A multi-channel MRI console for ultra-high field up to 14 T

被引:0
|
作者
Lin, Kaisheng [1 ]
Wang, Miaotian [1 ]
Wang, Yaohui [2 ]
Li, Hongwei [3 ]
Wang, Weimin [1 ,4 ]
机构
[1] Peking Univ, Sch Elect, Beijing 100871, Peoples R China
[2] Chinese Acad Sci, Inst Elect Engn, Beijing 100190, Peoples R China
[3] Peking Univ, Coll Chem & Mol Engn, Beijing NMR Ctr, Beijing 100871, Peoples R China
[4] Peking Univ, Inst Biomed Engn, Shenzhen Grad Sch, Shenzhen 518055, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2025年 / 96卷 / 03期
基金
中国国家自然科学基金;
关键词
RESONANCE; GRADIENT;
D O I
10.1063/5.0239183
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Ultra-high field magnetic resonance imaging (MRI) offers significant advantages in terms of signal-to-noise ratio and spatial resolution. In this study, we detail the development of a multi-channel home-built MRI console operating at 14 T. We propose a hybrid analog-digital framework that shifts high-frequency radio frequency transmission and reception issues to lower frequencies, utilizing software-defined radio technology to process these low-frequency signals. Digital pre-emphasis is used in gradient calculations to counteract the effects of eddy currents during gradient switching. Our console can transmit and receive at center frequencies up to 600 MHz. The pulse programmer module achieves a timing resolution of 20 ns, while the transmitter can independently generate waveforms with varying amplitude, frequency, phase, and envelope. The receiver's dual-stage gain control provides 63 dB of adjustable range, optimizing the magnetic resonance (MR) signal's dynamic range. After frequency conversion, the MR signals are digitized with 16-bit resolution and 100 MHz sampling rate. High-resolution water phantom images are acquired on the 14 T Bruker Ascend 600 nuclear magnetic resonance magnet, demonstrating its potential for clinical research and application.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Multi-channel SQUID system for MEG and ultra-low-field MRI
    Zotev, Vadim S.
    Matlachov, Andrei N.
    Volegov, Petr L.
    Sandin, Henrik J.
    Espy, Michelle A.
    Mosher, John C.
    Urbaitis, Algis V.
    Newman, Shaun G.
    Kraus, Robert H., Jr.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2007, 17 (02) : 839 - 842
  • [2] Design of a multi-channel ultra-high resolution superconducting Gamma-ray spectrometer
    Friedrich, S
    Terracol, SF
    Miyazaki, T
    Drury, OB
    Ali, ZA
    Cunningham, MF
    Niedermayr, TR
    Barbee, TW
    Batteux, JD
    Labov, SE
    HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS VI, 2004, 5540 : 156 - 164
  • [3] Multi-channel helical-antenna inner-volume RF coils for ultra-high field MR scanners
    Athalye, Pranav S.
    Ilic, Milan M.
    van de Moortele, Pierre-Francois
    Kiruluta, Andrew J. M.
    Notaros, Branislav M.
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2018, 48B (04)
  • [4] Ultra-high field MRI in multiple sclerosis
    Inglese, M.
    MULTIPLE SCLEROSIS JOURNAL, 2017, 23 : 79 - 79
  • [5] Challenges and Opportunities of Ultra-High Field MRI
    Dula, A. N.
    Welch, E. B.
    Creasy, J. L.
    Gatenby, J. C.
    Stringer, E. A.
    Chen, L. M.
    Anderson, A. W.
    Avison, M. J.
    Gore, J. C.
    THIRD INTERNATIONAL CONFERENCE ON THE DEVELOPMENT OF BIOMEDICAL ENGINEERING IN VIETNAM, 2010, 27 : 1 - +
  • [6] Leveraging ultra-high field (7T) MRI in psychiatric research
    Calabro, Finnegan J.
    Parr, Ashley C.
    Sydnor, Valerie J.
    Hetherington, Hoby
    Prasad, Konasale M.
    Ibrahim, Tamer S.
    Sarpal, Deepak K.
    Famalette, Alyssa
    Verma, Piya
    Luna, Beatriz
    NEUROPSYCHOPHARMACOLOGY, 2024, 50 (01) : 85 - 102
  • [7] Intervertebral disc lesions: visualisation with ultra-high field MRI at 11.7 T
    Nikolaus Berger-Roscher
    Fabio Galbusera
    Volker Rasche
    Hans-Joachim Wilke
    European Spine Journal, 2015, 24 : 2488 - 2495
  • [8] Intervertebral disc lesions: visualisation with ultra-high field MRI at 11.7 T
    Berger-Roscher, Nikolaus
    Galbusera, Fabio
    Rasche, Volker
    Wilke, Hans-Joachim
    EUROPEAN SPINE JOURNAL, 2015, 24 (11) : 2488 - 2495
  • [9] Epilepsy under the scope of ultra-high field MRI
    Rondinoni, Carlo
    Magnun, Celso
    da Silva, Alexandre Vallota
    Heinsen, Helmut Manfred
    Amaro, Edson, Jr.
    EPILEPSY & BEHAVIOR, 2021, 121
  • [10] Ultra-high field MRI: looking through the 'macroscope'
    Kilsdonk, Iris D.
    Wattjes, Mike P.
    Geurts, Jeroen J. G.
    JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2014, 85 (01): : 4 - 4