Biomechanical analysis of partial mandibular implants with various lattice designs of different material properties: In vitro study and finite element analysis

被引:0
|
作者
Zhang, Hao [1 ]
Fuh, Lih Jyh [1 ,2 ]
Hsu, Jui Ting [3 ]
Lim, Zhe Min [1 ]
Huang, Heng Li [1 ,4 ]
机构
[1] China Med Univ, Sch Dent, Taichung, Taiwan
[2] China Med Univ Hosp, Dept Dent, Taichung, Taiwan
[3] China Med Univ, Dept Biomed Engn, Taichung, Taiwan
[4] Asia Univ, Dept Bioinformat & Med Engn, Taichung, Taiwan
关键词
Mandibular segmental resection; Porous lattice design; Lattice size; Rod diameter; Mandibular implant; Abutment; In vitro experiment; Finite element analysis; PORE-SIZE; RECONSTRUCTION; BONE; PROSTHESIS; STRESS;
D O I
10.36922/ijb.3943
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For patients with mandibular bone defects, although reconstruction plates can be used for repair, achieving both occlusal function and facial aesthetics is challenging. In the present study, in vitro experiments and finite element analysis (FEA) were conducted to determine the biomechanical characteristics of multiple porous lattice structures of varying shapes and diameters that were used for mandibular implants. Additionally, an abutment designed to carry occlusal forces was added to the top of the implants. The stress distribution of four lattice designs (tetrahedron, quad-diametral-cross, hex-star, and hex-vase) of three sizes (2.5, 3.0, and 3.5 mm) in cubic porous models were analyzed by FEA. Subsequently, two optimal designs for 3D-printed titanium alloy were selected. These designs, featuring different lattice diameters (0.5, 0.7, and 0.9 mm), were tested to determine their elastic modulus, which was used in another FEA of a mandibular implant designed for a patient with a malignant tumor in the right mandible. This model, which included an abutment design, was subjected to a vertical force of 100 N and muscle forces generated by biting. This analysis was conducted to determine the elastic modulus of the implant and the values of stress and strain on the implant and surrounding bone. The lattice designs of quad-diametral-cross and hex-vase exhibited smaller high-stress regions than those of tetrahedron and hex-star. In vitro tests revealed that the elastic modulus of the lattices increased with the rod diameter. When these values were applied to mandibular implants, Young's modulus decreased, which in turn increased the frictional stress observed at the interface between the abutment and the implant. However, the implant's maximum stress remained below its yield strength (910 MPa), and the strain on the surrounding bone varied between 1500 and 3000 mu strain. As indicated by Frost's theory, these implants are unlikely to damage the surrounding bone tissue and are likely to support bone growth.
引用
收藏
页码:567 / 581
页数:15
相关论文
共 50 条
  • [31] Comparative Finite Element Analysis of Short Implants with Different Treatment Approaches in the Atrophic Mandible
    Doganay, Ozge
    Kilic, Erdem
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2020, 35 (04) : E69 - E76
  • [32] Biomechanical analysis of distalization of mandibular molars by placing a mini-plate: A finite element study
    Park, Myungsoon
    Na, Yonghyun
    Park, Minbong
    Ahn, Janghoon
    KOREAN JOURNAL OF ORTHODONTICS, 2017, 47 (05) : 289 - 297
  • [33] Biomechanical Factors Associated with Mandibular Cantilevers: Analysis with Three-Dimensional Finite Element Models
    Gonda, Tomoya
    Yasuda, Daiisa
    Ikebe, Kazunori
    Maeda, Yoshinobu
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2014, 29 (06) : E275 - E282
  • [34] Finite-element-analysis and in vitro study of bioabsorbable polymer stent designs
    Schultze, C.
    Grabow, N.
    Martin, H.
    Schmitz, K. -P.
    4TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2009, 22 (1-3): : 2175 - 2178
  • [35] Influence of Mechanical Properties of Dental Implants on Biomechanical Behavior: A Three-Dimensional Finite Element Analysis
    Goradia, Youcef
    Gouasmi, Sadek
    Reffaas, Sid Ahmed
    JOURNAL OF BIOMIMETICS BIOMATERIALS AND BIOMEDICAL ENGINEERING, 2024, 64 : 39 - 54
  • [36] Stress analysis of mandibular Kennedy class I partial overdenture submitted to different implant length and clasp designs - An in vitro study
    Mahross, Hamada Z.
    Atito, Ehab M.
    Mansour, Mamdouh M.
    Badr, Wesam E.
    Osman, Mohamed S.
    Quassem, Mohamed A.
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2024,
  • [37] Biomechanical Analysis of Functionally Graded Root Analog Implants on Alveolar Bone: A 3D Finite Element Study
    Wang, Feilong
    Jiang, Zhaoliang
    Si, Pengchao
    Lan, Jing
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2021, 36 (04) : E63 - E71
  • [38] Comparison of Biomechanical Behaviors of Different Designs and Configurations of Titanium and Zirconium Dental Implants With Finite Elements Analysis in Anterior Maxilla
    Izgi, Eda
    Simsek, Mehmet Baris
    JOURNAL OF ORAL IMPLANTOLOGY, 2024, 50 (03) : 277 - 287
  • [39] Biomechanical study of tarsometatarsal joint fusion using finite element analysis
    Wang, Yan
    Li, Zengyong
    Zhang, Ming
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (11) : 1394 - 1400
  • [40] Biomechanical finite element analysis of self-tapping implants with different dimensions inserted in two bone qualities
    Hasan, Istabrak
    Heinemann, Friedhelm
    Bourauel, Christoph
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 (03): : 203 - 211