs, the number of participants by arm circumference, or the differences between methods by cuff subgroup. Due to the considerable differences between validation protocols, 53% of the devices analyzed were approved under the Universal Standard. The study contributes to expanding the validated pool of self-measurement devices under the Universal Standard.
引用
收藏
页码:35 / 47
页数:13
相关论文
共 21 条
[1]   Evaluating medical tests: introducing the Cochrane Handbook for Systematic Reviews of Diagnostic Test Accuracy [J].
Bossuyt, Patrick M. ;
Deeks, Jonathan J. ;
Leeflang, Mariska M. ;
Takwoingi, Yemisi ;
Flemyng, Ella .
COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2023, (07)
[2]   Blood pressure measurement device selection in low-resource settings: Challenges, compromises, and routes to progress [J].
Brady, Tammy M. ;
Padwal, Raj ;
Blakeman, Drew E. ;
Farrell, Margaret ;
Frieden, Thomas R. ;
Kaur, Prabhdeep ;
Moran, Andrew E. ;
Jaffe, Marc G. .
JOURNAL OF CLINICAL HYPERTENSION, 2020, 22 (05) :792-801
[3]   History and Justification of a National Blood Pressure Measurement Validated Device Listing [J].
Cohen, Jordana B. ;
Padwal, Raj S. ;
Gutkin, Michael ;
Green, Beverly B. ;
Bloch, Michael J. ;
Germino, F. Wilford ;
Sica, Domenic A. ;
Viera, Anthony J. ;
Bluml, Benjamin M. ;
White, William B. ;
Taler, Sandra J. ;
Yarows, Steven ;
Shimbo, Daichi ;
Townsend, Raymond R. .
HYPERTENSION, 2019, 73 (02) :258-264
[4]  
Com I, 2020, WHO technical specifications for automated non-invasive blood pressure measuring devices with cuff WHO medical device technical series Internet
[5]   Wearable cuffless blood pressure monitoring devices: a systematic review and meta-analysis [J].
Islam, Sheikh Mohammed Shariful ;
Chow, Clara K. ;
Daryabeygikhotbehsara, Reza ;
Subedi, Narayan ;
Rawstorn, Jonathan ;
Tegegne, Teketo ;
Karmakar, Chandan ;
Siddiqui, Muhammad U. ;
Lambert, Gavin ;
Maddison, Ralph .
EUROPEAN HEART JOURNAL - DIGITAL HEALTH, 2022, 3 (02) :323-337
[6]   The 2020 "WHO Technical Specifications for Automated Non-Invasive Blood Pressure Measuring Devices With Cuff" [J].
John, Oommen ;
Campbell, Norm R. C. ;
Brady, Tammy M. ;
Farrell, Margret ;
Varghese, Cherian ;
Berumen, Adriana Velazquez ;
Gaitan, Laura A. Velez Ruiz ;
Toffelmire, Nicola ;
Ameel, Mohammad ;
Mideksa, Mulugeta ;
Jaffe, Marc G. ;
Schutte, Aletta E. ;
Khan, Taskeen ;
Lopez Meneses, Laura Patricia .
HYPERTENSION, 2021, 77 (03) :806-812
[7]   Validation of Blood Pressure Devices as per 2020 World Health Organization Technical Specifications: A Scoping Review of Global Literature [J].
Murthy, Shruti ;
Rajwar, Eti ;
Johnson, Rosemol ;
Shrivastava, Krati ;
Kamath, Prashanthi ;
Rahman, Rinsy ;
Devaria, Abhinav ;
Ismavel, Vijay Anand ;
Vetter, Beatrice N. ;
Jacob, Anil G. ;
John, Oommen .
HYPERTENSION, 2023, 80 (05) :1110-1116
[8]   The quest for accuracy of blood pressure measuring devices [J].
O'Brien, Eoin ;
Stergiou, George S. ;
Turner, Martin J. .
JOURNAL OF CLINICAL HYPERTENSION, 2018, 20 (07) :1092-1095
[9]  
Page MJ, 2021, BMJ-BRIT MED J, V372, DOI [10.1136/bmj.n160, 10.1136/bmj.n71]
[10]  
Pan American Health Organization, 2021, HEARTS in the Americas Regulatory Pathway to the Exclusive Use of Validated Blood Pressure Measuring Devices Internet. HEARTS in the Americas Regulatory Pathway to the Exclusive Use of Validated Blood Pressure Measuring Devices