Sensitive LC-MS/MS Assay for Total Testosterone Quantification on Unit Resolution and High-Resolution Instruments

被引:0
|
作者
Wolken, Jill K. [1 ]
Peterson, Meghan M. [1 ]
Cao, Wenjing [1 ,2 ]
Challoner, Keith [1 ]
Jin, Zhicheng [1 ,2 ]
机构
[1] Univ Wisconsin Hosp & Clin, Madison, WI 53792 USA
[2] Univ Wisconsin Madison, Dept Pathol & Lab Med, 600 Highland Ave, Madison, WI 53792 USA
关键词
testosterone; multiple reaction monitoring; high-resolution mass spectrometry; assay harmonization; parallel reaction monitoring; TANDEM MASS-SPECTROMETRY; SERUM TESTOSTERONE; VALIDATION;
D O I
10.3390/jcm13237056
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Testosterone is an androgenic hormone that plays important roles in both males and females. The circulating levels of total testosterone vary from 1 to 1480 ng/dL. High-throughput immunoassays often lack accuracy in lower concentration ranges (below 100 ng/dL), particularly when used for females or children. To address this limitation, we developed a total testosterone LC-MS/MS assay on three instruments. Methods: Sample preparation began with the dilution and conditioning of 200 mu L of serum. A supported liquid extraction cartridge was used to extract the analyte from biological matrices. Chromatographic separation was achieved using a C18 column with a runtime of 5 min per sample. This assay was validated on a Triple Quad 6500 and an API 4500 instrument. Results: Method validation was carried out according to the CLSI C62-ED2 guideline and our hospital protocol. The within-day coefficient of variation (CV) was less than 10% and the between-day CV was less than 15%. The assay had a limit of quantitation of 0.5 ng/dL with an analyte measure range of 2-1200 ng/dL. A comparison using Deming regression and Bland-Altman plots showed that this assay correlated well with a reference method. The results from the API 4500 and an Orbitrap were consistent with those from the TQ 6500. Both serum-separator tubes (BD) and serum-activator tubes were found to be suitable. Conclusions: We successfully developed and validated a robust total testosterone LC-MS/MS assay for routine clinical testing. This assay was harmonized across two triple quadrupole instruments and one high-resolution mass spectrometer.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Quantification of free and total sialic acid excretion by LC-MS/MS
    van der Ham, Maria
    Prinsen, Berthil H. C. M. T.
    Huijmans, Jan G. M.
    Abeling, Nicolaas G. G. M.
    Dorland, Bert
    Berger, Ruud
    de Koning, Tom J.
    de Sain-van der Velden, Monique G. M.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 848 (02): : 251 - 257
  • [32] High-resolution MS: first choice for peptide quantification?
    Dillen, Lieve
    Cuyckens, Filip
    BIOANALYSIS, 2013, 5 (10) : 1145 - 1148
  • [33] Highly sensitive feature detection for high resolution LC/MS
    Ralf Tautenhahn
    Christoph Böttcher
    Steffen Neumann
    BMC Bioinformatics, 9
  • [34] A robust LC-MS/MS assay with online cleanup for measurement of serum testosterone
    Zhou, Yu
    Wang, Sihe
    JOURNAL OF SEPARATION SCIENCE, 2019, 42 (15) : 2561 - 2568
  • [35] Highly sensitive feature detection for high resolution LC/MS
    Tautenhahn, Ralf
    Bottcher, Christoph
    Neumann, Steffen
    BMC BIOINFORMATICS, 2008, 9 (1)
  • [36] 13C-labelled yeast as internal standard for LC-MS/MS and LC high resolution MS based amino acid quantification in human plasma
    Hermann, G.
    Schwaiger, M.
    Volejnik, P.
    Koellensperger, G.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 155 : 329 - 334
  • [37] maXis High Resolution LC-MS Makes the Most of Ultrafast LC Separations
    Stacey, Catherine
    Goetz, Sebastian
    Baessmann, Carsten
    LC GC EUROPE, 2008, : 40 - 41
  • [38] Development of a CDC-certified total testosterone assay for adult and pediatric samples using LC-MS/MS
    Star-Weinstock, Michal
    Dey, Subhakar
    CLINICAL MASS SPECTROMETRY, 2019, 13 : 27 - 35
  • [39] Simultaneous quantification of total and free testosterone in human serum by LC–MS/MS
    Rongmei Huang
    Yi Hong
    Yike Wu
    Weifeng Li
    Wenlan Liu
    Analytical and Bioanalytical Chemistry, 2023, 415 : 6851 - 6861
  • [40] A high-throughput on-line extraction LC-MS/MS assay for testosterone in human serum
    Nelson, RE
    Grebe, SK
    Singh, RJ
    CLINICAL CHEMISTRY, 2005, 51 : A242 - A242