Quantitative analysis of protective Kirschner wire diameters in lateral opening wedge distal femoral osteotomy: A finite element study

被引:0
作者
Baris, Alican [1 ]
Ozmen, Emre [1 ]
Circi, Esra [1 ]
Yuksel, Serdar [1 ]
Beytemur, Ozan [2 ]
机构
[1] Istanbul Phys Med & Rehabil Training & Res Hosp, Dept Orthoped & Traumatol, Istanbul, Turkiye
[2] Bagcilar Training & Res Hosp, Dept Orthoped & Traumatol, Istanbul, Turkiye
关键词
Biomechanics; distal femoral osteotomy; finite element analysis; hinge fracture; protective K-wire; FRACTURE; BONE;
D O I
10.52312/jdrs.2025.1806
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: This study aims to investigate quantitatively the protective effect of a 1.6-mm or a 2.5-mm Kirschner wire (K-wire) on the medial hinge at different gap distances through finite element analysis (FEA) and to establish whether using a 2.5-mm K-wire can offer benefits compared to a 1.6-mm in preventing medial hinge fractures. Materials and methods: Between June 2024 and July 2024, three different models simulating a lateral opening wedge (LOW) osteotomy of the distal femur were created from a femoral computed tomography (CT) scan of a 36-year-old male patient: no K-wire (Model I), 1.6-mm K-wire (Model II), and 2.5-mm K-wire (Model III). Finite element analysis was performed to simulate 7- to 13-mm gaps at the osteotomy site. Loads, principal stress, strain, and equivalent stress were analyzed around the medial hinge. Results: Model I required 123.0 +/- 5.2 N, Model II required 181.7 +/- 12.2 N, and Model III required 228.7 +/- 13.6 N (p<0.001). Cracked shell elements were the lowest in Model II and the highest in Model I. While the average equivalent/yield stress ratio was not significantly lower in Model II compared to Model III (87.0 +/- 10.9% vs. 92.7 +/- 12.1%), the maximum equivalent/yield stress ratio values in Model II were significantly lower than both Model I and Model III (1206.2 +/- 138.3% vs. 1836.2 +/- 165.4% and 1689.1 +/- 404.0%, respectively), suggesting a superior dispersion of forces. Conclusion: Using a 1.6-mm K-wire during LOW osteotomy of the distal femur provides a balance between structural reinforcement and stress distribution, significantly improving stability and reducing the risk of medial hinge fractures compared to a 2.5-mm K-wire or no K-wire. The 1.6-mm K-wire optimizes stress dispersion, making it the preferred choice for surgical planning in lateral opening wedge distal femoral osteotomy.
引用
收藏
页码:97 / 106
页数:10
相关论文
共 32 条
[21]   Assessing the Local Mechanical Environment in Medial Opening Wedge High Tibial Osteotomy Using Finite Element Analysis [J].
Pauchard, Yves ;
Ivanov, Todor G. ;
McErlain, David D. ;
Milner, Jaques S. ;
Giffin, J. Robert ;
Birmingham, Trevor B. ;
Holdsworth, David W. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2015, 137 (03)
[22]   Ten-year minimal follow-up of lateral opening wedge distal femoral osteotomy for lateral femorotibial osteoarthritis: Good survivorship and high patient satisfaction [J].
Cance, Nicolas ;
Batailler, Cecile ;
Lording, Timothy ;
Schmidt, Axel ;
Lustig, Sebastien ;
Servien, Elvire .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2025, 33 (02) :675-685
[23]   Biomechanical evaluation of opening-wedge high tibial osteotomy with composite materials using finite-element analysis [J].
Koh, Yong-Gon ;
Son, Juhyun ;
Kwon, Sae Kwang ;
Kim, Hyo Jeong ;
Kang, Kyoung-Tak .
KNEE, 2018, 25 (06) :977-987
[24]   Biomechanical Behavior Analysis of Tibia According to Types of Fractures in Opening Wedge High Tibial Osteotomy Based on Finite Element Method [J].
Sim, On ;
Rhee, Seung Joon ;
Lee, Chiseung .
TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2023, 47 (12) :607-614
[25]   Computer-aided Design of Distal Femoral Osteotomy for the Valgus Knee and Effect of Correction Angle on Joint Loading by Finite Element Analysis [J].
Wu, Yanfei ;
Jin, Xin ;
Zhao, Xingwen ;
Wang, Ying ;
Bai, Haohao ;
Lu, Bin ;
Tong, Xue ;
Ma, Jianxiong ;
Ma, Xinlong .
ORTHOPAEDIC SURGERY, 2022, 14 (11) :2904-2913
[26]   Effects of K-wire diameter and insertion angle on femoral bone medial closing-wedge osteotomies: a finite element study [J].
Ayda K. Dastgerdi ;
Alireza Y. Bavil ;
Markus T. Berninger ;
Imke A. K. Fiedler ;
Björn Busse ;
Matthias Krause ;
Felix N. von Brackel .
Scientific Reports, 15 (1)
[27]   A quantitative biomechanical study of positive buttress techniques for femoral neck fractures: a finite element analysis [J].
Wang Gang ;
Wang Bin ;
Tang Yong ;
Yang Hui-Lin .
中华医学杂志英文版, 2019, 132 (21) :2588-2593
[28]   A quantitative biomechanical study of positive buttress techniques for femoral neck fractures: a finite element analysis [J].
Wang, Gang ;
Wang, Bin ;
Tang, Yong ;
Yang, Hui-Lin .
CHINESE MEDICAL JOURNAL, 2019, 132 (21) :2588-2593
[29]   Mechanical stability of the proximal tibia with different bone formations after plate removal in medial opening-wedge high tibial osteotomy: a finite element analysis [J].
Jang, Chul-Young ;
Kang, Kyoung-Tak ;
Hong, Hyongtaek ;
Jung, Min ;
Kim, Sungjun ;
Yoo, Je-Hyun ;
Kim, Sung-Hwan .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01)
[30]   Favorable rates of return to activity and work following lateral closing wedge distal femoral osteotomy for femoral-based symptomatic varus malalignment: an analysis at a mean 6-year follow-up [J].
Marco-Christopher Rupp ;
Alexander Themessl ;
Michael Merkle ;
David Insam ;
Maximilian Hinz ;
Franziska L. Breulmann ;
Andrea Achtnich ;
Julian Mehl ;
Sebastian Siebenlist ;
Lukas N. Muench .