No Difference in Survivorship Following Primary Total Hip Arthroplasty Using Ultraporous Acetabular Cups and Cross-Linked Polyethylene With and Without Acetabular Screws in a Large United States Health Care System

被引:0
作者
Chen, Foster [1 ]
Chan, Priscilla H. [2 ]
Prentice, Heather A. [2 ]
Paxton, Elizabeth W. [2 ]
Fasig, Brian H. [2 ]
Sheth, Dhiren S. [3 ]
机构
[1] Washington Permanente Med Grp, Dept Orthopaed Surg, 125 16th Ave E, Seattle, WA 98112 USA
[2] Kaiser Permanente, Med Device Surveillance & Assessment, San Diego, CA USA
[3] Southern Calif Permanente Med Grp, Dept Orthopaed Surg, Irvine, CA USA
关键词
total hip arthroplasty; acetabular cups; screws; registry; aseptic loosening; periprosthetic fracture; POROUS METAL; OSTEOLYSIS; COMPONENTS; MICROMOTION;
D O I
10.1016/j.arth.2024.10.019
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Modern cementless acetabular cups for total hip arthroplasty (THA) typically have screw options. Historically, screws were thought to improve stability, but came at the cost of pathways for osteolysis. Modern cups and liners may have made both concerns obsolete, and the utility of screws are now contested. We sought to determine modern implant survivorship relative to screw use. Methods: We conducted a cohort study. A US health care system's Total Joint Replacement Registry was used to identify patients > 18 years who underwent uncomplicated primary THA for osteoarthritis (2010 to 2021) with an ultraporous cup and cross-linked polyethylene liner, with or without one to two acetabular screws. The primary outcome was acetabular revision for aseptic loosening. Secondary outcomes were aseptic revision for acetabular fracture and any revision for acetabular/femoral loosening and periprosthetic fracture. Multiple Cox proportional hazard regression was used to evaluate revision risk. There were 46,785 THAs identified. Screw use declined from 65.3 to 49.9%. Results: No difference was observed in 10-year revision risk for acetabular loosening (0.2 versus 0.1%, hazard ratio 1.97, 95% confidence interval = 0.84 to 4.59, P = 0.119). There was one revision for acetabular fracture with and three revisions without screws. There was no difference in risk of overall acetabular or femoral revision, loosening, or periprosthetic fracture. There remained no difference in acetabular-sided loosening between routine screw users and nonusers (0.15 versus 0.06%, hazard ratio 1.26, 95% confidence interval 0.42 to 3.75, P = 0.683). Conclusions: In this study of survivorship following routine uncomplicated primary THA with modern cups and liners, screw usage patterns were associated with neither an advantage nor disadvantage- neither screw usage nor avoidance was associated with differences in acetabular loosening revision risk. Screw use was not associated with harm but remains debatable if there is an added benefit. Level of evidence: Level III. (c) 2024 Elsevier Inc. All rights are reserved, including those for text and data mining, AI training, and similar technologies.
引用
收藏
页码:986 / 991.e2
页数:8
相关论文
共 30 条
[1]  
Abdelaal MS, 2021, SURG TECHNOL INT, V38, DOI 10.52198/21.STI.38.OS1398
[2]  
American Joint Replacement Registry (AJRR), 2023, Annual Report
[3]  
Banerjee S, 2013, HIP INT, V23, P509, DOI 10.530/hipint.5000064
[4]  
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[5]   The use of acetabular screws in total hip arthroplasty and its influence on wear and periacetabular osteolysis in the long-term follow-up [J].
Brodt, Steffen ;
Bischoff, Kathleen ;
Schulze, Marcel ;
Nowack, Dimitri ;
Roth, Andreas ;
Matziolis, Georg .
INTERNATIONAL ORTHOPAEDICS, 2022, 46 (04) :717-722
[6]   Impact of Acetabular Implant Design on Aseptic Failure in Total Hip Arthroplasty [J].
Coden, Gloria ;
Matzko, Chelsea ;
Hushmendy, Shazaan ;
Macaulay, William ;
Hepinstall, Matthew .
ARTHROPLASTY TODAY, 2021, 7 :60-68
[7]  
Curvo, 2023, Orthopaedic Network News, V9
[8]   Wear, Osteolysis, and Aseptic Loosening Following Total Hip Arthroplasty in Young Patients with Highly Cross-Linked Polyethylene: A Review of Studies with a Follow-Up of over 15 Years [J].
Deans, Christopher F. ;
Buckner, Brandt C. ;
Garvin, Kevin L. .
JOURNAL OF CLINICAL MEDICINE, 2023, 12 (20)
[9]   Midterm Outcomes of Revision Total Hip Arthroplasty With the Use of a Multihole Highly-Porous Titanium Shell [J].
Delanois, Ronald E. ;
Gwam, Chukwuweike U. ;
Mohamed, Nequesha ;
Khlopas, Anton ;
Chughtai, Morad ;
Malkani, Arthur L. ;
Mont, Michael A. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (09) :2806-2809
[10]   The Otto Aufranc Award -: Highly cross-linked polyethylene in Total Hip Arthroplasty -: Randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis [J].
Digas, G ;
Kärrholm, J ;
Thanner, J ;
Malchau, H ;
Herberts, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :6-16