Effectiveness of Remote Patient Monitoring Equipped With an Early Warning System in Tertiary Care Hospital Wards: Retrospective Cohort Study

被引:1
作者
Lakshman, Pavithra [1 ]
Gopal, Priyanka T. [1 ]
Khurdi, Sheen [1 ]
机构
[1] Ramaiah Mem Hosp, Hosp Adm, New BEL Rd, Bengaluru 560054, Karnataka, India
关键词
continuous vitals monitoring; remote patient monitoring; early warning system; hospital wards; retrospective; cohort study; early deterioration monitoring; patient care; decision making; clinical information; SCORE MEWS; INTERMITTENT; IMPACT; UNIT; DETERIORATION;
D O I
10.2196/56463
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Monitoring vital signs in hospitalized patients is crucial for evaluating their clinical condition. While early warning scores like the modified early warning score (MEWS) are typically calculated 3 to 4 times daily through spot checks, they might not promptly identify early deterioration. Leveraging technologies that provide continuous monitoring of vital signs, combined with an early warning system, has the potential to identify clinical deterioration sooner. This approach empowers health care providers to intervene promptly and effectively. Objective: This study aimed to assess the impact of a Remote Patient Monitoring System (RPMS) with an automated early warning system (R-EWS) on patient safety in noncritical care at a tertiary hospital. R-EWS performance was compared with a simulated Modified Early Warning System (S-MEWS) and a simulated threshold-based alert system (S-Threshold). Methods: Patient outcomes, including intensive care unit (ICU) transfers due to deterioration and discharges for nondeteriorating cases, were analyzed in Ramaiah Memorial Hospital's general wards with RPMS. Sensitivity, specificity, chi-square test for alert frequency distribution equality, and the average time from the first alert to ICU transfer in the last 24 hours was determined. Alert and patient distribution by tiers and vitals in R-EWS groups were examined. Results: Analyzing 905 patients, including 38 with deteriorations, R-EWS, S-Threshold, and S-MEWS generated more alerts for deteriorating cases. R-EWS showed high sensitivity (97.37%) and low specificity (23.41%), S-Threshold had perfect sensitivity (100%) but low specificity (0.46%), and S-MEWS demonstrated moderate sensitivity (47.37%) and high specificity (81.31%). The average time from initial alert to clinical deterioration was at least 18 hours for RPMS and S-Threshold in deteriorating participants. R-EWS had increased alert frequency and a higher proportion of critical alerts for deteriorating cases. Conclusions: This study underscores R-EWS role in early deterioration detection, emphasizing timely interventions for improved patient outcomes. Continuous monitoring enhances patient safety and optimizes care quality.
引用
收藏
页数:14
相关论文
共 38 条
[1]   The future of postoperative vital sign monitoring in general wards: improving patient safety through continuous artificial intelligence-enabled alert formation and reduction [J].
Aasvang, Eske K. ;
Meyhoff, Christian S. .
CURRENT OPINION IN ANESTHESIOLOGY, 2023, 36 (06) :683-690
[2]   Modified Early Warning Score (MEWS) as a predictor of intensive care unit admission in cancer patient on chemotherapy with positive blood culture: A retrospective cohort study [J].
Allarakia, Jawad ;
Felemban, Taher ;
Alghamdi, Amer ;
Ashi, Abdullah ;
Al Talhi, Yousef M. ;
Alsahafi, Ashraf ;
Alamri, Abdulfatah ;
Aldabbagh, Mona .
JOURNAL OF INFECTION AND PUBLIC HEALTH, 2023, 16 (06) :865-869
[3]   Utopia versus dystopia: Professional perspectives on the impact of healthcare artificial intelligence on clinical roles and skills [J].
Aquino, Yves Saint James ;
Rogers, Wendy A. ;
Braunack-Mayer, Annette ;
Frazer, Helen ;
Win, Khin Than ;
Houssami, Nehmat ;
Degeling, Christopher ;
Semsarian, Christopher ;
Carter, Stacy M. .
INTERNATIONAL JOURNAL OF MEDICAL INFORMATICS, 2023, 169
[4]   A controlled trial of electronic automated advisory vital signs monitoring in general hospital wards [J].
Bellomo, Rinaldo ;
Ackerman, Michael ;
Bailey, Michael ;
Beale, Richard ;
Clancy, Greg ;
Danesh, Valerie ;
Hvarfner, Andreas ;
Jimenez, Edgar ;
Konrad, David ;
Lecardo, Michele ;
Pattee, Kimberly S. ;
Ritchie, Josephine ;
Sherman, Kathie ;
Tangkau, Peter .
CRITICAL CARE MEDICINE, 2012, 40 (08) :2349-2361
[5]   Continuous Monitoring in an Inpatient Medical-Surgical Unit: A Controlled Clinical Trial [J].
Brown, Harvey ;
Terrence, Jamie ;
Vasquez, Patricia ;
Bates, David W. ;
Zimlichman, Eyal .
AMERICAN JOURNAL OF MEDICINE, 2014, 127 (03) :226-232
[6]   Effectiveness of continuous or intermittent vital signs monitoring in preventing adverse events on general wards: a systematic review and meta-analysis [J].
Cardona-Morrell, M. ;
Prgomet, M. ;
Turner, R. M. ;
Nicholson, M. ;
Hillman, K. .
INTERNATIONAL JOURNAL OF CLINICAL PRACTICE, 2016, 70 (10) :806-824
[7]   Cost of hospital services in India: a multi-site study to inform provider payment rates and Health Technology Assessment [J].
Chauhan, Akashdeep Singh ;
Guinness, Lorna ;
Bahuguna, Pankaj ;
Singh, Maninder Pal ;
Aggarwal, Vipul ;
Rajsekhar, Kavitha ;
Tripathi, Surbhi ;
Prinja, Shankar .
BMC HEALTH SERVICES RESEARCH, 2022, 22 (01)
[8]   Association between intensive care unit transfer delay and hospital mortality: A multicenter investigation [J].
Churpek, Matthew M. ;
Wendlandt, Blair ;
Zadravecz, Frank J. ;
Adhikari, Richa ;
Winslow, Christopher ;
Edelson, Dana P. .
JOURNAL OF HOSPITAL MEDICINE, 2016, 11 (11) :757-762
[9]   The impact of continuous versus intermittent vital signs monitoring in hospitals: A systematic review and narrative synthesis [J].
Downey, C. L. ;
Chapman, S. ;
Randell, R. ;
Brown, J. M. ;
Jayne, D. G. .
INTERNATIONAL JOURNAL OF NURSING STUDIES, 2018, 84 :19-27
[10]   Reducing ICU Utilization, Length of Stay, and Cost by Optimizing the Clinical Use of Continuous Monitoring System Technology in the Hospital [J].
Dykes, Patricia C. ;
Lowenthal, Graham ;
Lipsitz, Stuart ;
Salvucci, Suzanne M. ;
Yoon, Catherine ;
Bates, David W. ;
An, Perry G. .
AMERICAN JOURNAL OF MEDICINE, 2022, 135 (03) :337-+