Microwave Radiometer Calibration Using Low-Noise Amplifier as Calibration Source

被引:0
|
作者
Neate, Reuben [1 ]
Stander, Tinus [1 ]
机构
[1] Univ Pretoria, Carl & Emily Fuchs Inst Microelect, Dept EEC Engn, ZA-0002 Pretoria, South Africa
来源
关键词
Calibration; Noise; Temperature measurement; Microwave radiometry; Noise measurement; Switches; Microwave theory and techniques; Gain; Power measurement; Microwave technology; Atmospheric monitoring; noise calibration; remote sensing; water vapor radiometer (WVR); Y-factor;
D O I
10.1109/LMWT.2024.3474971
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present a novel method for built-in noise and gain calibration of a total power radiometer using a low-noise amplifier (LNA) as the calibration source. A terminated LNA's biasing is varied to obtain multiple calibration points, with the novel postprocessing method accounting for both radiometer nonlinearity and unknown LNA parameters. This method allows for a built-in radiometer calibration without a calibrated resistive or diode noise source, instead using an internal thermometer measurement as a reference. The method is demonstrated by measuring a thermal load to a 24-GHz water vapor radiometer (WVR), achieving a mean error of 0.43 K over the band of interest under 45 degrees C ambient temperature variation. It is further shown that the calibration data can be used to estimate the radiometer's gain to within 1.41 dB and noise figure (NF) to within 0.55 dB.
引用
收藏
页码:127 / 130
页数:4
相关论文
共 50 条
  • [1] MEASURING AMPLIFIER NOISE ON A NOISE SOURCE CALIBRATION RADIOMETER
    WILLIAMS, GL
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1995, 44 (02) : 340 - 342
  • [2] Calibration of a Fully Polarimetric Microwave Radiometer Using a Digital Polarimeric Noise Source
    Lim, Boon H.
    Ruf, Christopher S.
    2006 IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, VOLS 1-8, 2006, : 3992 - 3995
  • [3] Microwave active cold/warm noise source for calibration of radiometer sensors
    Smith, MC
    Roeder, RS
    PASSIVE MILLIMETER-WAVE IMAGING TECHNOLOGY II, 1998, 3378 : 123 - 131
  • [4] A variable active cold noise source for calibration of medical microwave radiometer
    Vidyalakshmi, M. R.
    Arunachalam, Kavitha
    2015 IEEE APPLIED ELECTROMAGNETICS CONFERENCE (AEMC), 2015,
  • [5] Low-noise cryogenic microwave amplifier characterization with a calibrated noise source
    Malnou, M.
    Larson, T. F. Q.
    Teufel, J. D.
    Lecocq, F.
    Aumentado, J.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2024, 95 (03):
  • [6] A correlated noise calibration source for the internal calibration of a fully polarimetric radiometer
    Oh, Myeonggeun
    Kim, Yong-Hoon
    MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2019, 61 (01) : 73 - 76
  • [7] MICROWAVE WAVEGUIDE SWITCH FOR A LOW-NOISE RADIOMETER
    GABDRAKIPOVA, MN
    ILIN, VA
    FATYKHOV, KZ
    ETKIN, VS
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1984, 27 (01) : 161 - 162
  • [8] Calibration of scanning low frequency microwave radiometer
    Prytz, A
    Heron, ML
    Burrage, DM
    Goodberlet, A
    OCEANS 2002 MTS/IEEE CONFERENCE & EXHIBITION, VOLS 1-4, CONFERENCE PROCEEDINGS, 2002, : 2003 - 2007
  • [9] A LOW-NOISE MICROWAVE QUADRUPOLE AMPLIFIER
    ASHKIN, A
    PROCEEDINGS OF THE INSTITUTE OF RADIO ENGINEERS, 1961, 49 (06): : 1016 - &
  • [10] A New Linearly Polarized Microwave Radiometer Calibration Source
    Wang, Xinbiao
    Zeng, Chang
    Li, Jing
    Jiang, Jingshan
    ADVANCED MATERIALS IN MICROWAVES AND OPTICS, 2012, 500 : 341 - 347