Development and validation of a portable gas chromatograph method for quantitative determination of oxygen and toxic gas impurities in medical oxygen using planar microchromatographic columns and detectors

被引:0
|
作者
Galeeva, Ekaterina, V [1 ,3 ]
Galeev, Roman R. [1 ,2 ,3 ]
Sharma, Prachi [4 ,5 ]
Khokhlov, Alexander l. [3 ]
V. Somov, Dmitry [1 ]
Semanov, Dmitry A. [1 ,2 ]
Aryslanov, Ilshat R. [1 ]
Lezhnina, Natalia A. [1 ,3 ]
Platonov, Vladimir [5 ]
Tripathi, Nishant [5 ]
机构
[1] Recording & Anal Circulat Med Prod Roszdravnadzor, Informat & Methodol Ctr Expert Evaluat, 4 Bld 1,Slavyanskaya Sq, Moscow 109074, Russia
[2] Kazan Fed Univ, 18 Kremlevskaya St, Kazan 420008, Russia
[3] Yaroslavl State Med Univ, 5 St Revolutionary, Yaroslavl 150000, Yaroslavl, Russia
[4] Vellore Inst Technol VIT, Sch Elect Engn SENSE, Vellore 632014, Tamil Nadu, India
[5] Samara Natl Res Univ, 34 Moskovskoye Shosse, Samara 443086, Russia
关键词
oxygen; medical gases; gas chromatography; planar microchromatographic columns; quality control;
D O I
10.5806/AST.2024.37.5.261
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study examines portable Gas Chromatography (GC) for the quantitative analysis of oxygen and impurities, focusing on the development and validation of a method to determine oxygen, carbon monoxide, carbon dioxide, methane, and nitrogen in medical compressed oxygen gas. The goal is to ensure the quality of medicalgrade oxygen. The method's validation assessed its metrological characteristics, demonstrating specificity through clear chromatographic separation of the target gases and the absence of these peaks in the carrier gas chromatogram. It exhibited linearity within the designated concentration ranges, while precision met permissible standards, with the relative standard deviation for intermediate precision being less than 4% for carbon monoxide (0.00025 - 0.00099 %), less than 3 % for methane (0.0005 - 0.00246%) and carbon dioxide (0.0050 - 0.0150 %), less than 2% for nitrogen (0.1 - 0.7 %), and less than 0.01% for oxygen (99.27 - 99.98 %). Overall, the validation results confirm the suitability of this analytical method for the quantitative determination of the aforementioned gases in medical compressed oxygen using portable GC with microchromatographic columns and detectors.
引用
收藏
页码:261 / 270
页数:10
相关论文
共 9 条
  • [1] Development and Validation of a Gas-Chromatographic Method for Quantitative Determination of Phenol in Biologicals
    Kolesnikova, O. N.
    Runova, O. B.
    Ustinnikova, O. B.
    PHARMACEUTICAL CHEMISTRY JOURNAL, 2018, 52 (05) : 478 - 482
  • [2] Development and Validation of a Gas-Chromatographic Method for Quantitative Determination of Phenol in Biologicals
    O. N. Kolesnikova
    O. B. Runova
    O. B. Ustinnikova
    Pharmaceutical Chemistry Journal, 2018, 52 : 478 - 482
  • [3] Development and Validation of a Gas Chromatography Method for Quantitative Determination of Lavendar Oil in Lavaflam Preparation
    Aslanyan, M. A.
    Bobritskaya, L. A.
    Nazarova, E. S.
    Mirnaya, T. A.
    Zborovskaya, T. V.
    PHARMACEUTICAL CHEMISTRY JOURNAL, 2016, 50 (03) : 200 - 204
  • [4] Development and Validation of a Gas Chromatography Method for Quantitative Determination of Lavendar Oil in Lavaflam Preparation
    M. A. Aslanyan
    L. A. Bobritskaya
    E. S. Nazarova
    T. A. Mirnaya
    T. V. Zborovskaya
    Pharmaceutical Chemistry Journal, 2016, 50 : 200 - 204
  • [5] Development and Validation of an Automated Gas Chromatography Method for Determination of Dichloromethane in Ampicillin Sodium by Using Capillary Column Technology
    Trivedi, Manisha
    John, Elsy Raynil
    Ali, Faraat
    Prakash, Anuj
    Kumar, Robin
    Singh, Gyanendra Nath
    CURRENT PHARMACEUTICAL ANALYSIS, 2020, 16 (07) : 901 - 908
  • [6] Method development and validation for the determination of residual solvents in quinabut API by using gas chromatography. Message 2
    Golembiovska, Olena
    Voskoboinik, Oleksii
    Berest, Galina
    Kovalenko, Sergiy
    Logoyda, Liliya
    PHARMACIA, 2021, 68 (01) : 53 - 59
  • [7] Development and validation of a multiresidue method for the simultaneous determination of organophosphorus insecticides and their toxic metabolites in sugarcane juice and refined sugar by gas chromatography with flame photometric detection
    Ramasubramanian, Thirumalaiandi
    Paramasivam, Mariappan
    JOURNAL OF SEPARATION SCIENCE, 2016, 39 (11) : 2164 - 2171
  • [8] The development of new microelectrode gas sensors: An odyssey Part V. Simultaneous electrochemical determination of oxygen, carbon dioxide and nitrous oxide gas mixtures in a non-aqueous solvent using membrane shielded gold microelectrodes
    Floate, S
    Hahn, CEW
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2005, 583 (02) : 203 - 211
  • [9] Development and validation of a multi-residue method for the determination of pesticides in honeybees using acetonitrile-based extraction and gas chromatography-tandem quadrupole mass spectrometry
    Walorczyk, Stanislaw
    Gnusowski, Boguslaw