Statement of problem. Autonomous dental implant robot systems (ADIRSs) have been marketed for dental implant surgery; however, research on their accuracy is limited. Purpose. The purpose of this clinical study was to examine the accuracy of an ADIRS. Material and methods. Seventy-four participants with partial edentulism underwent implant surgery by using the ADIRS. Before surgery, a cone beam computed tomography (CBCT) scan was conducted, and its image data were fed into the ADIRS navigation and positioning system for iterative reconstruction and surgical path planning. After local anesthesia administration, area disinfection, and patient registration, the robot prepared the osteotomy and placed the implant under guidance of the navigation system. A postoperative CBCT scan was made to assess deviations between the planned and actual implant positions. The global coronal, apical, and angular deviations were measured and analyzed. A generalized linear mixed model (GLMM) was used to compare the differences in major outcome variables (alpha=.05). Results. All implant procedures were successfully carried out without any complications during surgery. A total of 86 implants were placed in 74 participants. The mean +/- standard deviation global coronal deviation, global apical deviation, and angular deviation was 0.61 +/- 0.20 mm (95% CI: 0.23 to 1.0 mm), 0.79 +/- 0.32 mm (95% CI: 0.17 to 1.41 mm), and 2.56 +/- 1.10 degrees (95% CI: 0.42 to 4.70 degrees), respectively. The findings from the GLMM revealed that the implant region (premolar, molar; maxilla, mandible; left, right), implant length, and surgeon did not significantly impact accuracy (P>.05). Conclusions. ADIRS showed high accuracy for implant placement because of the control of angular deviation and axial errors. The implant region, implant length, and surgeon had no significant influence on the accuracy of implant placement. Nevertheless, more clinical studies are required to support evidence-based clinical outcomes. (J Prosthet Dent 2025;133:764-770)
引用
收藏
页码:764 / 770
页数:7
相关论文
共 50 条
  • [21] The evolution of robotics: research and application progress of dental implant robotic systems
    Liu, Chen
    Liu, Yuchen
    Xie, Rui
    Li, Zhiwen
    Bai, Shizhu
    Zhao, Yimin
    INTERNATIONAL JOURNAL OF ORAL SCIENCE, 2024, 16 (01)
  • [22] Digital Technologies for Dental Implant Treatment Planning and Guided Surgery
    Greenberg, Alex M.
    ORAL AND MAXILLOFACIAL SURGERY CLINICS OF NORTH AMERICA, 2015, 27 (02) : 319 - +
  • [23] Review on Dental Implant and Infection Management Approach
    Alrakkad, Ibrahim Abdulkarim
    Alrakkad, Rakan Abdulkarim
    Altamimi, Mohammad Saud
    Alshammari, Nawaf Mubarak
    Alghuraymil, Abdulrahman Abdulaziz Salem
    John, Mohammed Ali M.
    Jalal, Rua Adnan
    Alsayegh, Sara Youssef
    Linjawi, Loulwah
    Otayn, Waseem Ahmed K.
    Alshabib, Zainab Yaseen Ahmed
    Alharthi, Shahad Hassan
    ARCHIVES OF PHARMACY PRACTICE, 2022, 13 (01) : 37 - 39
  • [24] Three-dimensional accuracy of partially guided implant surgery based on dental magnetic resonance imaging
    Schwindling, Franz Sebastian
    Juerchott, Alexander
    Boehm, Sophia
    Rues, Stefan
    Kronsteiner, Dorothea
    Heiland, Sabine
    Bendszus, Martin
    Rammelsberg, Peter
    Hilgenfeld, Tim
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (10) : 1218 - 1227
  • [25] Analysis of the accuracy of a dynamic navigation system when performing dental implant surgery with transcrestal sinus floor elevation: A pilot study
    Wu, Bin-Zhang
    Ma, Fei-Fei
    Sun, Feng
    JOURNAL OF DENTAL SCIENCES, 2023, 18 (04) : 1747 - 1755
  • [26] Is routine antibiotic prophylaxis warranted in dental implant surgery to prevent early implant failure? - a systematic review
    Momand, Palwasha
    Naimi-Akbar, Aron
    Hultin, Margareta
    Lund, Bodil
    Gotrick, Bengt
    BMC ORAL HEALTH, 2024, 24 (01):
  • [27] The relationship between dental implant papilla and dental implant mucosa around single-tooth implant in the esthetic area: A retrospective study
    Garabetyan, Julia
    Malet, Jacques
    Kerner, Stephane
    Detzen, Laurent
    Carra, Maria Clotilde
    Bouchard, Philippe
    CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (12) : 1229 - 1237
  • [28] Accuracy and operation procedure of robotic computer-aided implant surgery
    Liu, Yitian
    Zheng, Qingqing
    Xu, Zonghe
    Lin, Yanjun
    Zhang, Sihui
    Huang, Wenxiu
    Chen, Jiang
    Wu, Dong
    SCIENTIFIC REPORTS, 2025, 15 (01):
  • [29] Antibiotic prophylaxis patterns of Finnish dentists performing dental implant surgery
    Pyysalo, Mikko
    Helminen, Mika
    Antalainen, Anna-Kaisa
    Sandor, George K.
    Wolff, Jan
    ACTA ODONTOLOGICA SCANDINAVICA, 2014, 72 (08) : 806 - 810
  • [30] Analysis of the mylohyoid nerve in elderly Japanese cadavers for dental implant surgery
    Koga, Shintaro
    Sato, Iwao
    Li, Zhong-Lian
    Miyaso, Hidenobu
    Kawata, Shinichi
    Itoh, Masahiro
    CLINICAL AND EXPERIMENTAL DENTAL RESEARCH, 2021, 7 (01): : 20 - 32