Is Wear Still a Concern in Total Knee Arthroplasty With Contemporary Conventional and Highly Crosslinked Polyethylene Tibial Inserts in the mid- to Long-Term?

被引:0
作者
Asher, Devin P. [1 ]
Wright, Jennifer L. [1 ]
Hall, Deborah J. [1 ]
Lundberg, Hannah J. [1 ]
Van Citters, Douglas W. [2 ]
Jacobs, Joshua J. [1 ]
Levine, Brett R. [1 ]
Pourzal, Robin [1 ]
机构
[1] Rush Univ, Dept Orthoped Surg, Med Ctr, 1611 W Harrison St,Suite 204-F, Chicago, IL 60612 USA
[2] Dartmouth Coll, Thayer Sch Engn, Hanover, NH USA
关键词
Total knee arthroplasty; Tibial inserts; Polyethylene; Highly crosslinked; Wear; Delamination; OXIDATION; FAILURE; REPLACEMENT; VOLUME;
D O I
10.1016/j.artd.2024.101550
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Modern literature has brought into question if wear of tibial inserts made from conventional or highly crosslinked polyethylene (HXL PE) is still a factor limiting longevity of total knee arthroplasty (TKA) in the mid- to long-term. It is the objective of this study to determine: 1) most common causes of mid- to long-term TKA failure, 2) the prevalence of delamination, and 3) the medial/ lateral linear wear rates of conventional and HXL PE tibial inserts retrieved in the mid- to long-term. Methods: A tibial insert retrieval cohort of 107 inserts (79 conventional, 28 HXL PE) with a minimum time in situ of 6.5 years (mean 11.7 +/- 4) was studied. Failure causes were determined from chart-review, delamination presence was assessed microscopically, and medial/lateral linear wear was determined by minimal thickness changes measured with a dial-indicator. Results: The most common mid-to long-term etiologies for failure were instability (44.9%), PE wear 15%), aseptic loosening (14%), and infection (13.1%). Delamination occurred in 70% of inserts (72.1% conventional, 64.3% HXLPE). Gross material loss due to delamination appeared to be the underlying reason for at least 33.3% of cases exhibiting instability. Of the cases removed for infection, 75% exhibited no histopathological hallmarks of acute infection. The medial/lateral wear rates were 0.054/0.051 (conventional) and 0.014/0.011 (HXL) mm/y, respectively. Conclusions: Polyethylene wear still appears to be a major primary and secondary cause for TKA revision in the mid- to long-term. Wear may manifest as destabilizing delamination or as continuous release of fine wear particles potentially resulting in inflammatory responses and subsequent failure. (c) 2024 The Authors. Published by Elsevier Inc. on behalf of The American Association of Hip and Knee Surgeons. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/ licenses/by-nc-nd/4.0/).
引用
收藏
页数:7
相关论文
共 40 条
[1]  
AJRR, 2023, American joint replacement registry (AJRR): 2023 annual report
[2]  
American Joint Replacement Registry (AJRR), 2022, Annual Report
[3]   Knee Wear Measured in Retrievals: A Polished Tray Reduces Insert Wear [J].
Berry, Daniel J. ;
Currier, John H. ;
Mayor, Michael B. ;
Collier, John P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (07) :1860-1868
[4]   Gamma-irradiation Sterilization in an Inert Environment: A Partial Solution [J].
Berry, Daniel J. ;
Currier, Barbara H. ;
Mayor, Michael B. ;
Collier, John P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (07) :1805-1813
[5]   Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylene oxide [J].
Costa, L ;
Luda, MP ;
Trossarelli, L ;
del Prever, EMB ;
Crova, M ;
Gallinaro, P .
BIOMATERIALS, 1998, 19 (7-9) :659-668
[6]   Oxidation in Retrieved, Never-Irradiated UHMWPE Bearings What Can We Learn About in Vivo Oxidation? [J].
Currier, Barbara H. ;
Jevsevar, Kori C. ;
Van Citters, Douglas W. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2023, 105 (04) :293-301
[7]   Effectiveness of anti-oxidant polyethylene: What early retrievals can tell us [J].
Currier, Barbara H. ;
Currier, John H. ;
Holdcroft, Lindsay A. ;
Van Citters, Douglas W. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2018, 106 (01) :353-359
[8]   Comparison of Wear and Oxidation in Retrieved Conventional and Highly Cross-Linked UHMWPE Tibial Inserts [J].
Currier, Barbara H. ;
Currier, John H. ;
Franklin, Katherine J. ;
Mayor, Michael B. ;
Reinitz, Steven D. ;
Van Citters, Douglas W. .
JOURNAL OF ARTHROPLASTY, 2015, 30 (12) :2349-2353
[9]  
Currier BH, 1997, CLIN ORTHOP RELAT R, P111
[10]   What factors drive polyethylene wear in total knee arthroplasty? RESULTS OF A LARGE RETRIEVAL SERIES [J].
Currier, J. H. ;
Currier, B. H. ;
Abdel, M. P. ;
Berry, D. J. ;
Titus, A. J. ;
Van Citters, D. W. .
BONE & JOINT JOURNAL, 2021, 103B (11) :1695-1701