Intrapartum monitoring of the fetal heart rate using transabdominal electrocardiography: A reliability and accuracy study

被引:0
|
作者
Wojtanowski, Anne [1 ,2 ]
Garabedian, Charles [2 ,3 ]
Charlier, Pierre [1 ,2 ]
Prot, Pierre [4 ]
Ghesquiere, Louise [2 ,3 ]
De Jonckheere, Julien [1 ,2 ]
机构
[1] CHU Lille, CIC IT 1403, F-59000 Lille, France
[2] Univ Lille, ULR 2694 METRICS, F-59000 Lille, France
[3] CHU Lille, Dept Obstet, F-59000 Lille, France
[4] BioSerenity, F-75013 Paris, France
关键词
Fetal heart rate monitoring; Non-invasive fetal electrocardiography; Signal quality; AUTONOMIC RESPONSE; RATE-VARIABILITY; LABOR;
D O I
10.1016/j.jogoh.2025.102942
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Introduction: In this study, we evaluated the performance of an intrapartum monitoring system (the TOCONAUTE) for fetal heart rate estimation using transabdominal electrocardiography. Method: This prospective study was conducted in Lille University Hospital's maternity ward between August 2022 and November 2023 to assess the performance of the TOCONAUTE compared to a cardiotocograph (CTG). Pregnant women in labor at 36 weeks or more of gestation with a singleton in cephalic presentation were eligible for enrollment in the study. In addition to the cardiotocograph, the participants received an abdominal electrode patch after providing their written informed consent. The positive percent agreement with the cardiotocograph indicated reliability, and the Bland-Altman analysis determined accuracy. Results: Of the 60 patient recruited, 35 patients were included in the recording analysis. The recordings had a mean duration of 519 f 210 min. The TOCONAUTE's fetal heart rate (FHR) overall positive percent of agreement (PPA) was 84.2 f 10.4 %. The PPA was greater than 80 % for all labor phases, except for expulsion (44.7 %). The overall mean bias was 1.7 f 2.4. For the FHR, the mean bias was lower than 2 bpm for all labor phases, except for expulsion (14.13 f 15.04 bpm). Conclusion: Compared with traditional CTG, this study demonstrated the TOCONAUTE's reliability and accuracy for FHR estimation in all labor phases except expulsion.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] INTRAPARTUM FETAL HEART-RATE MONITORING .4. OBSERVATIONS ON ELECTIVE AND NON-ELECTIVE FETAL HEART-RATE MONITORING
    KREBS, HB
    PETRES, RE
    DUNN, LJ
    SEGRETI, A
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1980, 138 (02) : 213 - 219
  • [42] INTRAPARTUM FETAL HEART-RATE MONITORING .3. ASSOCIATION OF MECONIUM WITH ABNORMAL FETAL HEART-RATE PATTERNS
    KREBS, HB
    PETRES, RE
    DUNN, LJ
    JORDAAN, HVF
    SEGRETI, A
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1980, 137 (08) : 936 - 943
  • [43] Importance of electronic monitoring of intrapartum fetal heart rate as a predictor of neonatal outcomes
    Fernanda Nozar, Maria
    Fiol, Veronica
    Martinez, Alma
    Enrique Pons, Jose
    Alonso, Justo
    Briozzo, Leonel
    REVISTA MEDICA DEL URUGUAY, 2008, 24 (02): : 94 - 101
  • [44] Intrapartum electronic fetal heart rate monitoring and the prevention of perinatal brain injury
    Graham, Ernest M.
    Petersen, Scott M.
    Christo, Dana K.
    Fox, Harold E.
    OBSTETRICS AND GYNECOLOGY, 2006, 108 (03): : 656 - 666
  • [45] THE CLINICAL-SIGNIFICANCE OF ACCELERATIONS IN INTRAPARTUM FETAL HEART-RATE MONITORING
    RICHTER, R
    HOHL, M
    LUSCHER, KP
    STUCKI, D
    ZEITSCHRIFT FUR GEBURTSHILFE UND PERINATOLOGIE, 1980, 184 (03): : 195 - 199
  • [46] INTRAPARTUM FETAL HEART-RATE ASSESSMENT - MONITORING BY AUSCULTATION OR ELECTRONIC MEANS
    MORRISON, JC
    CHEZ, BF
    DAVIS, ID
    MARTIN, RW
    ROBERTS, WE
    MARTIN, JN
    FLOYD, RC
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1993, 168 (01) : 63 - 66
  • [47] Risk Management in Intrapartum Fetal Monitoring Accidental Recording of the Maternal Heart Rate
    Hanson, Lisa
    JOURNAL OF PERINATAL & NEONATAL NURSING, 2010, 24 (01) : 7 - 9
  • [48] Utility of Intrapartum Fetal Heart Rate Monitoring in a Contemporary Labor and Delivery Unit
    Taylor, Rachel
    Burk, Camden
    Sugiyama, Nathaniel
    Friedman, Perry
    Ogunyemi, Dotun
    OBSTETRICS AND GYNECOLOGY, 2017, 129 : 174S - 174S
  • [49] THE DUBLIN RANDOMIZED CONTROLLED TRIAL OF INTRAPARTUM FETAL HEART-RATE MONITORING
    MACDONALD, D
    GRANT, A
    SHERIDANPEREIRA, M
    BOYLAN, P
    CHALMERS, I
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1985, 152 (05) : 524 - 539
  • [50] Bayesian interpretation of trials: the example of intrapartum electronic fetal heart rate monitoring
    Hornbuckle, J
    Vail, A
    Abrams, KR
    Thornton, JG