摘要
Background:Identifying plans at risk of patient-specific quality assurance (PSQA) failure through complexity metrics can reduce the workload while maintaining quality. This study evaluates complexity metrics as predictors of PSQA outcomes.Materials and Methods:A retrospective analysis was conducted on 192 IMRT plans for head-and-neck cancer. Complexity metrics were calculated using an in-house Python program. PSQA was performed with 3%/2-mm gamma passing rate (GPR) criteria, with plans classified as "Pass" (GPR >= 95%) or "Fail." Statistical analyses, including Spearman's correlation and receiver operating characteristic analysis, assessed the metrics' predictive value.Results:Passing plans had an average GPR of 98.64 +/- 1.33%, compared to 92.17 +/- 2.35% for failing plans. The mean small area segment (MSAS) 5mm metric, with a threshold of 0.085, achieved a true positive rate of 38.17% and a false positive rate of 3.1%. Beam modulation and beam area indices also significantly differed between passing and failing plans.Conclusion:MSAS5 and edge metrics showed strong potential for identifying high-risk plans. These metrics can guide targeted PSQA, improving workflow efficiency without compromising treatment safety.
引用
收藏
页码:46 / 54
页数:9
相关论文
共 50 条
  • [21] Optimization and quality assurance of an image-guided radiation therapy system for intensity-modulated radiation therapy radiotherapy
    Tsai, Jen-San
    Micaily, Bizhan
    Miyamoto, Curtis
    MEDICAL DOSIMETRY, 2012, 37 (03) : 321 - 333
  • [22] Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the radiation therapy oncology group
    Molineu, A
    Followill, DS
    Balter, PA
    Hanson, WF
    Gillin, MT
    Huq, MS
    Eisbruch, A
    Ibbott, GS
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : 577 - 583
  • [23] Patient-specific quality control for intensity-modulated radiation therapy and volumetric-modulated arc therapy using electronic portal imaging device and two-dimensional ion chamber array
    Kausar, Abu
    Mani, Karthick R.
    Azhari, Hasin A.
    Zakaria, Golam A.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (01) : 26 - 31
  • [24] Integrating plan complexity and dosiomics features with deep learning in patient-specific quality assurance for volumetric modulated arc therapy
    Ce Han
    Ji Zhang
    Bing Yu
    Haoze Zheng
    Yibo Wu
    Zhixi Lin
    Boda Ning
    Jinling Yi
    Congying Xie
    Xiance Jin
    Radiation Oncology, 18
  • [25] A quality control method for intensity-modulated radiation therapy planning based on generalized equivalent uniform dose
    Pang, Haowen
    Sun, Xiaoyang
    Yang, Bo
    Wu, Jingbo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (03): : 276 - 282
  • [26] An Ensemble Approach to Knowledge-Based Intensity-Modulated Radiation Therapy Planning
    Zhang, Jiahan
    Wu, Q. Jackie
    Xie, Tianyi
    Sheng, Yang
    Yin, Fang-Fang
    Ge, Yaorong
    FRONTIERS IN ONCOLOGY, 2018, 8
  • [27] Quality assurance-based optimization (QAO): Towards improving patient-specific quality assurance in volumetric modulated arc therapy plans using machine learning
    Wall, Phillip D. H.
    Fontenot, Jonas D.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 87 : 136 - 143
  • [28] Does fluence smoothing reduce the complexity of the intensity-modulated radiation therapy treatment plan? A dosimetric analysis
    Saroj, Dinesh Kumar
    Yadav, Suresh
    Paliwal, Neetu
    JOURNAL OF MEDICAL PHYSICS, 2022, 47 (04) : 336 - 343
  • [29] Patient Specification Quality Assurance for Glioblastoma Multiforme Brain Tumors Treated with Intensity Modulated Radiation Therapy
    Al-Mohammed, H. I.
    INTERNATIONAL JOURNAL OF MEDICAL SCIENCES, 2011, 8 (06): : 461 - 466
  • [30] Evaluation criteria for film based intensity modulated radiation therapy quality assurance
    Anjum, Muhammad Naeem
    Parker, William
    Ruo, Russell
    Afzal, Muhammad
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2010, 26 (01): : 38 - 43