Impact of temperature on ophthalmic viscosurgical devices and clinical implications: a pilot study

被引:0
|
作者
Henry, Caoimhe [1 ]
Castro, Alejandro Thompson [2 ]
Belkessa, Mohamed [3 ]
Rycroft, Ewan [2 ]
Syntouka, Ioanna M. [2 ]
Silva, Andreia F. [4 ,5 ]
Sillars, Fiona [3 ]
Oliveira, Monica S. N. [2 ]
Lockington, David [1 ]
机构
[1] Gartnavel Royal Hosp, Tennent Inst Ophthalmol, 1053 Great Western Rd, Glasgow G12 0YN, Scotland
[2] Univ Strathclyde, Dept Mech & Aerosp Engn, James Weir Fluids Lab, Glasgow, Scotland
[3] Univ Strathclyde, Adv Mat Res Lab, Glasgow, Scotland
[4] Univ Edinburgh, Edinburgh Complex Fluids Partnership ECFP, SUPA, Edinburgh, Scotland
[5] Univ Edinburgh, Sch Phys & Astron, Edinburgh, Scotland
来源
关键词
CATARACT-SURGERY; BEHAVIOR; RHEOLOGY; FLOW;
D O I
10.1097/j.jcrs.0000000000001513
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate how temperature affects the rheology of common ophthalmic viscosurgical devices (OVDs) and clinical implications. Setting: Tennent Institute of Ophthalmology, Glasgow, with Department of Mechanical and Aerospace Engineering, and Advanced Materials Research Laboratory, University of Strathclyde, Glasgow, United Kingdom. Design: Laboratory pilot study. Methods: The viscous and elastic responses of 3 OVDs (Eyefill-SC/Eyefill-C/Eyefill-HD) were measured using rotational and extensional rheometers at clinically relevant temperatures (5 degrees C, 25 degrees C, 37 degrees C). Thermal properties were evaluated using differential scanning calorimetry and laser-flash analysis. Results: The OVDs tested exhibited viscoelastic properties and shear-thinning behavior. Apparent viscosities and relaxation time were higher at lower temperatures. The Eyefill-C and Eyefill-SC exhibited predominantly viscous character at low frequencies with a transition to predominantly elastic behavior at high frequencies. An increase in temperature led to a decrease in relaxation time under shear and extension. At low frequencies, Eyefill-C and Eyefill-SC moduli increase with decreasing temperatures. Eyefill-HD at 25 degrees C and 37 degrees C displays 2 crossover points, with the storage modulus dominating at low and high frequencies indicating a predominantly elastic behavior. Thermal property analysis revealed Eyefill-C had the lowest thermal conductivity. Conclusions: This pilot study confirms our clinical experience that OVD properties are affected by low temperatures, with increased viscosities at low shear rates and higher relaxation times. Cold OVD can cause greater resistance to initiation of intraocular lens (IOL) injection system forces (compared with warmer OVD). Excessively forced injection using cold OVD could contribute to inadvertent cannula detachment if incorrectly assembled, or uncontrolled IOL release leading to avoidable injury.
引用
收藏
页码:1173 / 1179
页数:7
相关论文
共 50 条
  • [1] Impact of Ophthalmic Viscosurgical Devices in Cataract Surgery
    Malvankar-Mehta, Monali S.
    Fu, Angel
    Subramanian, Yasoda
    Hutnik, Cindy
    JOURNAL OF OPHTHALMOLOGY, 2020, 2020
  • [2] Ophthalmic viscosurgical devices
    Bissen-Miyajima, Hiroko
    CURRENT OPINION IN OPHTHALMOLOGY, 2008, 19 (01) : 50 - 54
  • [3] Endotoxins in ophthalmic viscosurgical devices
    Dick, HB
    Augustin, AJ
    Pakula, T
    Pfeiffer, N
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2003, 13 (02) : 176 - 184
  • [4] New terminology: Ophthalmic viscosurgical devices
    Arshinoff, S
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (05): : 627 - 628
  • [5] Prefilled syringes and usability of ophthalmic viscosurgical devices
    Shiba, Takuya
    Tsuneoka, Hiroshi
    CLINICAL OPHTHALMOLOGY, 2014, 8 : 1697 - 1702
  • [6] Silicone oil bubbles in ophthalmic viscosurgical devices
    Ohrstrom, A
    Behndig, A
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (03): : 389 - 389
  • [7] Detecting Endotoxin Contamination of Ophthalmic Viscosurgical Devices
    Buchen, Shelley Y.
    Calogero, Don
    Hilmantel, Gene
    Eydelman, Malvina B.
    OPHTHALMOLOGY, 2012, 119 (07) : E11 - E18
  • [8] Silicone oil bubbles in ophthalmic viscosurgical devices
    Lanzl, IM
    Merté, RL
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (12): : 2069 - 2069
  • [9] Influence of ophthalmic viscosurgical devices on intraoperative aberrometry
    Masket, Samuel
    Fram, Nicole R.
    Holladay, Jack T.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (07): : 990 - 994
  • [10] New classification of ophthalmic viscosurgical devices - 2005
    Arshinoff, SA
    Jafari, M
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (11): : 2167 - 2171